1 / 59

Tri-Tech Medical Inc.

Tri-Tech Medical Inc. Medical Gas Outlets. Medical Gas Outlets. DISS Ohmeda style Allied style Puritan style Recessed wall DISS Recessed ceiling Surface wall Console / Headwall Ambulance / Surgical Arm. Common Backs. Recessed/Console 360º swivel inlet tube 1/2” OD copper tubing

anisa
Télécharger la présentation

Tri-Tech Medical Inc.

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Tri-Tech Medical Inc. Medical Gas Outlets

  2. Medical Gas Outlets • DISS • Ohmeda style • Allied style • Puritan style • Recessed wall • DISS Recessed ceiling • Surface wall • Console / Headwall • Ambulance / Surgical Arm

  3. Common Backs • Recessed/Console • 360º swivel inlet tube • 1/2” OD copper tubing • Accommodates all styles of Outlet Latch Plates • ETL listed • Ambulance/ Surgical Arm • 360º swivel inlet tube • DISS or hose barb • Accommodates all styles of Outlet Latch Plates • ETL listed

  4. DISS Outlets • Frontall™ design • Gangable • Full faceplate color coding • 5 inch centers • Primary & secondary checks • 360° positioning of extension tube

  5. DISSSurface Outlets • Frontall™ design • Full faceplate color coding • Primary & secondary checks

  6. Frontall™ Feature • Allows routine service of both the primary and secondary check valves without removing the nameplate or gas specific portions. • Very fast & easy to service • Primary check cartridge

  7. Repairing DISS Front Bodies • These parts would be ordered separately • Cartridge (Negative and Positive gases) • Primary Check Seat and Ball Check (Positive gases only) • Chrome-Plated Body

  8. Ohmeda Style Outlets

  9. Ohmeda Style Outlets • Accepts Ohmeda gas specific compatible male adaptors • Gangable • 5” centers • Allows routine service of both the primary and secondary check valves. • Full faceplate color coding

  10. Ohmeda Style Surface Outlet • Accepts Ohmeda gas specific compatible male adaptors • Allows routine service of both the primary and secondary check valves. • Full faceplate color coding

  11. Repairing Ohmeda style fronts • Remove screws • Remove body and plate

  12. U-clip and washer remain. • Front body and plate

  13. Replace body • Replace plate and screw

  14. XQ1000-D Ohmeda style Front Latch Body • Can be ordered as an assembly, or in parts

  15. Allied Style Outlets

  16. Allied Style Outlets • Accepts Allied style gas specific compatible male adapters • Gangable, 5” centers • Full faceplate color coding • 1/2” OD copper tubing • Metal back and front plates • 360º swivel inlet tube

  17. Allied Style Surface Outlets • Accepts Allied gas specific compatible male adapters • Full faceplate color coding

  18. Repairing an Allied Style Front • Remove screws • Remove body and plate

  19. Position o-ring and washer on new body • Place body on housing and replace plate

  20. Replace Screws

  21. XA1000-D Front Latch Body • Can be ordered as an assembly, or in parts

  22. Puritan Style Outlets

  23. Puritan Style Outlets • Accepts Puritan style gas specific compatible male adapters • Gangable, 5” centers • Full faceplate color coding • 1/2” OD copper tubing • Metal back and front plates • 360º swivel inlet tube

  24. Puritan Style Outlets • Accepts Puritan gas specific compatible male adapters • Full faceplate color coding

  25. Repairing a Puritan Style Front • Using manual screwdriver, remove screws • Remove body and small plate

  26. Remove ONLY these parts

  27. Replace parts or entire body • Screw plate back in place with Manual screwdriver. • Do not over tighten

  28. XP1000-E Puritan Latch Body • Can be ordered as an assembly or in parts

  29. Console / Headwall Outlets • Available in DISS, Ohmeda, Puritan and Allied styles • DISS style includes Frontall™ features • Full faceplate color coding • Optional trim plate available

  30. Console / Headwall Outlets

  31. Console / Headwall Outlets

  32. Ambulance / Surgical Arm Outlets • Available in DISS and Quick Connect styles • DISS and hose barb inlet connections • DISS includes Frontall™ features • Full faceplate color coding

  33. DISS x ¼” npt outlets and DISS x hose barb outlets

  34. Fairfield Rail

  35. Fairfield Rail Outlets • Three different styles

  36. Outlet Conversion Kits

  37. Conversion Kits

  38. Conversion Kit - Disassembly

  39. Conversion Kits - Disassembly

  40. Conversion Kits - Installation

  41. Trim Plates • Single gang through six gang • Large array of on center spacing • Can be customized to meet your needs

  42. Ohmeda to Ohmeda Conversion Kits • Available for Diamond I to Diamond III • Available for Diamond II to Diamond III • Easy to install • Easy to service

  43. Other Ohmeda Conversion Kits • Available for Melco/Foregger/Linde 9/16-18 and Chemetron/NCG to Diamond III • Easy to install and service

  44. NFPA 99 REQUIREMENTS FOR • STATION OUTLETS/INLETSTRI-TECH PRODUCT • GAS SPECIFIC YES • STATION OUTLET TO HAVE PRIMARY AND SECONDARY CHECK YES • STATION INLET TO HAVE PRIMARY CHECK (SECONDARY CHECK PERMITTED) YES • SECONDARY VALVE TO CLOSE WHEN PRIMARY VALVE IS REMOVED YES • EACH OUTLET/INLET LEGIBLY IDENTIFIED ACCORDING TO 5.1.11.3 YES • THREADED OUTLETS/INLETS SHALL COMPLY WITH CGA V-5 (DISS) YES • STATION OUTLET/INLET COMPONENTS THAT ARE GAS SPECIFIC • NOT INTERCHANGEABLE YES • COPPER TUBES ON STATION OUTLETS SHALL BE 3/8 OD MIN • AND LESS THAN 8 INCHES LONG YES • COPPER TUBES ON STATION INLETS SHALL BE ½ OD MIN • AND LESS THAN 8 INCHES LONG YES • ADDITIONAL TESTING PERFORMED BY TRI-TECH • INITIAL PRESSURE TEST AND FINAL TIE IN TEST REQUIRED FOR LEAKAGE • AS WELL AS THE OPERATIONAL PRESSURE TEST REQUIRED FOR FLOW • IS PERFORMED ON ALL STATION OUTLETS/INLETS PRIOR TO SHIPPING

  45. Tri-Tech Medical Inc. Gas Control Panels and Gas Control Panel Conversion Kits

  46. Stainless steel front Self-relieving regulator Frontall™ DISS outlet features High quality ball valve Painted steel rough-in box 2” Inlet & outlet pressure gauges Gas Control Panel

More Related