Reprocessing Endoscopes and Other Reusable Medical Equipment (RME): Lessons from the NMVA

Reprocessing Endoscopes and Other Reusable Medical Equipment (RME): Lessons from the NMVA. Carol Moore, MSN, RN, CIC Infection Control Practitioner, NMVAHCS. Updated Guidelines. VA Health Care in the News. Valve Problem Cited In Colonoscopy Issue (Lambert, 2009)

bailey
Télécharger la présentation

Reprocessing Endoscopes and Other Reusable Medical Equipment (RME): Lessons from the NMVA

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Reprocessing Endoscopes and Other Reusable Medical Equipment (RME): Lessons from the NMVA Carol Moore, MSN, RN, CICInfection Control Practitioner,NMVAHCS

  2. Updated Guidelines

  3. VA Health Care in the News Valve Problem Cited In Colonoscopy Issue (Lambert, 2009) VA Hospital: Patients Exposed To HIV, Hepatitis (JustNews, 2009) VA Continues Notification Process for Veterans Affected by Reprocessing Issues (Reuters, 2009) Congressional panel to analyze VA hospitals' mistakes (Poovey, 2009) Inspections find many VA facilities lack colonoscopy safety (CNN, 2009) VA Endoscope Problem is Department-Wide (House Committee on Veterans' Affairs, 2009 ) VA Allocates $26M for Endoscopy Sterilization Equipment (EndoNurse, 2009)

  4. Safety Communication from FDA, CDC, and the VA • Preventing Cross-Contamination in Endoscope Processing, November 2009 (FDA, 2009a)

  5. The Joint Commission Infection Control Focus: • Updated Position on Steam Sterilization (JC, 2009a) • Reducing Infection Risks Associated with Medical Equipment, Devices, and Supplies(JC, 2009b)

  6. Reusable Medical Equipment (RME) • RME is medical equipment designed by the manufacturer to be reused for multiple patients. • Medical equipment designed to be used multiple times by one patient is not RME unless the manufacturer provides instructions for reprocessing between patients. • Single-use disposable medical devices are not reprocessed for reuse.

  7. Reprocessing RME • Reprocessing (cleaning, disinfection and/or sterilization) RME is essential to ensure that the equipment does not transmit infectious pathogens to patients. • Failure to properly clean, disinfect or sterilize equipment creates a risk for: • person-to-person transmission (e.g., Hepatitis B virus, HIV, Salmonella, M. tuberculosis, etc.) • transmission of environmental pathogens (e.g., Pseudomonas, atypical mycobacteria, etc.) • transmission of multi-drug-resistant organisms (MDRO) (Rutala, Weber, & HICPAC, 2008)

  8. Reprocessing Methods • The selection of reprocessing method (cleaning, disinfection or sterilization) is determined by the: • intended use of the equipment • instructions of the RME’s manufacturer • compatibility of the device with the methods available • impact of the method on the device over time (Rutala, Weber, & HICPAC, 2008)

  9. Intended Use The Spaulding Scheme classifies RME according to the degree of risk for infection based on the intended use of the item. • Critical • Semi-Critical • Non-Critical (Rutala, Weber, & HICPAC, 2008)

  10. Reprocessing Method by Intended Use

  11. Transmission Risk to Patients • The most common reasons pathogens are transmitted to patients from RME include: • Inadequate cleaning • Improper selection of disinfecting agent • Failure to follow recommended procedures for set-up and use or reprocessing the RME • More healthcare-associated outbreaks have been linked to endoscopes than to any other RME (Rutala, Weber, & HICPAC, 2008)

  12. Use and Reprocessing RMEThe Process • Set-up & Use • Pre-cleaning / Cleaning • Disinfection or Sterilization • Storage

