RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY

IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology. RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. Part 19.07: Optimization of protection in Mammography Practical exercise. Overview.

velvet
Télécharger la présentation

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION INDIAGNOSTIC ANDINTERVENTIONAL RADIOLOGY Part 19.07: Optimization of protection in Mammography Practical exercise

  2. Overview • To be able to apply quality control protocol to mammography equipment • To determine the spatial resolution (lp/mm) both parallel and perpendicular to the tube axis using • This is an evaluation of the whole system (blur due to focal spot combined with the film-screen system) 19.07 : Optimization of protection in Mammography

  3. IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology Part 19.07: Optimization of protection in Mammography Phantom image quality evaluation

  4. Image quality - spatial resolution • The spatial resolution can be measured by imaging two resolution lead bar patterns, (20 line pairs/mm) • These should be placed on top of the PMMA plates (45 mm total thickness) to measure worst-case resolution. 19.07 : Optimization of protection in Mammography

  5. Image quality - spatial resolution • Image the patterns at the reference point • Determine the resolution in both directions using a 6X magnifying loupe. Note: In some cases the three bars may blur but there are smaller bars visible. This is known as false resolution—there will be fewer bars visible in the area with false resolution. The resolution is the smallest group where all bars are just visible. 19.07 : Optimization of protection in Mammography

  6. Image quality - spatial resolution 19.07 : Optimization of protection in Mammography

  7. Image quality - spatial resolution 19.07 : Optimization of protection in Mammography

  8. Image quality phantom 19.07 : Optimization of protection in Mammography

  9. Image quality - spatial resolution 19.07 : Optimization of protection in Mammography

  10. Image quality - spatial resolution • The distance from the chest wall edge is critical, but the position parallel to the thorax side is not critical within ± 5 cm from the reference point. • Resolution is generally lower parallel to the tube axis due to the asymmetrical shape of the focal spot. 19.07 : Optimization of protection in Mammography

  11. Image quality - spatial resolution Limiting value : > 12 lp/mm acceptable > 15 lp/mm desirableat the reference point Frequency :Weekly Equipment : PMMA plates 150 x 240 mm, resolution pattern(s) up to 20 lp/mm, 6X magnifying loupe 19.07 : Optimization of protection in Mammography

  12. Where to Get More Information European protocol for the quality control of the physical and technical aspects of mammography screening. http://euref.org/index.php?option=com_phocadownload&view=category&id=1&Itemid=8 American College of Radiology Mammography Quality Control Manual, Reston VA, 1999. 19.07 : Optimization of protection in Mammography