1 / 43

Nuclear Medicine SPECT & Gamma Camera

Nuclear Medicine SPECT & Gamma Camera. Eugen Kvasnak, PhD. Department of Medical Biophysics and Informatics 3rd Medical Faculty of Charles University. SPECT. Why SPECT?

tadhg
Télécharger la présentation

Nuclear Medicine SPECT & Gamma Camera

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Nuclear Medicine SPECT & Gamma Camera Eugen Kvasnak, PhD. Department of Medical Biophysics and Informatics 3rd Medical Faculty of Charles University

  2. SPECT Why SPECT? Similar to X-ray Computed Tomography (CT) or Magnetic Resonance Imaging (MRI), Single Photon Emission Computed Tomography (SPECT) allows us to visualize functional information about a patient's specific organ or body system.

  3. SPECT How does SPECT manage to give us functional information? Internal radiation is administered by means of a pharmaceutical which is labeled with a radioactive isotope. This so called radiopharmaceutical, or tracer, is either injected, ingested, or inhaled. The radioactive isotope decays, resulting in the emission of gamma rays. These gamma rays give us a picture of what's happening inside the patient's body.

  4. SPECT But how do these gamma rays allow us to see inside? By using the most essential tool in Nuclear Medicine, the gamma camera. The gamma camera can be used in planar imaging to acquire 2-dimensional images, or in SPECT imaging to acquire 3-dimensional images.

  5. SPECT What is SPECT? SPECT is short for Single Photon Emission Computed Tomography. As its name suggests (single photon emission), gamma ray emissions are the source of information, rather than X-ray transmissions as used in conventional Computed Tomography.

  6. SPECT How are these gamma rays collected? The Gamma camera collects gamma rays that are emitted from within the patient, enabling us to reconstruct a picture of where the gamma rays originated. From this, we can determine how a particular organ or system is functioning.

  7. SPECT … history • Although the first instance of SPECT was when Kuhl and Edwards produced the first tomographs from emission data in 1963, the history of SPECT detectors begins earlier. In the 1940's crude spatial information about radioactive source distributions within the brain were produced using a single detector positioned at various locations around the head. Ben Classen improved this method in the 1950's when he invented the rectilinear scanner. This device produced planar images by mechanically scanning a detector in a raster-like pattern over the area of interest. By today's standards, this technique required very long imaging times because of the sequential nature of the scanning. A pin-hole in lead was used to project a gamma ray image of the source distribution in 1953 by Hal Anger. The image was projected onto a scintillating screen with photographic film behind it. This technique required extremely long exposure times because of the huge inefficiencies in the system (principally due to losses in the film). The inefficiencies in the system resulted in extremely high radiation doses to patients.

  8. SPECT … history • In the late 1950's, Anger replaced the film and screen with a single NaI crystal and PMT array. This formed the basis for the "Anger Camera" which is now the standard clinical nuclear imaging device. Modern Anger Cameras use a lead collimator perforated with many parallel, converging or diverging holes instead of the original pin-hole configuration. Kuhl and Edwards were the first to present tomographic images produced using the Anger Camera in 1963. Everett, Fleming, Todd and Nightengale suggested the use of the Compton effect for gamma-radiation imaging in 1977. This technique is currently in use in astronomy. It's adaptation to SPECT is non-trivial because of the vastly different source distributions and geometry involved. The investigation of the Compton Camera for SPECT began in 1983. Manbir Singh and David Doria proposed and experimented with a basic design using solid state detectors, performed an analysis of possible detector materials, and produced a small prototype for testing.

  9. Once a radiopharmaceutical has been administered, it is necessary to detect the gamma ray emissions in order to attain the functional information.  The instrument used in Nuclear Medicine for the detection of gamma rays is known as the Gamma camera.  The components making up the gamma camera are the collimator, detector crystal, photomultiplier tube array, position logic circuits, and the data analysis computer.  The purpose of each is briefly described below.

  10. 1. Camera Collimator The first object that an emitted gamma photon encounters after exiting the body is the collimator. The collimator is a pattern of holes through gamma ray absorbing material, usually lead or tungsten, that allows the projection of the gamma ray image onto the detector crystal.  The collimator achieves this by only allowing those gamma rays traveling along certain directions to reach the detector;  this ensures that the position on the detector accurately depicts the originating location of the gamma ray.

