Radon purification facility, effect inside NEMO3

Radon purification facility, effect inside NEMO3. Radon purification facility. Running since Oct. 4th, 2004 in FréjusUnderground Lab. 1 ton charcoal @ -50 o C, 7 bars Flux: 150 m 3 /h Activity of 222 Rn : Before Facility = 15 Bq/m 3 After Facility < 15 mBq/m 3.

Télécharger la présentation

Radon purification facility, effect inside NEMO3

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Radon purification facility, effect inside NEMO3

  2. Radon purification facility Running since Oct. 4th, 2004 in FréjusUnderground Lab. 1 ton charcoal @ -50oC, 7 bars Flux: 150 m3/h Activity of 222Rn : Before Facility = 15 Bq/m3 After Facility < 15 mBq/m3

  3. Now : A(222Rn) in the LSM ~10 Bq/m3 Fall 2003 : A(222Rn) ~5 Bq/m3 Mid 2004 : A(222Rn) ~0.2 Bq/m3 Anti-radon tent for NEMO3 detector S E W N

  4. (all operations involving crystals and PMTs -including photos- in HP N2 atmosphere) detectors during installation; in the central and right up detectors the new shaped Cu shield surrounding light guides (acting also as optical windows) and PMTs was not yet applied installing DAMA/LIBRA detectors assembling a DAMA/ LIBRA detector filling the inner Cu box with further shield view at end of detectors’ installation in the Cu box closing the Cu box housing the detectors