1 / 15

Clinical Microscopy

Clinical Microscopy. “The Microscope”. Illuminator. Halogen Bulb Usually controlled by a rheostat (knob that controls the intensity). . Condenser. Compound lens system that focuses the light from the illuminator onto the specimen.

issac
Télécharger la présentation

Clinical Microscopy

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Clinical Microscopy “The Microscope”

  2. Illuminator • Halogen Bulb • Usually controlled by a rheostat (knob that controls the intensity).

  3. Condenser • Compound lens system that focuses the light from the illuminator onto the specimen. • Gathers a large amount of light and condenses it into a small area.

  4. Aperture Diaphragm • Lever on the front of the condenser that allows you to control the amount of light leaving the condenser. • Opens and closes like the iris of you eye.

  5. Stage • Simply the place where the slide sits. • This area becomes very dirty. Keep it clean.

  6. Slide Holder • Mechanical device that holds the slide in place on the stage. Where ever the stage goes the slide goes!

  7. Objectives • Traditionally come in four flavors in the lab • 10x, 40x, 50x, and 100x • Magnify the image • 50x and 100x require immersion oil. • The oil bends light waves so more light enters the objective

  8. Oculars • Usually have a magnification power of 10x. • Form the final image. • Total Magnification: equals the oculars time the objectives. (40x times 10x = 400)

  9. Focus Knobs • Two flavors of these: • Course • Fine • Some microscopes control the course and fine focus with one knob.

  10. Stage Adjustment Knobs • Controls the stage movement. • Top knob controls forward and back movement. • Bottom knob controls left and right movement.

  11. Cleaning the Microscope • Keeping the microscope clean is extremely important. • Use a clean, lint-free lens tissue. • Don’t use a dry tissue to clean the lenses. Use a moist tissue first and then a dry one to mop up. 4 3 5 2 1

  12. Cleaning... • Start by cleaning the illuminator. • Move to the condenser • Next clean the objectives, but clean the 100x last because there may be oil on it. Once you have cleaned the 100x, get a new tissue. 4 3 5 2 1

  13. Cleaning... • Clean the oculars • Lastly, wipe off the stage. Clean it last because it usually has the most dust and oil. • Make this a daily routine. 4 3 5 2 1

More Related