User Guide KI 734x Series Two way Loss & ORL Tester

User Guide KI 734x Series Two way Loss & ORL Tester . www.kingfisher.com.au. Two-way Loss Tester. T he fastest and easiest two-way loss tester Real time display of bi-directional average multi-  link loss on both instruments 3 year warranty & calibration cycle No warm up required

kaili
Télécharger la présentation

User Guide KI 734x Series Two way Loss & ORL Tester

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. User Guide KI 734x Series Two way Loss & ORL Tester www.kingfisher.com.au

  2. Two-way Loss Tester • The fastest and easiest two-way loss tester • Real time display of bi-directional average multi- link loss on both instruments • 3 year warranty & calibration cycle • No warm up required • 360 hours battery life & interchangeable connectors • 1200 storage space for dual  test results

  3. Fast & Easy For Bi-directional Testing Step 1: Set references Step 2: Press the “Auto test” button Real time bi-directional average multi- link loss display on both instruments

  4. Step 1 - Set Reference At Both Ends Test lead Two-way port Meter port An optic fiber link End B End A

  5. Set Reference Key Press Sequence 1310nm Source off Source Return Loss Power Meter Auto test - 52.73dBm Pushing “AUTOTEST” Alternates between 1310 and 1550 nm if 1310 & 1550 nm lasers are installed - + 1310nm Source on Auto test Menu Abs / Rel Hold - 4.01dBm -4.00dBm Pushing “Abs / Rel” - + 1310nm Source on Auto test Menu Set Ref Abs / Rel Hold Holding “Set Ref” for 3 seconds to set reference for multi- - 0.00dB R -4.00dBm Pushing “Menu” to be back to the turn-on display 1310nm Source off Source Return Loss Power Meter Auto test - 52.73dBm

  6. Step 2 - Measure Insertion And Return Loss An optic fiber link End B End A 1310nm Source off Source Return Loss Power Meter Auto test - 52.73dBm Pushing “Auto test” at one End.

  7. Measure Insertion And Return Loss - Continue Alternates between 1310 and 1550 if 1310 & 1550 nm lasers are installed - + 1310nm Source on Auto test Menu Hold - 34.01 A ◄► B dB -4.00dBm Displays the average bi-directional link loss of multi- - + Press ‘Menu’ Press ‘Select’ Displays Loss/Optical Power from A-to-B (‘A ►B’) Displays Loss/Optical Power from B-to-A (‘A◄ B’) Displays ORL at A end (‘A ►’) - + 1310nm Source on Auto test Menu Hold - 54.17 Displays ORL at B end (‘◄ B’) A ◄ B dB RL -4.00dBm Back to the average bi-directional link loss display Pushing “STORE” to store average Loss, A-to-B Loss, B-to-A Loss, ORL at end A, ORL at end B of multi- STORE Testing is completed for the link.

  8. Communication During Testing End A End B An optic fiber link - + - + 1310nm Source on Auto test Menu Hold Source on Auto test Menu Hold - 34.01 - 34.01 A ◄► B dB A ◄► B dB -4.00dBm -4.00dBm After completing one link test, pushing “Hold” will freeze the display on both units and the unit at End B beeps twice. This tells End B that a test has been completed. Then, End B connects to the next link and pushes “Hold”. This releases the display on both units and the unit at End A beeps once. This tells End A that he is ready for the next test. Increases the productivity in high fiber-count systems.

  9. Set Ref In Autotest - + Displays average link loss for multi-. Hold “Set Ref” for 3 seconds will set references on both units for multi-. 1310nm Source on Auto test Menu Set Ref Hold - 34.01 A ◄►B dB

  10. View Ref Value In Autotest Displays the average bi-directional link loss of multi- Press ‘Select’ Press ‘Menu’ Displays Loss/Optical Power from A-to-B (‘A ►B’) Displays Loss/Optical Power from B-to-A (‘A◄ B’) Displays ORL at A end (‘A ►’) Displays ORL at B end (‘◄ B’) Back to the average bi-directional link loss display - + Displays link loss and reference value for multi- for A-to-B. 1310nm Source on Auto test Menu Abs / Rel Hold - 34.01 A ►B dB R -8.20dBm Unit A reference value - + Displays link loss and reference value for multi- for B-to-A. 1310nm Source on Auto test Menu Abs / Rel Hold - 32.24 A ◄ B dB R -7.80dBm Unit B reference value - +

  11. Pass/Fail Assessment In Two-way testing mode, press ‘Level’ to enter Pass/Fail display mode. To exit press ‘Level’ again, or press ‘Select’. Level Select - + 1310nm Source on <<< Auto test Menu Hold PASS A

  12. Insertion Loss Pass/Fail Threshold Set Up Multi- of Insertion Loss Pass/Fail criteria can be set. Enter Power Meter mode.Select two-way operational wavelength. - + 1310nm Shift Source off SEt Menu 15.73 dB Use ‘Shift’ + ‘Level’ to set Average Insertion Loss Pass/Fail threshold, push ‘Set’ to store. Shift Default Level Cancel Set Select another two-way operational wavelength to set up Average Insertion Loss Pass/Fail threshold. - +

  13. Return Loss Pass/Fail Threshold Set Up Multi- of Return Loss Pass/Fail criteria can be set. - + 1310nm Shift Source off SEt Menu Enter Return Loss mode.Select wavelength. -10.57 dB RL Use ‘Shift’ + ‘Level’ to set Return Loss Pass/Fail threshold, push ‘Set’ to store. Shift Default Level Cancel Set Select another wavelength to set up ReturnLoss Pass/Fail threshold. - +

  14. The Contents of User Memory • Average loss for multi- • Reference of both units for multi- • ORL for multi- • Nominal output power for multi- • Location info by flashing symbol of either A or B

  15. KITS™ testing software