  11. 2. Scintillation Detector In order to detect the gamma photon we use scintillation detectors.  A Thallium-activated Sodium Iodide [NaI(Tl)] detector crystal is generally used in Gamma cameras.  This is due to this crystal's optimal detection efficiency for the gamma ray energies of radionuclide emission common to Nuclear Medicine.  A detector crystal may be circular or rectangular. It is typically 3/8" thick and has dimensions of 30-50 cm. A gamma ray photon interacts with the detector by means of the Photoelectric Effect or Compton Scattering with the iodide ions of the crystal.  This interaction causes the release of electrons which in turn interact with the crystal lattice to produce light, in a process known as scintillation.

  12. 3. Photomultiplier Tubes Only a very small amount of light is given off from the scintillation detector.  Therefore, photomultiplier tubes are attached to the back of the crystal.  At the face of a photomultipler tube (PMT) is a photocathode which, when stimulated by light photons, ejects electrons.  The PMT is an instrument that detects and amplifies the electrons that are produced by the photocathode.  For every 7 to 10 photons incident on the photocathode, only one electron is generated.  This electron from the cathode is focused on a dynode which absorbs this electron and re-emits many more electrons (usually 6 to 10).  These new electrons are focused on the next dynode and the process is repeated over and over in an array of dynodes.  At the base of the photomultiplier tube is an anode which attracts the final large cluster of electrons and converts them into an electrical pulse.

  13. A Photomultiplier Tube Array Each gamma camera has several photomultiplier tubes arranged in a geometrical array.  The typical camera has 37 to 91 PMT's.

  14. 4. Position Circuitry The position logic circuits immediately follow the photomultiplier tube array and they receive the electrical impulses from the tubes in the summing matrix circuit (SMC).  This allows the position circuits to determine where each scintillation event occurred in the detector crystal.

  15. 5. Data Analysis Computer Finally, in order to deal with the incoming projection data and to process it into a readable image of the 3D spatial distribution of activity within the patient, a processing computer is used.  The computer may use various different methods to reconstruct an image, such as filtered back projection or iterative reconstruction, both of which are further described in this tutorial.

  16. Acquisition Protocols Various different acquisitions can be performed with a SPECT camera…. • Planar Imaging • Planar Dynamic Imaging • SPECT Imaging • Gated SPECT Imaging

  17. 1. Planar Imaging •     The simplest acquisition protocol is the planar image.  With planar imaging, the detector array is stationary over the patient, and acquires data only from this one angle.  The image created with this type of acquisition is similar to an X-ray radiograph.  Bone scans are done primarily in this fashion.

  18. 2. Planar Dynamic Imaging •     Since the camera remains at a fixed position in a planar study, it is possible to observe the motion of a radiotracer through the body by acquiring a series of planar images of the patient over time.  Each image is a result of summing data over a short time interval, typically 1-10 seconds.  If many projections are taken over a long time, then an animation of the tracer movement can be viewed and data analysis can be performed.  The most common dynamic planar scan is to measure glomerular filtration rate in the kidneys.

  19. 3. SPECT Imaging •     If one rotates the camera around the patient, the camera will acquire views of the tracer distribution at a variety of angles. After all these angles have been observed, it is possible to reconstruct a three dimensional view of the radiotracer distribution within the body. 

  20. 4. Gated SPECT Imaging • As the heart is a moving object, by performing a regular SPECT of the heart, the end image obtained will represent the average position of the heart over the time the scan was taken.  It is possible to view the heart at various stages of its contraction cycle however, by subdividing each SPECT projection view into a series of sub-views, each depicting the heart at a different stage of it's cycle.  In order to do this, the SPECT camera must be connected to an ECG machine which is measuring the heart beat.

  21. Reconstruction • The most common algorithm used in the tomographic reconstruction of clinical data is the filtered backprojectionmethod.  Other methods also exist, please refer to the section on iterative reconstruction methods. • 1. Projection of Original Data2. Transformation of Data into Fourier Domain3. Filtering of Data4. Transformation of Data Back into Spatial Domain5. Backprojection

  22. Data Projection As a SPECT camera rotates around a patient, it creates a series of planar images called projections.   At each stop, only photons moving perpendicular to the camera face pass through the collimator. As many of these photons originate from various depths in the patient, the result is an overlapping of all tracer emitting organs along the specific path, much in the same manner that an X-ray radiograph is a superposition of all anatomical structures from three dimensions into two dimensions. A SPECT study consists of many planar images acquired at various angles.

  23. A set of projections taken of a patient's bone scan.

  24. After all the projections are acquired, they are subdivided by taking all the projections for a single, thin slice of the patient at a time. All the projections for each slice are then ordered into an image called a sinogram as shown below. It represents the projection of the tracer distribution in the body into a single slice on the camera at every angle of the acquisition.

