Stretching machine for biomedical research

Stretching machine for biomedical research. Mark Bradford Kevin Feeley Tony Martinelli Jeff Snyder Jacob Stephens. Background. Our sponsor sought a device capable of recreating in vivo conditions on cell cultures.

liesel
Télécharger la présentation

Stretching machine for biomedical research

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Stretching machine for biomedical research Mark Bradford Kevin Feeley Tony Martinelli Jeff Snyder Jacob Stephens

  2. Background • Our sponsor sought a device capable of recreating in vivo conditions on cell cultures. • This device applies the mechanical stresses and strains that a particular type of cell culture would encounter within the body. • There is a device available commercially for this purpose; however, it is prohibitively expensive. • The purpose of this project was to design an alternative that is less expensive, thus enabling more researchers access to this type of machine.

  3. Customer Requirements • Must be considerably less expensive to manufacture • Must use standard well plates • Must be computer controlled • Must fit in an incubator • Must tolerate humidity up to 100% • Must be easy to use • Must allow easy access to well plate • Must measure forces applied to cultures • Must measure displacement applied to cultures • Must have high repeatability

  4. Initial Concepts • Several concepts were brainstormed • All concepts accomplish same functions but with different components • Those components included the clipping mechanism, driver, and type of well plate Figure 1: Clipping Mechanism Concept

  5. Initial Concepts • One of these concepts is sketched below Figure 2: Concept 2

  6. Final Design, Clips • The clips secure the membrane to the stretch plate • The version shown below was determined to be the most user-friendly, while accomplishing their primary function Figure 3: Solid Model of the Clip

  7. Final Design, Stretch Plate • The stretch plate transfers force from the driver to the membranes • The final version is designed for standard 8-well plates Figure 4: Solid Model of Stretch Plates

  8. Final Design, Base • The base constrains the stretch plate and holds the well plate in place • It was also designed to allow easy removal of the well plate Figure 5: Base with mounting sub-assembly

  9. Final Design, Driver • The driver in the final design was picked to be a linear actuator • The driver provides force to stretch the membrane Figure 6: RRA-23 Linear Actuator

  10. Final Design, Sensors • Two sensors were needed, one that measured force and the other to measure displacement • The force sensor was a Load Cell strain gage, and the displacement sensor was a Baumer Inductive sensor Figure 7: Strain Gage Figure 8: Baumer Inductive Sensor

  11. Final Design, Software • The machine needs to be computer controlled, and use software that is easy to use • National Instruments’ LabVIEW was chosen

  12. Evaluation Concepts Final Design • Clips • Used staples or screwed-down clamp • Well Plates • Used 6-well or 8-well plates • Driver • Used solenoid or piezoelectric motors • Base • Requires removal of stretch plate to access well plate • Clips • Uses ridged clamps • Well Plates • Uses 8-well plates • Driver • Uses electric linear actuator • Base • Well plate can be removed without touching stretch plate

  13. Recommendations • A prototype was produced, but a production run would use different materials • Of particular interest are the polymers whose resins are commercially available through Solvay. They provide the necessary mechanical properties and environmental stability needed for this product.

  14. Recommendations • The displacement sensor may need to be replaced with a more accurate linear laser sensor • The motor might interfere with the force sensor’s signal and may need replacement

  15. Acknowledgements • We would like to thank Cook Biotech, Dr. Omar El-Mounayri, Dr. Hazim El-Mounayri, and Mr. Rudy Earlson.