An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Central placement of screw fixation in scaphoid fracture—acomputational approach and biomechanical study Yang Guo MD, FHKCOS Associate Consultant Department of Hand Surgery, JiShui Tan hospital, beijing, China

  2. OPTIMAL position of fixation for scaphoid fractures continues to be debated • Central placement: supported by(Tremble,Mccallister,2000) biomechanical studies /clinical evaluations Has been accepted as a prefered method of fixation • Recently, some studies :(Luria 2010,2012) biomechanical studies /finite element analysis suggested no difference in stability between central and eccentric placement The authors did not receive any outside funding or grants in support of their research for or preparation of this work. Neither they nor a member of their immediate families received payments or other benefits.

  3. To be central or Not to be?Our study consisted of 2 parts • Establishment “central” concept computational approach • Biomechanical test central Vs peripheral

  4. 1Establishing “Central”

  5. Shrinkage in a uniform manner in software-- pre-op planning system Scaphoid surface model was extracted. One pixel space was allowed by one time of shrinkage in the program Scaphoid model: smaller and smaller

  6. Central one-third zone and central zone Scahpoid distal pole Further shrinkage of 1.5mm Central zone Central one third Waist portion Scahpoidproximal pole

  7. Central zone-geometry Screw axis

  8. Part 2: Biomechanical Study Design:Types of fixation

  9. Biomechanical study: osteotomy design

  10. Screw axis Group 1 Group 2 thread Group 4 Group 3 Fracture plane

  11. Materials and Test machine

  12. Estimation of sample size and power 10 of scaphoids,consistent with a previous data • Data analysis: Wilcoxon nonparametric analysis Failure load of the specimens (P <0.05). • The mode of failure :Screw migration fracture at the screw–bone interface, sliding along the fracture. • None of the specimens dislodged from the potting fixture • None of screw was bent

  13. Stroke-Force curve

  14. Results

  15. Brief Disccussion • 1. Concept of Central need to be clarified Straight central axis does not exist • 2. Central placement : important as long as possible

  16. Limitations • Samewith any cadaveric biomechanical study • The screw itself may contributed resistance ,when it was placed in a more horizontal direction • Saw bone material: Failure load lower than previous tests • We simplified the forces into a single vector and did not perform cyclic loading • Fx plane differed slightly, which affects its stability.

  17. Thanks!