Basics of Biodosimetry Part 1

Basics of Biodosimetry Part 1. Lecture Module 2. What is Biodosimetry?. Biodosimetry is the use of any biological change in an irradiated person that can be sufficiently quantified to indicate their radiation dose. The ideal specification for a biological dosemeter. Specific to radiation

jtarkington
Télécharger la présentation

Basics of Biodosimetry Part 1

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Basics of BiodosimetryPart 1 Lecture Module 2

  2. What is Biodosimetry? Biodosimetry is the use of any biological change in an irradiated person that can be sufficiently quantified to indicate their radiation dose

  3. The ideal specification for a biological dosemeter Specific to radiation Low background Low donor variability Low doubling dose Dose response calibration Persistent effect Easy sampling Rapid result Cost

  4. What are our options? For high doses: >2 Gy Prodromal nausea and vomiting Differential white blood cell counts Localised > 3 Gy Erythema Conjunctivitis

  5. Erythema and blistering

  6. Changes in neutrophil numbers after irradiation Gy <1 Infections, fever, death 1 - 2 2 - 5 Normal >5 - 6 100 % of normal 50 Critical period,infections,fever 0 15 45 0 30 180 Time, d

  7. Other possible endpoints Altered levels of biochemicals in body fluids Gene mutations Genes up- or down-regulated Increased numbers of DNA double strand breaks

  8. Our best options Cytogenetic indicators: • Dicentrics • Translocations • Micronuclei • Aberrations in prematurely condensed chromosomes

  9. Why do we need biodosimetry ? Doctors treat symptoms, not doses • Biodosimetry can detect false positives and false negatives • Biodosimetry can forewarn to expect later clinical developments • Biodosimetry can indicate partial body or inhomogeneous exposure

  10. Why do we need biodosimetry? Dose information can inform doctors in the dose range where medical intervention is needed Doses <1 Gy will need no treatment, only counselling False alarms, i.e., no dose, need reassurance Large events result in epidemiology follow-up

  11. Why do we need biodosimetry? Physical dosemeter badges do not necessarily reflect the wearer’s dose People, sometimes members of the public, who are not routinely monitored can be involved in radiation events

  12. Overall place of biodosimetry in radiation dose monitoring It supplements, but does not replace, physical monitoring It works for overdoses, it does not have the sensitivity of a badge It resolves anomalies It helps quantify dose in the absence of reliable physics It relieves anxiety

  13. Brief early history of cytogenetic dosimetry Early in the 20th century it was known that radiation damages chromosomes 1960 breakthrough; method to visualize human chromosomes in white blood cells Calibrations at Oak Ridge, USA Biodosimetry performed on mid-1960s criticality accidents in USA

  14. The dicentric assay The first type of aberration to be used Most frequently still used Therefore a large body of experience Now a routine procedure Has been described as the biodosimetry ‘gold standard’

  15. A dicentric in a metaphase chromosome spread

  16. What is the ‘dose’ that we are measuring? • Absorbed dose • Unit is gray (Gy) • We relate the chromosome aberration yield to dose by reference to a dose response calibration curve in Gy • We are measuring dose to lymphocytes • We do NOT operate in sieverts (Sv)

  17. What are the lymphocytes that we use

  18. Lymphocytes PHA stimulates T-lymphocytes Lymphocytes in the blood are non-dividing They are a synchronised population of cells that can be stimulated to divide and can be harvested at first metaphase In non-dividing cells dicentrics are predominantly induced by radiation

  19. The lymphocyte in vitro cell cycle

  20. Conclusions • This lecture has described: • Requirements for biological dosemeter • Clinically observable effects of high doses; poorly quantitative • Best options for more quantitative dosimetry; cytogenetics • Why we need biodosimetry; how it helps • Early beginnings in 1960s with the dicentric • Consideration of dose that we specify • Some basic points about peripheral blood lymphocyte