An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Cryodet - March 2006 The ZEPLIN III Detector Results from surface calibrations(Coimbra, Edinburgh, ICL, ITEP, RAL, Rochester, Texas, UCLA)

  2. Cryodet - March 2006 Simultaneous Measurement of Scintillation and Ionisation

  3. Cryodet - March 2006 ZEPLIN III Features • 8kg fiducial mass • 31 PMTs in liquid • 3.5 cm drift depth • 0.5 cm electroluminescent gap • reverse field region • position sensitivity • open plan – no surfaces - reduced feedback

  4. Cryodet - March 2006 Simulations • Light collection • liquid phase • gas phase • Position sensitivity • gas electroluminescence • Internal gamma/neutron backgrounds • Fully implemented GEANT4 model • Test data set production

  5. Cryodet - March 2006

  6. year 2 year 3 Araujo et al., 2006

  7. Cryodet – March 2006 ZEPLIN III is built CAPAC - May 2001

  8. Cryodet - March 2006

  9. Cryodet - March 2006

  10. Cryodet - March 2006 Three cool-downs so far • Check-out of thermal control system, xenon transfer and PMT response using internal radioactive sources. • Thermal design verified – 1litre/hour LN boil off! • All PMTs operational • Xenon transferred several times during 6 week run • Am241 spectra obtained with up to 17pe/keV! • Dual-range 62 channel DAQ readout • High precision waveforms collected (including neutron source)

  11. Cryodet - March 2006 Three cool-downs so far • First two-phase run (after removal of radioactive sources) • Level setting with capacitive sensors • HV established • Two-phase signal recovery from full depth • HV breakdown seen • Xenon leak into vacuum jacket (☺mass spec.) • Xenon recovery from vacuum jacket – no loss! Open system, located fractured HV feedthrough, replaced, closed up again

  12. Cryodet - March 2006 Three cool-downs so far • Second two-phase run • HV established (~4kV/cm (at the moment) – liquid – sustained for several days without trip) • Two-phase signal recovery • Event classification started • 57Co spectrum from external source – as expected – working on light yield • Purification cycles started – to recover from recent opening of detector

  13. First cool-down 11-14% FWHM • Single phase (no liquid/gas interface) 7-11 pe/keV • Dual phase (with liquid/gas interface) ~17pe/keV • MC simulations imply Xe yield probably > 60photons/keV • Linear to better than 10% - 8 to 60 keV

  14. Second cool-down

  15. Third cool-down Primary scintillation

  16. Third cool-down

  17. Cryodet - March 2006 Summary • ZEPLIN II is operating in Boulby mine. • Full depth operation • Final in-situ calibrations ongoing • ZEPLIN III is up and running. • Performance so far is as expected. • Some indication of higher photon yield than expected in xenon. • Further surface evaluation and optimisation in progress – another 3 months • FP 7 Design Study proposal for ton scale – ELIXIR – LoI submitted to ILIAS/ApPEC • FP7 Single Project proposal for ZEPLIN III deployment - LoI submitted to ILIAS/ApPEC