An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Brazilian Tunable Filter Imager(BTFI)Preliminary Design Review (PDR)‏ iBTF module Prototyping Version 1.0 USP-IAG Universidade de São Paulo 18-19th June 2008

  2. iBTF module (as built) z-stage VPHG1 • VPHG2 with • 40mm D-Gs • 2-by-2 mosaic

  3. iBTF – 3 axes motors (as built) • VPHG1 • Rotation only • VPHG1 • Rotation • Translation

  4. iBTF Grating mount(as built) D-G 2-by-2 mosaic (Ondax) on substrate Grating holder (individual) Grating mount (adjustable in roll)

  5. It works!Thanks to Rene, Reitano (INPE); Giseli, Fernando (USP);Cesar & Ligia de Oliveira (LNA)

  6. Close up view

  7. Control electronics bird’s nest

  8. Fundamentals • In order to effect dispersion cancellation at all angles of the VPH grating (VPHG) pair (between 25° and 45°): • The grating rulings of the VPHGs are required to be parallel; • ~1’ or ~30m at edge • The surfaces of the VHG gratings (VPHGs) are required to be parallel; • RX ~3” ; TX ~1’ – constrained not by counter-dispersion requirement but to avoid image motion with scan • Furthermore, in order to scan the iBTF in wavelength: • The image transmitted by the VPHG pair should remain stationary during the wavelength scan; • 1-pixel ~10” – but more relaxed with flexure monitoring

  9. Coordination • Motor identification: • M1 – Fixed rotation stage • M2 – Moving rotation stage • M3 – Linear translation stage on which M2 is moved • Fundamental axes of the iBTFP: • Z-axis: defined by the linear translation stage on which M2 is mounted • X-axis: defined by the rotation axis of M1 • Y-axis: experimentally defined by the gravitational vector in the plane of rotation of M1

  10. iBTFP geometry • Input optical axis (O-axis) • Translation stage (Z-axis – M3) • O&Z-axes should be coincident • Rotation axis of M1 (X-axis) • X-axis & O/Z-axis should be orthorgonal • Rotation axis of M2 should be parallel to M1 • Y-axis defined by gravity vector (arbitrary) • Grating surfaces (VPHG1&2) should be coincident with X-axes of M1&M2 • Grating rulings of VPHG1&2 should be parallel • Requires rotation @ y-axis … to what tolerance?

  11. Alignment Procedure • Mounting of iBTFP and laser on optical bench • Laser directed along z-axis; y-axis defined by gravity vector • Zero-point and rotation direction of M1 and M2 • Defined using software control of M1 and M2 • Establishing Y-axis alignment with the gravitational vector • Accurate spirit level • Establishing parallelism between rotation axes of M1 and M2 • Rotation @ y-axis unavailable; require shims on motor mount • Establishing laser alignment with the Z-axis • Use pin-hole and z-translation stage (M3) • Establishing orthogonality of M2 mirror surface with Z-axis • No rotation @ y-axis – used mirror on 2D mount • Establishing orthogonality of M1 mirror surface with Z-axis • Again no y-axis adjustment on M1 • Establishing orthogonality of the Z-axis with the X-Y plane • Uncritical (tbd)

  12. Still to be done • Final alignment using: • Laser, 2D-adjustable mirrors, shims and spirit-level • Install VPHGs • from Ondax (Rx ; doped-glass) • Check dispersion cancellation • Check stationarity of image during scan

  13. Preliminary conclusions • Basic z-configuration design works? • Electro-mechanically – Yes • Optically • Reflection tests – Yes • Dispersion tests – Probably (tbc) • What needs to be changed? • Y-axis adjustment of M2 • Shims will be adequate • Y-axis adjustment of grating mount • Design modification in hand • Fundamental conclusion • iBTFP = iBTF module in BTFI