100 tons of LAr (Liquid Argon) Detectors T2K (Tokai 2 Kamiokande) (2km from accelerator)

R&D for liquid Argon TPC detectors for Neutrino and astroparticle experiments. F ~ 70 m. X00km. 295km. 2km. 100 ktons Liquid Argon TPC (Time Projection Chamber) for Astro particle Physics T2K second phase CP violation? LAGUNA: proton decay. 100 tons of LAr (Liquid Argon)

oriole
Télécharger la présentation

100 tons of LAr (Liquid Argon) Detectors T2K (Tokai 2 Kamiokande) (2km from accelerator)

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. R&D for liquid Argon TPC detectors for Neutrino and astroparticle experiments F ~ 70 m X00km 295km 2km 100 ktons Liquid Argon TPC (Time Projection Chamber) for Astro particle Physics T2K second phase CP violation? LAGUNA: proton decay 100 tons of LAr (Liquid Argon) Detectors T2K (Tokai 2 Kamiokande) (2km from accelerator) 2020 ? T2K phase II, LAGUNA, 100 Ktons LAr, 100 000 channels or more

  2. Electronic choices due to the detector principle ne QE • Detection with a system of wires of primary ionization drifted in ultra-pure LAr by induction or collection, at minimum ~ 18000 electrons per channel • No charge amplification at the level of the detector  low noise preamp • 400 ns sampling to reconstruct the drift time and the amount of charge • 3D image reconstruction with several planes of wires + drift time • Multichannel 3fC to 120fC (0.5μs pulse) pre-amplifier • 1000 e- ENC with 250pF Detector capacitance • Able to work in LAr vapours @ -150°C with very few power dissipation

  3. Test setup in Liquid Nitrogen N2 @ 77K Qin Vin Vout Qnoise en Vnoise en Cinj CD Cpa(A+1) externalshaper Pre-Amplifier Detector H -A CD 250pF Cpa Rpa τ= [20ns - 1s] -A • 8100μ/0.35μ input transistor T1.4 supplied by 10 mA • In-between phase margin compensation Equivalence without Rpa

  4. Inside PATOP test chip 350Ω : compensation resistance 5pF : compensation capacitance 1/81 pattern of the input trans. Half of the cascode trans.

  5. Two Test chip versions • TOPEST expected for mid-july 2008 • PATOP received in October 2007 AMS 0.35µ PA shaper buffer biasing circuit

  6. Test setup in Liquid Nitrogen N2 @ 77K

  7. Typical 3fC input charge signal externalshaper Pre-Amplifier H -A CD 250pF Cpa Rpa τ= [5ns - 1s] ,q=1.6 10-19 Vnoise Vpeak

  8. Performances for various temperatures externalshaper Qin Vin Vout Qnoise en Vnoise H en Cinj CD Cpa(A+1) τ= [5ns - 1s] -A Shaping time [5ns-1000ns] Vpeak Vnoise ENC 2MΩ 20MΩ - Ambient ~30°C - Vapour~ -110°C - LN2 : 77K

  9. Powered with 3V battery supply Vpeak Vnoise ENC No more noise (sdev) increasing for 20MΩ channel, its shows a lack of power supply decoupling. - Ambient ~30°C - LN2 : 77K

  10. Noise performance in temperature • The noise target that we expect is 1000 e-, considering an input capacitance of 250pF • Due to decoupling weakness, the ENC is not as low as expected. • Theoretical vs. Experimental • ENC table for various temperature with or without shaper for 2MΩ and 20MΩ channels

  11. Resistance for various temperature Measured Using a QuadTech 7600 RLC meter Rpolyhc 2MΩ ambiant to -196°C Rpoly2 350Ω ambiant to -196°C

  12. Threshold voltage NMOS Wtot=199.8μ, L=0.7μ, ngate=9 PMOS Wtot=100μ, L=0.35μ, ngate=10 Vth

  13. members • LAr R&D electronics IPNL for T2K and Astroparticle Physics : • D. Autiero, E. Bechetoille, B.Carlus, S. Gardien, C.Girerd, J. Marteau, H.Mathez • In collaboration with ETHZ and Bern groups Thank you !

  14. Appendix

  15. SMD Capacitances mounted on PCB Power supply Decoupling : 10μF X5R oxyde Substitute detector capacitance : 250pF COG oxyde +30°C; -110°C; -194 +30°C -194

  16. Quadtech RLC measures of SMD tested in LN2

  17. Dynamic measures Measures last ⋍ 2 minutes PMOS test from 0 to 3.3 => starts with high current • NMOS test from 0 to 3.3 • starts with small currents • Progressive warming

  18. ENC as a function of Cdet ENC = 620.4 + 7.5 * Cdet; 20M_30_0uENC = 510.4 + 6.2 * Cdet; 20M_-100_0uENC = 438.6 + 6.9 * Cdet; 20M_-200_0u ENC = 257.1 + 7.2 * Cdet; 20M_30_1uENC = 255.5 + 5.1 * Cdet; 20M_-100_1uENC = 252.1 + 5.1 * Cdet; 20M_-200_1u ENC = 262.0 + 14.8 * Cdet; 2M_30_0uENC = 513.2 + 8.3 * Cdet; 2M_-100_0uENC = 512.8 + 9.1 * Cdet; 2M_-200_0u ENC = 426.1 + 5.0 * Cdet; 2M_30_1uENC = 318.2 + 3.9 * Cdet; 2M_-100_1uENC = 248.2 + 4.4 * Cdet; 2M_-200_1u