Four-Year Results of the Pivotal U.S. Multicenter Trial of the Endologix Powerlink Endograft for EVAR

Four-Year Results of the Pivotal U.S. Multicenter Trial of the Endologix Powerlink Endograft for EVAR. Rodney White, MD Harbor UCLA Medical Center Torrance, CA. Conflicts of Interest & Off-Label Use Disclosure. Investigational trial support from multiple manufacturers

Télécharger la présentation

Four-Year Results of the Pivotal U.S. Multicenter Trial of the Endologix Powerlink Endograft for EVAR

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Four-Year Results of the Pivotal U.S. Multicenter Trial of the Endologix Powerlink Endograft for EVAR Rodney White, MD Harbor UCLA Medical Center Torrance, CA

  2. Conflicts of Interest &Off-Label Use Disclosure • Investigational trial support from multiple manufacturers • Paid consultant and advisory boards for • Medtronic, Boston-Scientific, WL Gore, Endologix, General Electric – OEC, and Volcano

  3. Powerlink® System: • Unibody-Bifurcated Design • Long Main Body • Low-Porosity Proprietary ePTFE Formulation • Cobalt Chromium Alloy Stent • Single-wire Main Body Construction • Fully Supported

  4. Powerlink® U.S. Pivotal Trial • 192 patients enrolled between July 2000 and March 2003 • Eligibility Criteria • Proximal Infrarenal Neck: • >15mm length • <60° angle • 26mm maximum diameter, 18mm minimum diameter • AAA >4.0 cm diameter or rapidly growing AAA • Iliac diameter >7mm on at least one side (for access) • Dispensable inferior mesenteric artery • Preservation of at least one hypogastric artery • Iliac seal zone of >15mm length (<18mm diameter) • Aortic bifurcation diameter >18mm • >18 years old • Not pregnant • Candidate for open AAA repair • Serum creatinine <1.7mg/dlL • Willingness to comply with follow-up schedule • No bleeding disorders • Life expectancy >2 years • No connective tissue disorders

  5. Summary of Significant Early Clinical Findings • † Defined as death, MI, stroke, AAA rupture, conversion, secondary procedure, coronary intervention, renal failure, or respiratory failure • * Not device related Carpenter JP, et al. Midterm results of the multicenter trial of the Powerlink bifurcated system for endovascular aortic aneurysm repair. J Vasc Surg 2004;40:849-59.

  6. Secondary Procedures through 60 months* 34 procedures in 26 patients • Endoleak –23 (Cuffs, embolization, balloon dilatation) • Type II Endoleak – 17 • Type I Endoleak – 6 • Graft Limb Occlusion – 7 • Embolectomy,Stent, PTA, or Lytic Therapy • Native Artery Procedures – 3 • Aortic Neck Dilation - 1 * As of December 16, 2005

  7. Conversion, Rupture, Migrationthrough 60 months* • Surgical conversions – 4 • Intraoperative – 3 • @12 mo. – 1 • Aneursym rupture – 0 • 5 Migrations (> 10mm) • All caudal • None leading to secondary procedure * As of December 16, 2005

  8. Core Lab AnalysisEndoleaks (All Types, New and Persistent)† †As of December 16, 2005

  9. Sac Diameter Over Time† † As of December 16m 2005 * 192 patients enrolled, 3 patients’ CT’s lost before submission to core lab

  10. Sac Volume Over Time† • † As of December 16, 2005 • * 192 patients enrolled, 3 patients’ CT’s lost before submission to core lab • ** Some CT scans are not evaluable for some parameters most often due to poor image quality, no contrast, CT’s taken at greater than 3mm slices, etc.

  11. Morphology Change Sub-Analysis Method: • Patients with paired data (both a 1-month and either a 2-year, 3-year, 4-year, or 5-year follow-up CT) as of December 16, 2005 were included in this analysis • Current Analysis N = 78 pts. • All CT’s were analyzed by study core lab utilizing proprietary 3-D reconstruction software (Medical Metrx Solutions, West Lebanon, NH)

  12. Analysis: • Mean max. sac diameter • Mean sac volume • Distance between distal renal and aortic bifurcation • Δ in maximum sac angulation • Sac angle calculation was determined from distal proximal neck to maximum inflection point in the sac to aortic bifurcation (180° = straight line through sac)

  13. Overall Results:

  14. Paired Results:

  15. Example: Pt. w/ 22.6° Straighter Sac Angle 1 – Mth. 3 – Yr.

  16. Distal End of Bifurcated Region

  17. 4-29-05 5-13-05

  18. Summary of Late Clinical Findings • 97.9 % Freedom from AAA-Related Mortality at 4 years with the Powerlink System • No aneurysm ruptures • Only 1 late conversion (@ 1 yr.) • No ePTFE graft material failures at 4 years • No cobalt chromium stent graft failure or fatigue at 4 years

  19. Conclusions: • Powerlink System appears safe and effective in protecting patients from AAA rupture • All cohorts treated w/ Powerlink® experienced: • Aneurysm sac diameter & volume reduction • Remodeling or shrinking of aneurysm sac around device as evidenced by straightening of the aorta • Minimal change in length of aorta from distal renal to bifurcation regardless of sac angle change • Sac remodeling does not appear to contribute to graft migration with the Powerlink stent graft • The column strength and long main body of the Powerlink® may directly influence the classical remodeling of the aneurismal aorta