1 / 10

Problem Management Scenario # 2

Problem Management Scenario # 2. September 17,2005. Scenario # 2. Weekly EM (static readings, at rest) in one of the facility’s BSCs is outside the acceptable range. What do you do?. General Investigational Approach. Determine the severity of the excursion. Evaluate recent trending data.

vesta
Télécharger la présentation

Problem Management Scenario # 2

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Problem Management Scenario # 2 September 17,2005

  2. Scenario # 2 • Weekly EM (static readings, at rest) in one of the facility’s BSCs is outside the acceptable range. • What do you do?

  3. General Investigational Approach • Determine the severity of the excursion. • Evaluate recent trending data. • Initiate investigation: limited or full • Identify products processed in BSC. • Review sterility results for affected products. • Based on investigation results, evaluate potential impact on products, patients & operations.

  4. What should be done if… • Alert level is exceeded for either the particle counts or the viable air sample.

  5. Alert Level Exceeded: Investigation • Limited investigation to include: • Inspect cleanliness of BSC and adjacent areas • Review BSC equipment log (maintenance & cleaning records, calibration date) • Review EM trending data for the BSC • Review sampling device PM & calibration • Identification of organism (if viable air sample in question)

  6. Alert Level Exceeded: Follow-Up • Document the investigation and any corrective/preventative actions. • If problems identified, possible actions include: • BSC Service to verify proper operation – remove from service if fails • Revision of BSC cleaning procedure • Increased monitoring frequency and sites • Staff retraining

  7. Alert Level Exceeded: Follow-Up • Ensure product document includes a cross-reference of the investigation document. • Affected Products: Investigation should determine if stored products are suitable for use. • Product recall unlikely unless BSC malfunction identified.

  8. What should be done if… • Action level is exceeded for either the particle counts or the viable air sample.

  9. Action Level Exceeded: Investigation • Limited or full investigation depending on product, processing, trending data, size of excursion • Same as for Alert Level Investigation • Possible additional actions include: • Verify proper operation of BSC, sampling device • Remove hood from service if excursion serious or BSC fails verification • Thoroughly clean & sanitize BSC • Increase monitoring frequency and sites • Evaluate EM trending for adjacent areas

  10. Action Level Exceeded: Follow-Up • Same as for the Alert Limit Scenario. • Follow Deviation Reporting SOP to notify appropriate medical/ reporting groups • May consider product ‘recall’ based on conclusions of investigation (e.g. organisms identified, nature of BSC malfunction). • Immediately notify FDA district office if recall intended.

More Related