  13. SET-UP & USE

  14. Patient Safety AlertsVHA Dec 08 / Olympus Feb 09

  15. Washing Tube vs Auxiliary Water Tube

  16. Set-up & Use • Record use of each endoscope, including model, serial number, and unique identifier • Document the patient upon whom the endoscope was used, the date and time of use, the location of use, and the type of procedure • Record the system (model and serial number of the Automatic Endoscope reprocessor (AER), if applicable, used to reprocess the endoscope and the identification of the person(s) responsible for reprocessing the scope. (FDA, 2009a)

  17. PRE-CLEANING / CLEANING

  18. Pre-cleaning • At the point of use (immediately after use) • Removes visible biomaterial so it does not dry and harden • Flush all channels • May include enzymatic detergents – follow manufacturers guidelines • May use enzymatic sprays to prevent drying (Catalone & Koos, 2005)

  19. Cleaning • Cleaning is the removal of biomaterial and microorganisms using water with detergents or enzymatic products • Accomplished manually or mechanically • Cleaning alone reduces bioburden by an average of 4 logs ( 99.99%) • Is essential before high-level disinfection and sterilization because debris will interfere with these processes • Cleaning and low-level disinfection of non-critical RME is sometimes completed at the same time with a combination detergent and disinfectant product (Rutala, Weber, & HICPAC, 2008)

  20. Enzymatic Detergent Errors • Improper dilution of the detergent • measure the detergent and the water • Reuse of the Enzymatic Detergent Solution • dispose after each use • Device is not fully submerged • make sure the sink or container is large enough • Inadequate contact time • use a timer • Inadequate rinse • determine the rinse volume and total number of rinses required (Catalone & Koos, 2005)

  21. Cleaning Endoscopes • Leak test • Enzymatic detergent recommended by the manufacturer (or equivalent) • Channel brushes – the right size, single use or reusable brushes that are processed after each use, inspect before use • Clean all channels, even if not used during the procedure • Some scopes have unique channels • Use cleaning adapters supplied by the manufacturer • If using an AER, compare AER and endoscope manufacturer’s instructions – ensure they are compatible • Prevent splash and aerosol contamination of processed scopes if cleaning and disinfection occur in the same room

  22. Preventing Exposure to Infectious Agents & Chemicals • Transporting Soiled RME • Transport to the reprocessing location in covered containers (impervious bags, covered/closed carts, tote-boxes, or red bags) identifiable as BIOHAZARDOUS • Handling Contaminated RME • Personal Protective Equipment (PPE) is required to reduce the risk of healthcare worker exposure to pathogens and to the chemicals used for cleaning/disinfection • The choice of PPE is determined by the risk of exposure during the cleaning process or from the environment • Disinfection is required before handling RME without PPE

  23. Disinfection & Sterilization

  24. Reprocessing MethodsChoose the correct process • Sterilization • Destroys or eliminates all forms of microbial life • High level disinfection • Eliminates all microorganisms, except large numbers of bacterial spores • Clostridium difficile spores are reliably killed by high-level disinfection methods • Intermediate level disinfection • Kills vegetative bacteria, mycobacteria, most viruses, most fungi, but not bacterial spores • Effective against bloodborne pathogens • Low level disinfection • Kills most vegetative bacteria, some fungi, and some viruses (Rutala, Weber, & HICPAC, 2008)

  25. Reprocessing Methods • Sterilization methodsinclude:Heat (steam or hot air), Ethylene oxide (EtO), Peracetic acid (e.g. Steris System 1) and Hydrogen peroxide gas plasma (e.g. Sterrad) • High level disinfection methods include: Orthophthalalde-hyde (Cidex OPA), >2% Glutaraldehyde , Hydrogen Peroxide (H2O2) and Wet Pasteurization • Intermediate level disinfection methods include: Ethyl or Isopropyl alcohol (70-90%), Sodium hypochlorite (diluted bleach solution) Phenolic or Iodophor germicidal detergents • Low level disinfection methods include: Same as Intermediate plus Quaternary ammonium germicidal detergents (Rutala, Weber, & HICPAC, 2008)

