Reliable Clinical Monitoring using Wireless Sensor Networks: Experience in a Step-down Hospital Unit. Yetta. Outline. Introduction Monitoring system Clinical study Clinical deterioration detection Conclusion. Introduction. Clinical deterioration detection

saddam
Télécharger la présentation

Yetta

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Reliable Clinical Monitoring using Wireless Sensor Networks: Experience in a Step-down Hospital Unit Yetta

  2. Outline • Introduction • Monitoring system • Clinical study • Clinical deterioration detection • Conclusion

  3. Introduction • Clinical deterioration detection • ICU / step-down unit / general care unit • IEEE 802.15.4 / IEEE 802.11 • Heart rate (HR) and blood oxygenation (spO2)

  4. Monitoring System • TelosB / OxiLink pulse-oximeter

  5. Monitoring System • CTP (collection tree protocol) • Low reliability because of user mobility • DRAP (Dynamic Relay Association Protocol) • Isolate the mobility from multi-hop routing • Single-hop to first relay • Relay to base station

  6. neighbor table of node A E A D B C

  7. Monitoring System • Radio power management • Sensor component (OxiLink pulse-oximeter) • Control by TelosB • average over 8 sec • Logging component • Batching flash writing

  8. Clinical Study • 1200m2 • 18 relays • 41 patients • Pulse and oxygenation were measured at 30- and 60-second intervals

  9. Reliability • Network reliability • Sensing reliability • Time-to-failure • Time-to-recover

  10. Network Reliability Mean = 22.4 min 95% < 2.5 min

  11. Sensing Reliability • Significantly affected by patient movement, sensor disconnections, sensor placement, and nail polish

  12. Improvement of Sensing Reliability • Oversample • Median reliability: 84%(30sec), 75%(60sec) Median = 1.81 min 75% < 1 min => short burst Long-tailed => sensor disconnection

  13. Improvement of Sensing Reliability • Disconnection alarms

  14. Clinical Deterioration Detection

  15. Clinical Deterioration Detection • CUSUM algorithm • detecting statistically significant changes in a series of measurements • Sliding window

  16. Conclusion • High network reliability • System reliability dominated by sensor reliability • Oversampling • Disconnection alarms • Show the potential of real-time detection of clinical deterioration