An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Dr.Wahid Helmypediatric consultant. Basics of Mechanical Ventilation in Neonates

  2. Ventilator care requires a team effort. Everyone involved has to get along and trust one another!

  3. Prevention of alveolar collapse ◘ Functional residual capacty (FRC). ◘ Surfactant . ◘Elatic-recoil ( compliance). ◘ Intrapleural pressure(-4mmHg) during inspiration and (+4mmHg) during inspiration . ◘ If surfactant is absent , Intrapleural pressure negativity may be increased up to (-20mmHg).

  4. What is it?

  5. Pulmonary Mechanics

  6. 1)Tidal Volume (Vt) ◘ (Vt) = 6-10 mL/kg/Breath. ◘RR is usually 30-60 BPM. 2) minute volume = (Vt- Dead space)x RR. ↑ (PIP)→↑Tidal Volume →↑minute volume .

  7. 3) Compliance = 0.004 L/cmH2O. = Change in volume (mL) = 0.004 L/cmH2O. Change in pressure(cmH2O) 4)Resistance = 30cm H2O/L/sec Change in pressure (cmH2O)= 30cm H2O/L/sec Change in flow (L/sec) NB., Resistance X Compliance = 1Time constant 1Time constant =0.004 L/cmH2O. X 30cm H2O/L/sec =12

  8. 4)one Time constant = Resistance X Compliance ◘ one time constant → 63% equilibration of pressure inside & outside the alveoli. ◘ we need 3 time constant →97% equilibration of pressure inside & outside the alveoli. • If resistance =30cm H2O/L/sec • compliance = 0.004 L/cmH2O. • One time constant =30 X 0.004 = 0.12 seconds. • We need 3time constant to inflate and deflate the lung (3 X 0.12 seconds = 0.36 seconds=Ti ). • as aresult Te= 2 or 3 X 0.36seconds. • So I/E ratio = 1:2 or 1:3 .

  9. Types of Mechanical Ventilators

  10. Types of Mechanical Ventilators • Volumev- cycled ventilators.لمجرد المعرفة • Pressure ventilators . لمجرد المعرفة • Pressure-limited, time-cycled, continuous-flow ventilators .هام جدا • Patient–triggered ventilators (PTV).هام جدا

  11. What is it?

  12. Pressure-limited, time-cycled, continuous-flow ventilators Ventilators • You select (PIP)→ (pressure-limited). • You select inspiratory time → (time-cycled). • (Continuous flow) →Fresh heated humidified gas is delivered to the patient throughout the respiratory cycle.

  13. Parameters of mechanical ventilation

  14. (PIP) minus(PEEP) • (PIP) → The maximum pressure reached during inspiration. If PIP is too low → low VT. If PIP too high → high VT → Hyperinflation and air leak → Impedance مقاومةof venous return. • (Optimum (PEEP) is 4-6 cmH2O). • High PEEP >8 cmH2O .,→ –Reduces gradient between PIP & PEEP→ (↓ VT) . –Decreases venous return . –Increases pulmonary air leaks . –Produces CO2 retention .

  15. (FiO2) • why Increase in FiO2 improves oxygenation ? ↑ oxygen tension inside the alveoli→ ↑ r diffusion gradient → good oxygenation. • Why Oxygen and Paw balance is essentiaL ? to minimize lung damage. • Why Paw should be ↓ before a very low FiO2 is reached During weaning. to avoid a high incidence of air leak is observed.

  16. RR, secrets • ↑ RR → ↑ (CO2 wash). • RR(60 BPM) allows for PIP reduction in PIP → ↓ incidence of pneumothorax with about 50% . • Most neonates have short time constants so they can tolerate (RR60-70 Bpm) and short (Te) without marked gas trapping . • RR Determinesيحدد minute ventilation(RR×VT),thus CO2 elimination.

  17. Minute alveolar ventilation Minute alveolar ventilation = (Tidal volume – Dead space) X Frequency. • Tidal volume,is determined mainly with pressure gradient between inspiration and expiration i.e. (PIP) minus (PEEP).