  26. Environmental Conditions: Central Processing & Endoscopy • Physical barrier between the decontamination area and the clean area to prevent droplet contamination • Work flows from contaminated to clean without backtracking • Constructed of impervious surfaces that withstand scrubbing (floors, walls & ceilings) • Flush ceilings with enclosed fixtures, no exposed pipes • Appropriate PPE (e.g. gloves, fluid-impervious gowns, face-shields, etc) available • Daily (when in use) cleaning schedule • Eye Wash Station(s) • Dedicated handwashing sink • Traffic restricted to authorized personnel (AIA, 2006; AAMI, 2006)

  27. Ventilation, Temperature & Humidity (AIA, 2006; AAMI, 2006)

  28. Reprocessing Location “In hospitals, perform most cleaning, disinfection, and sterilization of patient-care devices in a central processing department in order to more easily control quality” Category II • Issues: • Do you chemically sterilize or high-level disinfect in the decontamination room or the prep (clean) room? • How do you transport items that are packaged & sterilized? • Are chemical sterilization or high-level disinfection point of use methods only? (Rutala, Weber, & HICPAC, 2008)

  29. The Joint Commission Update: Steam Sterilization • Flash Sterilization • Unwrapped instruments for 3 minutes at 2700F at 27-28 pounds of pressure • Survey process-related issues: • Flash sterilization utilization • Cleaning process before flash • Chemical indicators • Transport back to sterile field (JC, 2009a)

  30. Laparoscopes / Arthroscopes / Cytoscopes • Sterilize or at least high level disinfect, follow high level disinfection with sterile water rinse • Endoscopes used in normally sterile body sites are critical and must be sterilized. • Steam • Low temp chemical gas/plasma/vapor • Liquid chemical sterilants only if incompatible (Rutala, Weber, & HICPAC, 2008) (FDA, 2009a)

  31. Endoscope Accessories • Reusable accessories inserted into endoscopes that break the mucosal barrier are critical items – sterilize between uses (e.g. biopsy forceps) • Check manufacturers instructions - sterile processing is recommended for many accessories used with endoscopes (e.g. in-sheath baskets, savory guide wires and water bottles) (Rutala, Weber, & HICPAC, 2008)

  32. Protective Sheaths • OK to use probe covers to reduce the level of contamination • Condoms Can FailDo not use a lower category of disinfection or cease to follow the appropriate disinfectant recommendations • Ultrasound probes used during surgical procedures also can contact sterile body sites - Use sterile sheath • Sterilize between each patient use or if not possible at a minimum high-level disinfect • Replace non-immersible probes with fully immersible probes as soon as possible (Rutala, Weber, & HICPAC, 2008)

  33. FDA notice and recommendations for Steris System 1 Dec 2009: • FDA has not determined whether the SS1 is safe or effective for its labeled claims, including claims that it sterilizes medical devices • Assess patient care needs and disinfection and sterilization requirements and transition to an alternative as soon as possible (within 3-6 months) • Some reports of malfunctions of the SS1 that had the potential to cause or contribute to serious injuries to patients, such as infections. • No confirmed infections (no patient notification required) • Reports of injuries, mostly burns from exposure to the sterilant solution Feb 2010: • Steris Corporation has chosen not to seek FDA clearance of this device, discontinue use as soon as practicable. • Transition period extended to 18 months (FDA, 2009b; FDA, 2010b)

  34. FDA-cleared Endoscope Washer Disinfectors / Automated Endoscope Reprocessors

  35. FDA-Cleared Low Temperature Sterilizers Using Hydrogen Peroxide or Ozone

  36. Ortho-phthaldehyde

  37. Glutaraldehyde • Glutaraldehyde is noncorrosive to metal and does not damage lensed instruments, rubber. or plastics. • Might have toxic effects on the gametes and embryos –consider other HLD for transvaginal transducers • Acute or chronic exposure can result in skin irritation or dermatitis, mucous membrane irritation (eye, nose, mouth), or pulmonary symptoms • Enforcement of the OSHA ceiling limit was suspended in 1993 by the U.S. Court of Appeals 577 • ACGIH recommends limiting employee exposure to 0.05 ppm (>0.05 can irritate the eyes, throat, and nose) • Ducted exhaust hoods, ductless fume hoods with absorbents, provide 7-15 AEH, tight-fitting lids on immersion baths • PPE: nitrile or butyl rubber gloves but not natural latex gloves and goggles to minimize skin or mucous membrane contact • Schedule Monitoring (AAMI, 2005)

