1 / 7

Comparison of the Diaton Transpalpebral Tonometer Versus Goldmann Applanation

Comparison of the Diaton Transpalpebral Tonometer Versus Goldmann Applanation

adamdaniel
Télécharger la présentation

Comparison of the Diaton Transpalpebral Tonometer Versus Goldmann Applanation

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Comparison of the Diaton Transpalpebral Tonometer Versus Goldmann Applanation R. S. Davidson1; N. Faberowski2; R. J. Noecker3; M. Y. Kahook11. Ophthalmology, Rocky Mountain Lions Eye Institute, Aurora, CO, USA. 2. Ophthalmology, Denver Health Medical Center, Denver, CO, USA. 3. Ophthalmology, UPMC, Pittsburgh, PA, USA.

  2. Financial Disclosure The authors have no financial interest in the subject matter being presented

  3. Background Diaton tonometry is a unique approach to measuring intraocular pressure (IOP) through the Eyelid. It is a non-contact (no contact with cornea), pen like, hand-held, portable tonometer. It requires no anesthesia or sterilization.

  4. Purpose To investigate the agreement in the measurement of intraocular pressure (IOP) obtained by transpalpebral tonometry using the Diaton tonometer versus Goldmann applanation in adult patients presenting for routine eye exams.

  5. Methods Retrospective chart review of consecutive IOP measurements performed on 64 eyes of 32 patients age 34-91 years with both the Diaton tonometer and Goldmann applanation. Results between groups were examined using analysis of variance (ANOVA) where appropriate.

  6. Results Mean IOP was 15.09 +/-4.31 mm Hg in the Goldmann group and 15.70 +/-4.33 mm Hg in the Diaton group (p=0.43). Mean IOP variation between groups was 1.74 +/-1.42 mm Hg (range 0-8). 83% of all measurements were within 2 mm Hg of each other.

  7. Conclusions The transpalpebral method of measuring IOP with the Diaton tonometer correlates well with Goldmann applanation. Diaton applanation may be a clinically useful device for measuring IOP in routine eye exams.

More Related