  25. The aim of the reconstruction process is to retrieve the radiotracer spacial distribution from the projection data as it is illustrated below. This surface rendered image was reconstructed using a fully 3D OSEM algorithm.

  26. 2. Fourier Transform of Data     If the projection sinogram data were reconstructed at this point, artifacts would appear in the reconstructed images due to the nature of the subsequent backprojection operation.  Additionally, due to the random nature of radioactivity, there is an inherent noise in the data that tends to make the reconstructed images rough.  In order to account for both of these effects, it is necessary to filter the data.  When we filter data, we can filter it directly in the projection space, which means that we convolute the data by some sort of smoothing kernel.      Convolution is a computationally intensive task however and so it is useful to avoid using it when possible.  It turns out that the process of convolution in the spatial domain is equivalent to a multiplication in the frequency domain.   This means that any filtering done by the convolution operation in the normal spatial domain can be performed by a simple multiplication when transformed into the frequency domain.  To see what is meant by the spatial domain and the frequency domain consider the following illustrative example.

  27.    Consider a picket fence surrounding Old Lady Fourier's yard.  Since Old Lady Fourier has lived here for a long time and has never looked after her picket fence, it is rather decrepit. At one time, the pickets were all evenly spaced apart and there were exactly 33 pickets over the 10 meter width of her yard when expressed in the spatial domain.  We can express this in the frequency domain however by saying that the picket frequency is 3.3 pickets per m, or 3.3 m-1.     When the fence was new, we can plot graphically in the spatial domain, the number of pickets vs the length in the yard.  The same plot is shown as plotted in the frequency domain.  In the spatial domain, there is one picket spaced every 0.33 m along the fence over the entire 10 m length.  When transformed, we see that there is a large peak at the frequency 3.33 m-1, which corresponds to all the pickets being spaced equally apart.

  28.     As some of the pickets disappear, there is a change in these plots to the ones shown below.  Some of the pickets are missing from the spatial plot, and we see that in the frequency space, there is a second peak emerging at 1 m-1 as now not all the pickets are 0.33 meters apart.  These pickets are now 1.0 meters apart and so the frequency has decreased to 1.0 m-1 for these pickets.     This change in the way the same data is displayed is called a transform.  In SPECT imaging we make a similar transform of the projection data into the frequency space whereby we can more efficiently filter the data.  The transform that we make use of is called the one dimensional Fourier Transform, so named after Old Lady Fourier.

  29. 3. Data Filtering     Once the data has been transformed to the frequency domain, it is then filtered in order to smooth out the statistical noise.  There are many different filters available to filter the data and they all have slightly different characteristics.  For instance, some will smooth very heavily so that there are not any sharp edges, and hence will degrade the final image resolution.  Other filters will maintain a high resolution while only smoothing slightly.  Some typical filters used are the Hanning filter, Butterworth filter, low pass cosine filter, Weiner filter, etc.  Regardless of the filter used, the end result is to display a final image that is relatively free from noise and is pleasing to the eye.  The next figure depicts three objects reconstructed without a filter true (left), without a filter noisy (middle) and with a Hanning filter (right).

  30. 4. Inverse Transform of the Data     As the newly smoothed data is now in the frequency domain, we must transform it back into the spatial domain in order to get out the x,y,z information regarding spatial distribution. This is done in the same type of manner as the original transformation is done, except we use what is called the one dimensional inverse Fourier Transform.  Data at this point is similar to the original (left) sinogram except it is smoothed as seen below (right).

  31. 5. Backprojection     The main reconstruction step involves a process known as backprojection.  As the original data was collected by only allowing photons emitted perpendicular to the camera face to enter the camera, backprojection smears the camera bin data from the filtered sinogram back along the same lines from where the photon was emitted from.  Regions where backprojection lines from different angles intersect represent areas which contain a higher concentration of radiopharmaceutical.

  32. The backprojection process.

  33. SPECT Applications … some examples of the many studies that can be performed with a SPECT camera … 1. Heart Imaging2. Brain Imaging3. Kidney/Renal Imaging4. Bone Scans

  34. 1. Heart Imaging The following figure is a myocardial MIBI scan taken under stress conditions.  Regions of the heart that are not being perfused will display as cooler regions.

  35. 2. Brain Imaging This figure is is a transverse SPECT image of the brain.  Note the hot spots present in the right posterior region.

  36. 3. Kidney/Renal Imaging The following is a renal planar scan using MAG3 tracer (a glucose analog). 

  37. 4. Bone Scans Bone scans are typically performed in order to assess bone growth and to look for bone tumours. The tumors are the dark areas seen in the picture below.

  38. Thank you for your attention

More Related