  38. Hydrogen Peroxide • Haag-Streit reusable tonometer tips (prisms): • 3% hydrogen peroxide and soak for 10 minutes • 3% hydrogen peroxide solution must be disposed of and replaced after 4 hours, once poured out of the bottle • Rinsing under running cold tap water suitable for drinking for ten minutes (VHA, 2007)

  39. Rinse & Drying • Remove all residual solution • Rinse according to manufacturers (device and HLD) instructions (duration, number of water exchanges, running water, water exchanges, water quality, etc) • Flush all channels multiple times with water, followed by air • Flush channels with alcohol, followed by forced air to facilitate drying (see manufacturers instructions) • External surfaces of instruments should be thoroughly wiped dry with a sterile, lint-free cloth. • Transport in a manner to protect from contamination (handle with clean gloves)

  40. Delayed Endoscope Reprocessing • More than 1 endoscope used during same procedure, scopes used in beginning of an OR case • Weekend and Emergent Use (GI scopes, bronchoscopes, TEE) • Some manufacturers' recommend a 10 hour soak in the event of delayed reprocessing, yet unnecessary or extended immersions may damage the scope (Catalone & Koos, 2005)

  41. STORAGE

  42. Storage RMEStore in a manner that reduces the potential for contamination • Sterile products • Limited access area that is well-ventilated, protected from dust, moisture, insects and temp/humidity extremes • 8 to 10 inches above the floor, at least 18 inches below the ceiling or the level of the sprinkler heads, and at least 2 inches from outside walls. • Positioned to prevent crushing or puncture • Time or event related shelf-life • Just-in-time processed products (Sterilization or HLD) • Protect from contamination or damage • Storage container must be cleanable (AAMI, 2006)

  43. Endoscope Storage Cabinets • Doors that close • Vented- allows air circulation • Adequate height to allow endoscope to hang vertically without touching the bottom • Internal surfaces with cleanable materials • Hangers constructed to support endoscopes • Sufficient space to store without touching • Do not store in the carrying case (Burlingame, 2009; Rutala, Weber, & HICPAC, 2008 )

  44. Reprocessing before use • No recommendation is made regarding the reprocessing of an endoscope again immediately before use if that endoscope has been processed after use according to the recommendations in this guideline. Unresolved issue. (Rutala, 2009) • American Association for the Study of Liver Diseases (AASLD), American College of Gastroenterology (ACG), American Gastroenterological Association (AGA) and American Society for Gastrointestinal Endoscopy (ASGE) jointly recommend the processed flexible endoscopes should be reprocessed within 10-14 days (VA memo, Dec 2, 2009) • VHA Policy: Reprocess after 12 days • Scopes not used frequently should be stored separately, consider sterilizing and storing sterile.

  45. QUALITY ASSURANCE

  46. Standard Operating Procedures • SOPs • Endoscopes are NOT all processed the same – review manufacturer guidelines for differences (e.g. you cant have one SOP for all GI scopes) • “Cheat sheets” – make sure they match

  47. Performance Measures • Supervision / Authority / Responsibility / Accountability • Training • Competency • Monitoring - Audits • Reporting

  48. Verification of processing • No recommendation is made about routinely performing microbiological testing of either endoscopes or rinse water for quality assurance purposes. Unresolved issue • VHA highly recommended: Testing to ensure bio-burden has been removed after reprocessing endoscopes that have been used for biopsy, (both the biopsy and suction channels). (VHA Directive, 2009) (Rutala, Weber, & HICPAC, 2008 )