Nuclear Magnetic Resonance

Nuclear Magnetic Resonance. ANIMATED ILLUSTRATIONS. MS Powerpoint Presentation Files. Uses Animation Schemes as available in MS XP or MS 2003 versions. A class room educational material. File-12 1hr_lecture; PULSED NMR. http://ugc-inno-nehu.com/links_from_web.html. Z. Z.

Télécharger la présentation

Nuclear Magnetic Resonance

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Nuclear Magnetic Resonance ANIMATED ILLUSTRATIONS MS Powerpoint Presentation Files Uses Animation Schemesas available in MS XP or MS 2003 versions A class room educational material File-12 1hr_lecture; PULSED NMR http://ugc-inno-nehu.com/links_from_web.html 'PULSED NMR' by Dr.S.Aravamudhan

  2. Z Z External Magnetic Field Y Y Z Z X X Y Y At t =0, the end of pulse Reference in phase at NMR (300MHz) frequency Phase Sensitive detector P-Xπ/2 X X Magnetization Y Y Y Y Chemical substance Spin ensemble X X X X Output (‘0’ freq) Phase Sensitive detector Reference in phase; offset from NMR frequency (300±0.003 MHz) Output at offset frequency (audio range) ~3KHz Rotating x,y axes :rotation about Lab z-axis A BLUE line for z-Axis indicates the view from within the rotating coordinate system. Viewed from within the rotating frame the RF field appears stationary Tilted Magnetization in xy plane viewed from Lab Frame. Precessing at resonance frequency. Z After the pulse: at t>0 Y Induced NMR signal at receiver (RF 300 MHz ) Apply the 90º, -X pulse now, P-Xπ/2 Magnetization in XY plane appears stationary when viewed in Rotating Frame from within the rotating frame No moreCLICKs.This show has automatic timings from this stage. Z A depiction of the Induced RF signal Characteristics would appear……… A rotating RF magnetic field results on application of RF at resonance frequency Free Induction NMR Signal x,y-axes Rotating about Lab Z-axis; frequemcy same as the precession frequency The rotating magnetic field tilts the magentizationaway from z-axis by 90º for a π/2 pulse D.C. Transverse Relaxation and magnetization decay in XY plane is not depicted. Rotating system viewed from within that system: STATIONARY Right CLICK mouse And CLICK on option “PREVIOUS” OR…………. CLICK toTransit. NOF.I.D.yet! 'PULSED NMR' by Dr.S.Aravamudhan

  3. Z External Magnetic Field Y X T1 (1-e-t/ ) I0 T1 It = -1/2 Magnetization Magnetization Magnetization +1/2 random Sample: Ensemble of spins I0 It CLICK ! t It Magnetization Builds up exponentially To repeat the above events:Right Click & Select option ‘previous” Initially, before the external magnetization is applied, the spins are randomly oriented When the magnetic field is turned on, the spins align at the characteristic longitudinal relaxation time T1 On the application of field….. CLICK ! Splitting is instantaneous & population redistribution requires more time called the relaxation time -1/2 CLICK for….... On-set of Longitudinal Relaxation No radiations are present CLICK ! Not stimulated transitions: but spontaneous relaxation transitions No net magnetization CLICK for…. Magnetic field Net magnetization along Z-direction & ZERO XY component +1/2 Degeneracy removed/Energy levels split Thermal equilibrium Boltzmann distribution 'PULSED NMR' by Dr.S.Aravamudhan

  4. Z Y X Y Y Y X X X In terms of Angular momenta, Izreplaces‘z’;forrotationabout z-axis= e-iφIz Represents rotation by angle φ about z-axis; Φ can be replaced by frequency of rotation in radians ‘ω’ multiplied by ‘t’ the time lapsed. Rotation about z-axis= e-i ω tIz Viewed from within the rotating frame the RF field appears stationary Z= unit vector along z-axis Rotation about z-axis= e-iφZ Represents rotation by angle φ about z-axis; Φ can be replaced by frequency of rotation in radians ‘ω’ multiplied by ‘t’ the time lapsed. Rotation about z-axis= e-i ω tZ RF field is along –X in the XY plane, the effect caused would be rotation about X-axis,unlike the precession about z-axis An equation representing this rotation would be displayed Repeat pulsing?.....Right Click and choose menu option ‘previous’ and CLICK! CLICK ! CLICK ! CLICK ! Z To repeatthe animated RF depictions “right click” and choose option:‘previous’ Click to end this slide A rotating RF magnetic field results on application of RF at resonance frequency x,y-axes Rotating about Lab Z-axis; frequency same as the precession frequency For a π/2 pulse the value of ‘ω1 t ‘=90º; ω1=γH1 The impulseoff… The impulseon… CLICK ! H1e-iI-xωt H1 e+iI-xωt A Pulse lasts only for a few μ Secs. For proton NMR a H1 of ~25Gauss along ‘-x’ , pulse widths are approximately 10-15μs + 2 H1I-x cos(ωt) = Only one of the rotating component is effective in causing resonance Rotating system viewed from within that system: STATIONARY RF source/ transmitter Connected to coil. Linearly oscillating field along the coil axis (X-axis) The linearly oscillating field can be resolved into two counter rotating components CLICK ! http://www.geocities.com/sankarampadi/eulexp.html 'PULSED NMR' by Dr.S.Aravamudhan

  5. Z Z Y Y X X Y X t=0 FID tp Acquisition time ~5T2 Rotating x,y axes :rotation about Lab z-axis A BLUE line for z-Axis indicates the view from within the rotating coordinate system. Viewed from within the rotating frame the RF field appears stationary Tilted Magnetization in xy plane viewed from Lab Frame. Precessing at resonance frequency. No More Clicks ! This show has automatic timings Z After the pulse: at t>0 Y If No T2…….. The F.I.D. Tilting of magnetization Apply the 90º, -X pulse now, P-Xπ/2 Magnetization in XY plane appears stationary when viewed in Rotating Frame from within the rotating frame At the end of pulse, time for F.I.D. begins with t=0 Described in rotating frame: Rotation about the X-axis I(tp) =e-iI-xφIz e+iI-xφwithφ=90º & tp is pulse duration When the XY magnetizationdecays with transverse relaxation time T2, immediately after the pulse…… When PSD reference is in phase off set from Resonance frequency; NMR signal at receiver (RF 300 MHz ) Induced NMR signal at receiver (RF 300 MHz ) When PSD reference is in phase at Resonance frequency; NMR signal at receiver (RF 300 MHz ) CLICK to Transit FreeInductionDecay Signal 'PULSED NMR' by Dr.S.Aravamudhan

  6. Relaxation Longitudinal and transverse Magnetic field I h NET Magnetization Alignment….. Random… CLICK ! Longitudinal T1 Relaxation Transverse T2 Relaxation Randomization in XY plane: Magnetization Decays t A π/2 pulse flips the z-magnetization to xy-plane 'PULSED NMR' by Dr.S.Aravamudhan