  18. Ti and Te ●(Ti)is .3 - . 5 seconds for LBW and .5 - .6 seconds for larger infants ●Depends on the pulmonary mechanics: – Compliance . – Resistance . –Time constant. I:E ratio ● It should NOT be reversed ● I:E ratio should NOT be less than 1:1.2

  19. mean airway pressure • MAP + FiO2 → determines oxygenation.why? • An ↑ in PIP and PEEP→ ↑ MAP → ↑ oxygenation more than ↑ in the I:E ratio. • NB., ↑↑↑ Paw →alveolar over distension with right to left shunt. Flow Flow rates of 6-10 liter/min are usually sufficient.

  20. Modes of venilation Who is theCommander?

  21. A)Non-triggeredModes. 1.Controlled Mandatory Ventilation (CMV) or IPPV: – IPPV (intermittent positive pressure ventilation ). –Ventilator rate is set > infant's spontaneous. – RR (usually 50-80 breaths/min). 2.Intermittent Mandatory Ventilation (IMV): –Ventilator rate is set < infant's spontaneous breaths. – RR (<30 breaths/min). – spontaneous breaths above the set rate are not assisted.

  22. B) Patient–Triggered Ventilators (PTV) • Modification of conventional ventilation ( IMV or IPPV) by adding synchorinization (S). • ASensor detect the Inspiratory efforts of the baby by so triggering ( the ventilator setting. • the patient is able to initiate (trigger) ventilator breaths.

  23. PTV is used in two modes • Assist Control Mode (A/C) or sippv • All breath initiated by patient is triggered= Assist. • Back up rate = ippv = ControL MV. • If apnea occur at any time baby will be ventilated. • Synchronized Intermttent Mandatory Ventilation (SIMV): • Preset rate that is triggered, • other patient breath is not assisted.

  24. Indications of Mechanical Ventilation • hypoxemia→ with PaO2 less than 50 mmHg despite FiO2 of 0.8. • Respiratory acidosis → pH of less than 7.20 to 7.25, or PaCO2 above 60 mmHg. • Severe prolonged apnea. • Frequent intermittent apnea unresponsive to drug therapy. • Relieving work of breathing in an infant with signs of respiratory difficulty.

  25. Blood Gases Changesby Ventilator Setting

  26. ET Size

  27. Initial Setting of Mechanical Ventilation

  28. Monitoring The Infant during Mechanical Ventilation • (ABG)) ., • Obtain a blood gas within 15-30 minutes of any change in ventilator settings. • Obtain a blood gas every 6 hours unless a sudden change in the infant's condition occurs. • Continuous monitoring of the O2 saturation level as well as the HR and RR is necessary.

  29. Paralysis and Sedation • It is not routinely indicated. • It may be used in irritable infants with their spontaneous respiration is out of phase with the ventilator( as in modes with preset rates as in ippv and imv) . • in infants with RDS→ ↓dynamic lung compliance →↑ airway resistance, the removal of the infant’s respiratory effort contribution to tidal breathing. • after initiation of neuromuscular blockadeit is necessary to increase ventilator pressure

  30. Weaning • Parameters gradually decreased (PIP 2 cm H2O, FiO2 5%, Rate 5 BPM). • 1. Reduce FiO2 to 80% before changing PIP, I:E or PEEP. • 2. Reduce PIP as clinically indicated. • 3. Reduce FiO2 to less than 60% • 4. Reduce inspiratory time. • 5. Reduce PIP to 10-14 cm H2O (Larger babies may be extubated with PIP 14-18) • 6. Reduce rate to 20 -40 /BPM then Te should be prolonged.

  31. Weaning (cont.) • 7.preterm infants → Use of nasal CPAP → to avoid atelectasis. • 8.prolonged intubation or previous failure of extubation → a short course of steroids may facilitate extubation. • 9.If stridor caused by laryngeal edema develops after extubation, →nebulization with adrenaline.