1 / 37

Canta Open Canal Manufacturing Procedure

Learn the step-by-step process of manufacturing the Canta StepVent, including tools needed, measurements, marking, drilling, vent wire insertion, UV material application, finishing, testing, and configuration.

walterhall
Télécharger la présentation

Canta Open Canal Manufacturing Procedure

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Canta Open Canal Manufacturing Procedure Alain S. Balucating Kessa Sem Global Technical Operations March 2004

  2. Manufacturing the Canta StepVent • Tools for Step Venting • Canta 710-OC StepVent Process • Canta 730-OC StepVent Process • StepVent Quality Inspection Criteria

  3. Tools needed for the Step Venting Process Plastic Step-Vent Measuring toolP/N 15178500 Vent Wires(Various Diameters) 6mm burr/mill carbide bit. Use one or the other. (Also known as Pineapple bit) 3.1mm drill bit P/N 12692-031

  4. 710-OC Step Venting Process 2. Using a marker, mark one end of the vent wire, where the top of the shell. • Place vent wire along side of the shell where the venting will be placed.

  5. 15% reduction size Length of vent 710-OC Step Venting Process 4. Place the vent wire against the left side of the plastic vent tool. Use the tool to measure the 15% reduction of the total vent length, shown on the right side of the tool. 3. Using a marker, mark the other end of the vent wire, where the bottom of the shell ends.

  6. 710-OC Step Venting Process 6. Use the same plastic vent tool to mark the 15% reduction location of the vent on the shell at the bottom of the canal. 5. Use a marker and the plastic vent tool to mark the 15% reduction location of the vent on the shell at the top of the canal.

  7. 710-OC Step Venting Process 7. Using the drill bit 3.1mm drill bit, drill vent holes on the side of shell where the 15% reduction was marked.

  8. 710-OC Step Venting Process 9. Insert the appropriate D-shape vent wire. Make the round side of the vent wire leaning on the trench created in the inner wall of the shell. 8. Grind a trench on the inner side of the shells connecting the two holes. Note:Always use the largest possible diameter vent for the best “vent effect” and occlusion reduction for the customer.

  9. 710-OC Step Venting Process 10. Apply UV material in two thin lines on each side of the vent wire. Make sure that the material floats gently around the wire. Cure the material completely in the UV-machine.

  10. 710-OC Step Venting Process 11. Completely cover the remaining part of the vent wire with the appropriate UV venting material. Also add additional UV Venting material on the 15% portion of top and bottom of the shell.Cure it in the UV-machine until completely hard.

  11. 710-OC Step Venting Process 13. Grind off the 15% portion of the shell on the top and bottom of the shell. Buff, round and smooth the edges of the shell . 12. After UV Vent material is cured, slowly remove the vent wire.

  12. 710-OC Step Venting Process 15. Case the shell on a standard Canta 710 faceplate assembly. Follow all standard manufacturing procedures. 14. Drill the receiver hole.

  13. 710-OC Step Venting Process 17. Buff , grind and smooth the faceplate around the shell. Be sure to cover the microphone and receiver openings with tape to protect them from dust and debris. 16. Remove excess faceplate material with a sonic cutter or standard hand tool.

  14. Removal line placement 710-OC Step Venting Process 19. Insert the hearing aid into the positive mold and check to make sure that the vent opening is not blocked on either end. 18. Place Removal line/cord as shown.

  15. Inside of 15% portion of shell Outside of Shell 710-OC Step Venting Process • Important Note! You are removing the outside wall of the hearing instrument shell. The inside wall of the 15% portion of the shell will become the “new” outside wall of the instrument. Be very careful not to cut through the inside wall of the vent, or to remove too much of the faceplate.

  16. 710-OC Step Venting Process Finished 710 OC. Perform all listening and visual inspections. You must now use FONIX II Canta test software to create a Canta Open device!

  17. 710-OC Configuration & Testing • Connect the final CIC instrument to the Fonix / DSA5000 test system • Use FONIX II Canta calibration & test software version 5.1.1 (or higher) • Select “710-OC” as the model type • Run the complete Final Device test program • The ComforTec™ software operating system is now installed, and the instrument will appear as a Canta 710-OC on Aventa Fitting Software v1.4.

