Summary of the Compton-PET project. V.Chepel. LIP-Coimbra and Department of Physics, University of Coimbra. V. Chepel. Jornadas do LIP, Lisboa, 22-24 April, 2012. Compton-PET stands for: “Feasibility study of using Compton scattering for medical imaging with positrons” (2008-2011).

Télécharger la présentation

Summary of the Compton-PET project

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Summary of the Compton-PET project V.Chepel LIP-Coimbra and Department of Physics, University of Coimbra V. Chepel Jornadas do LIP, Lisboa, 22-24 April, 2012

  2. Compton-PET stands for:“Feasibility study of using Compton scattering for medical imaging with positrons” (2008-2011) V. Chepel Jornadas do LIP, Lisboa, 22-24 April, 2012

  3. Motivation Scattering in the patient body: (~70 to 90%) Scattering in the detector: Probability of the first interaction to be a Compton: BGO – 53% LSO – 62% LXe – 79% LAr – 95% BGO – 78% LSO – 86% LXe – 95.6% LAr – 99.8% of coincidences involve scattered events Using these events for image reconstruction: improves statistics and/or allows reducing the dose administered to patient, but deteriorates resolution V. Chepel Jornadas do LIP, Lisboa, 22-24 April, 2012

  4. (E2, q2) (E4) (E1, q1) (E2, q2) Possible solution: Compton tracing This is what we see (or not) in the detector: Reconstruction of the true sequence depends strongly on resolutions on E and q (i.e. position) p´g pg pg, p´g, pe are in the same plane pe Knowing electron track orientation can help V. Chepel Jornadas do LIP, Lisboa, 22-24 April, 2012

  5. Objective: explore what can be done with the scattered events Compton tracing : • Compton electron track orientation: can we get some useful information on it ? • If one can measure polarization of the g-par – would it help ? • Do we really need to trace (i.e. reconstruct in detail) scattered g-rays ? Method: Monte Carlo and experimental • How Geant deals with the electron binding ? • (see talk by A.Mangiarotti on validation) V. Chepel Jornadas do LIP, Lisboa, 22-24 April, 2012

  6. 1. Attempting to measure electron track orientation in LXe The idea: for 500 keV electrons 5.5 MeV a-s 511 keV g V. Chepel Jornadas do LIP, Lisboa, 22-24 April, 2012

  7. 1. Attempting to measure electron track orientation in LXe 300 ns (expected ~25 ns) Neither field non-uniformity, nor diffusion in the gas explain such long rise time (unless diffusion in Xe vapour is much stronger than in non-saturated gas) LXe a-s, range 40 mm CHERNE2010, Coimbra, 7-9 June, 2010

  8. 2. Polarization of annihilation photons e+ + e-g + g What if we could measure polarization of each g ? Any help with Compton tracing ? Any information about scattering in the patient body ? V. Chepel Jornadas do LIP, Lisboa, 22-24 April, 2012

  9. 2. Compton scattering cross section for polarized photons Eg = 511 keV Maximum anisotropy = 5.5 (for   82º) V. Chepel Jornadas do LIP, Lisboa, 22-24 April, 2012

  10. 2. MC of g-pairs with GEANT4: Point source in vacuum 2 - 1 Point source in of water (20cm phantom) V. Chepel Jornadas do LIP, Lisboa, 22-24 April, 2012

  11. 3. Do we really need to do tracing ? Computation tool: Interactions (physics) Detector response Compton tracing and LOR definition Image reconstruction GEANT4 C++ Matlab CHERNE2010, Coimbra, 7-9 June, 2010

  12. 3. Choosing LOR: a “childish” approach CHERNE2010, Coimbra, 7-9 June, 2010

  13. 3. UTAH phantom simulated P r e l i m i n a r y Traditional PET: Single energy deposition in 350-650 keV Compton PET: Singles + doubles With total E= 350-650 keV Same contrast Statistics > x3 Detectors: LXe ring 80x5x5 cm3 Resolution 2 mm, dE/E=20% CHERNE2010, Coimbra, 7-9 June, 2010

  14. 3. Line source 4.5 cm off centre in 20 cm of water P r e l i m i n a r y Traditional PET: Scatter fraction  0.7 Compton PET: Scatter fraction  0.86 Detectors: LXe ring 80x5x5 cm3 Resolution 2 mm, dE/E=20% CHERNE2010, Coimbra, 7-9 June, 2010

  15. Conclusions • Compton electron tracking does not seem to be feasible for <=511 keV (at least with the readout we have tried). • The effect of gamma-ray polarization on the observable distributions is small. • Events involving Compton scattering in the detector can be recovered for the imaging with the advantage of higher image statistics and without jeopardizing the image quality. • Complex Compton tracing does not to be necessary. Simple methods can be used for that. CHERNE2010, Coimbra, 7-9 June, 2010