An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. FDA Human Factorsof Medical Devices Ensuring a safe, user friendly device-user interface

  2. Medical Errors in U.S. each Year Result In: • Up to 98,000 deaths • 5th leading case of death: exceeds auto accidents, breast cancer and AIDS • $29 Billion added cost

  3. FDA Medical Device Incident Reports: • 100,000 reports per year • More than 1/3 involve use error • Tip of the Iceberg

  4. FDA Recall Study • 44% of medical device recalls due to design problems. • Use error often linked to design.

  5. More user friendly Design • Operations - don’t exceed user capabilities • Information - Sufficient, legible and intelligible • Procedures - logical and intuitive • Operations - consistent with conventions • Dangerous error - designed out • Conditions of use - considered and addressed

  6. Key safety concepts in design • make things visible • simplify the operation • avoid reliance on memory • avoid reliance on vigilance • use natural mappings • use forcing functions • make it easy to reverse an error

  7. FDA’s authority • Quality Systems Regulation/ CGMP - Design Controls (1996)

  8. Design Controls • Regulatory Language: • “… design requirements … intended use … needs of the user and patient.” • “… testing production units under actual or simulated use conditions.” • “… conduct risk analysis.”

  9. Design Controls • Preamble Language: • “… conduct appropriate human factors studies, analyses, and tests …” • “… human interface includes both the hardware and software characteristics…”

  10. Design Controls - Human Factors Engineering (HFE) Process • HFE applied from concept stage to final design • Early Involvement of typical users is critical • The process is iterative

  11. Perform Studies & Analyses Develop Require- ments Develop Specs. Test Output Against Input Test Against Patient & User Needs Human Factors Engineering Process Concept Phase Design Input Design Output Verification Validation * HF Elements Analyses Expert Evaluation Rapid Prototyping Literature Complaints Observation Interviews Safety Environment Users Performance Drawings Mockups Computer Prototypes Production Units Full Usability Test Risk Assessment *Design and Development Planning Important Here

  12. Human Factors Engineering - Summary of Methodology • Study the user population and use conditions • Analyze function, tasks, and hazards • Incorporate findings in requirements

  13. Human Factors Engineering - Summary of Methodology • Test and analyze prototypes against requirements • Test production models in simulated environment • Conduct a risk assessment

  14. HFE Standards • ANSI/AAMI HE74:2001 Human factors design process for medical devices • IEC 60601-1-6, Ed1, Usability

  15. ANSI/AAMI HE74:2001 • Overview of the HFE process • Planning the HFE process • HFE process: a systems approach • Methods and techniques • Risk & cost benefit analysis

  16. ANSI/AAMI HE74:2001 • Need for user input • Scaling the HFE work • Documenting the HFE activities • Design evaluation: verification & validation

  17. ANSI/AAMI HE74:2001 • Relevant FDA regulations • Practical tips • Bibliography - General HFE references - Human error - Human computer interface - Usability testing

  18. FDA Web Site • http://www. fda.gov/cdrh/humanfactors.html fda.gov/cdrh/usererror.html