The Digital Triage Assistant
220 likes | 556 Vues
The Digital Triage Assistant. Cameron Abt Peter Backeris Glenn Shevach. Customer Problem/Need. Emergency Department Triage Determines who is seen, how long to wait Effective triage is crucial in crowded ERs Vital Signs Poor vital sign recording during emergency department triage
The Digital Triage Assistant
E N D
Presentation Transcript
The Digital Triage Assistant Cameron Abt Peter Backeris Glenn Shevach
Customer Problem/Need Emergency Department Triage • Determines who is seen, how long to wait • Effective triage is crucial in crowded ERs Vital Signs • Poor vital sign recording during emergency department triage • Vital signs incomplete or skipped • Respiration rate not automated • Errors in manual transcription
Engineering Problem Incomplete Vital Signs • Not simultaneous • Not automated • Not integrated • Time consuming • Trained professional required • High cost • Not mobile Respiration Rate • Requires manual measurement • Inaccurate and often estimated Manual Transcription • Error prone • Time consuming • Unnecessary paperwork
The Digital Triage Assistant • Solves current problems in emergency room triage • Four major subsystems • VSAD • PDA • PC Software Suite • Backend database vTriageTM
Vital Signs Acquisition Device • Miniature, portable, ergonomic design • Measures 5 vitals through 3 points of contact • Blood pressure • Heart rate • Oxygen Saturation (SpO2) • Temperature • Respiration Rate • Combines COTS and proprietary circuits
Vital Signs Acquisition Device • Custom designed motherboard • Based on dsPIC33 16 bit microcontroller • Responsible for • Power distribution and monitoring • Intelligent inter-module communication • Communication with PDA • Acquisition and analysis of respiration rate and temp
Vital Signs Acquisition Device • PCB design • 2 layer, bottom ground plane • Digital and analog sections • SMT 0805 passives • Analog test points • Critical component selection • 2.5” x 3.8”
Vital Signs Acquisition Device • Power monitoring • Tenergy Li-Ion battery • 7.4V, 2200 mAh • Proprietary voltage monitoring • Docking station with charging
Vital Signs Acquisition Device • 3 commercial modules • SunTech Medical Advantage A+ • Non-invasive blood pressure, heart rate • 9600 baud UART connection • Nonin Medical OEM III (SpO2, HR) • Oxygen saturation, heart rate • 9600 baud UART connection • Parani ESD 200 Bluetooth Module • Class 2 • 30m nominal range • 115200 baud • Bluetooth serial port profile
Respiration Rate • Measurable parameters • Chest expansion • Chest band/strain gauge • Airflow • Strain gauge • Sound • Bio-acoustic sensor • Exhaled air temperature • Thermopile
Respiration Rate • Respiration rate measurement • Thermocouples: J-type (iron-constantan) • Interested only in ΔT from thermopile • Amplification using INA128, instrumentation amp
PDA • HP iPAQ 210 Enterprise Handheld • Communicates with VSAD through Bluetooth • Custom software (mTriageTM) • Touch screen UI • Windows Mobile 6 • Developed with .NET Framework 3.5, MSVS 2008 • SocketMobile barcode scanner • Wireless LAN connection
PDA • mTriageTM • Patient identification and information display • Real-time data display • Intelligent data collection • Connects to SQL Server 2008 database
vTriageTM • Administrative software suite • Uses: • Entering patient information • Patient registration • Printing triage reports • Real-time patient monitoring • Intelligent health alert system
Network Database • Holds past and present patient information
Future Work/Applications • Modular sensor suite • Healthcare data mining • Triage Decision Support • Mass casualty events • Sporting events • Military applications • Connect to database using 3G • Ambulance, police, fire trucks
Conclusion • Designed and tested a novel respiration rate sensing technology • Developed a FULLY FUNCTIONAL system to streamline the triage process in hospital ERs • Come get triaged! Thank You!