Principles of Mechanical Ventilation

Principles of Mechanical Ventilation. RET 2284L Module 2.0 Set Up/Check of a Volume Ventilator. Set Up/Check of a Volume Ventilator. Ventilator Circuit Assembly. Set Up/Check of a Volume Ventilator. OVP – Operational Verification Procedure Usually an automated self-test

Télécharger la présentation

Principles of Mechanical Ventilation

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Principles of Mechanical Ventilation RET 2284L Module 2.0 Set Up/Check of a Volume Ventilator

  2. Set Up/Check of a Volume Ventilator • Ventilator Circuit Assembly

  3. Set Up/Check of a Volume Ventilator • OVP – Operational Verification Procedure • Usually an automated self-test • Usually includes a system-leak test to check the integrity of the ventilator circuit, humidifier and related equipment • Done before placing a patient on the ventilator for the first time and before reconnecting the patient to a ventilator if the circuit has been changed or disassembled for any reason • Must be documented

  4. Set Up/Check of a Volume Ventilator • Circuit Compressible Volume Loss • Determination of Circuit Compressible Volume Loss • Set VT to 100-200 and PEEP to zero • Set inspiratory pause at 2 seconds • Select a minimum flow rate and maximum pressure limit • Occlude Y-connection and initiate a mechanical breath • Record the exhaled volume (ml) and peak inspiratory pressure (cm H2O) • Divide exhaled volume by PIP (V/PIP) = circuit compression factor • Multiply compression factor by the patient’s PIP (PIP minus PEEP if PEEP is used) Example Circuit compression factor = 150 ml / 50 cm H2O = 3 ml/cm H2O Circuit compression volume = 3 ml/cm H2O x (60 cm H2O PIP – 10 cm H2O PEEP) 3 ml/cm H2O x 50 cm H2O = 150 ml (circuit volume Loss)

  5. Set Up/Check of a Volume Ventilator • Circuit Compressible Volume Loss • The amount of volume lost can be added to the VT setting to ensure that the patient is receiving the desired tidal volume Example: A patient’s estimated VT is 400 ml. Her peak pressure reading during inspiration is 30 cm H2O and circuit compression factor is 2.9 mL/cm H2O. What is the actual VT delivery to the patient? How should you set the desired VT? Volume lost = 2.9 mL/cm H2O x 30 cm h2O = 87 ml; actual volume received by the patient = 400 – 87 mL = 313 mL. To compensate, increase set VT to about 487 mL to deliver the 400 mL desired.