  18. 730-OC Step Venting Process • Place vent wire along side of the shell where the venting will be placed. 2. Using a marker, mark one end of the vent wire, where the top of the shell.

  19. 15% reduction size Length of vent 730-OC Step Venting Process 4. Place the vent wire against the left side of the plastic vent tool. Use the tool to measure the 15% reduction of the total vent length, shown on the right side of the tool. 3. Using a marker, mark the other end of the vent wire, where the bottom of the shell ends.

  20. 730-OC Step Venting Process 5. Use a marker and the plastic vent tool to mark the 15% reduction location of the vent on the shell at the top of the canal. 6. Use the same plastic vent tool to mark the 15% reduction location of the vent on the shell at the bottom of the canal.

  21. 730-OC Step Venting Process 7. Using the drill bit 3.1mm drill bit,, drill vent holes on the side of shell where the 15% reduction was marked.

  22. 730-OC Step Venting Process 9. Insert the appropriate D-shape vent wire. Make the round side of the vent wire leaning on the trench created in the inner wall of the shell. 8. Grind a trench on the inner side of the shells connecting the two holes. Note:Always use the largest possible diameter vent for the best “vent effect” and occlusion reduction for the customer.

  23. 730-OC Step Venting Process 10. Apply UV material in two thin lines on each side of the vent wire. Make sure that the material floats gently around the wire. Cure the material completely in the UV-machine.

  24. 730-OC Step Venting Process 11. Completely cover the remaining part of the vent wire with the UV venting material. Also add additional UV Venting material on the 15% portion of top and bottom of the shell. Cure it in the UV-machine until completely hard.

  25. 730-OC Step Venting Process 13. Grind off the 15% portion of the shell on the top and bottom of the shell. Buff, round and smooth the edges of the shell . 12. After UV Vent material is cured, slowly remove the vent wire.

  26. 730-OC Step Venting Process 15. Case the shell on a standard Canta 730 faceplate assembly. Follow all standard manufacturing procedures. 14. Drill the receiver hole.

  27. 730-OC Step Venting Process 16. Remove excess faceplate material with a sonic cutter or hand tool. 17. Buff , grind and smooth the faceplate around the shell. Be sure to cover the microphone and receiver openings with tape to protect them from dust and debris.

  28. 730-OC Step Venting Process 18. Insert the hearing aid into the positive mold and check to make sure that the vent opening is not blocked on either end.

  29. Inside of 15% portion of shell Outside of Shell 730-OC Step Venting Process • Important Note! You are removing the outside wall of the hearing instrument shell. The inside wall of the 15% portion of the shell will become the “new” outside wall of the instrument. Be very careful not to cut through the inside wall of the vent, or to remove too much of the faceplate.

  30. 730-OC Step Venting Process Finished 730 OC Perform all listening and visual inspections. You must now use FONIX II Canta test software to create a Canta Open device!

  31. 730-OC Configuration & Testing • Connect the final ITC instrument to the Fonix / DSA5000 test system • Use FONIX II Canta calibration & test software version 5.1.1 (or higher) • Select “730-OC” as the model type • Run the complete Final Device test program • The ComforTec™ software operating system is now installed, and the instrument will appear as a Canta 730-OC on Aventa Fitting Software v1.4.

  32. Quality Assurance Inspection The “Don’ts” in the Step Venting process

  33. 710-OC “Don’ts” Not properly tapered and stepped. Unacceptable Quality

  34. 710-OC “Don’ts” Vent bottom and faceplate are not rounded off Unacceptable Quality

  35. 730-OC “Don’ts” When reducing the vent, make sure that the edges are rounded off nicely and smoothly and the vent are stepped. Unacceptable Quality

  36. 730-OC “Don’ts” When stepping the vent, make sure that the edges are rounded off nicely and smoothly. Unacceptable Quality

  37. We wish you success with the GN ReSound Canta Open CIC & ITC!

More Related