In this photon counting test, an 8” MCP pair undergoes repairs and enhancements for improved gain map quality. Issues include broken corners, small charge clouds, and gain non-uniformity. Explore the results for better resolution and gain uniformity.
E N D
8” 40µm MCP Pair Photon Counting Test Top MCP 100 MΩ in air ~400MΩ in vacuum XDL detector with pair, 0.003” spacer Bottom MCP has corner Missing, 25 MΩ
8” 40µm MCP Pair Photon counting Tests Zoom in – see multifibers FIRST IMAGE - stripes due to anode/charge cloud too small –fix : have just rebuilt with 0.7mm biasable, spacer. Broken off corner on bottom MCP. Anode – X, photon pulses Gain ~106, <500µm resolution
8” 40µm MCP Pair Gain Map Y gain slice Y gain slice X gain slice FIRST Gain map –multifiber visible, lots of gain non-uniformity