1 / 40

Chromatography

Explore various techniques for the separation of complex mixtures based on the differential affinities of substances for a mobile and stationary phase. Learn about partitioning, capillary columns, and the factors that affect separation efficiency.

kellyd
Télécharger la présentation

Chromatography

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Chromatography • Various techniques for the separation of complex mixtures that rely on the differential affinities of substances for a gas or liquid mobile medium and for a stationary adsorbing medium through which they pass.

  2. Partitioning • Separation is based on the analyte’s relative solubility between two liquid phases or a liquid and solid Mobile Phase Stationary Phase Solvent or gas Bonded Phase Solid or liquid

  3. Capillary columns are a thin fused-silica capillary. Typically 10-100 m in length and 250 µm inner diameter. The stationary phase is coated on the inner surface. In GC Separation Separation Through Long Fused Silica Column

  4. Separation • The individual components are retained by the stationary phase differently. • The components separate from each other since they are running at different speeds through the column with the eluent gas. • At the end of the column they elute one at a time depending on their retention, governed by their boiling point and polarity.

  5. GC Separation • Boiling point- lower boiling point compounds spend more time in gas phase. • Temperature of column does not have to be above boiling point. Solids have vapor pressure. • High vapor pressure liquids used as solvents (water 25 mm Hg whereas ether 520 mm Hg/25C).

  6. Separation • Polarity “like absorbs likes”. • Polar compounds have longer retention time on Polar columns. • Non-polar compounds have longer retention on non-polar column. • Two types of columns in GC 3800 and 3900, a silicone based column and a PEG column (wax column).

  7. Separation • Carrier gas flow- high flow rate decreases retention time. • High gas flow may cause poor separation since components have little time to react with stationary phase. • There is optimum gas flow for various columns. Too high a flow causes excessive back pressure.

  8. Separation • Column length- longer column usually improves resolution, but increases back pressure. • Doubling length will not double resolution (resolution increases according to square root of length). • Generally a 30 meter column gives best resolution, analysis time and head pressure

  9. Separation • An effective means to increase resolution is to decrease column ID. • The efficiency of a capillary column increases (number of theoretical plates per meter) as the ID of column decrease. • Makes sharper peaks,& decreases column bleed • Decreasing ID decreases sample capacity. • Ideal (most popular) ID is 0.25 mm.

  10. Some Types of Chromatography Lid • Paper chromatography (PC), • Thin-layerchromatography (TLC), Two experiments in MCAL demonstrate: • Liquidchromatography (LC, including high-performance liquid chromatography, or HPLC), • Gas chromatography (GC). Paper Solvent front solvent

  11. Gas Chromatography (a review) • Chromatographic technique that can be used to separatevolatile organic compounds. • A gas chromatograph consists of a flowing mobile phase, an injection port, a separation column containing the stationary phase, detector and a controller (integrator) and data collection and storage. • Organic compounds are separated due to differences in their partitioning behavior between the mobile gas phase and the stationary phase in the column.

  12. Bruker Improving the Quality of our Beer – New 456-GC Off-Flavor Beer Analyzers

  13. Varian (Agilent) 3800 GC Gas tanks He, N2, H2, Air Autosampler Detectors Injectors Oven Computer control and data acquisition

  14. Stationary Phases • The most common stationary phases in gas-chromatography columns are polysiloxanes, which contain various substituent groups to change the polarity of the phase. • The nonpolar end of the spectrum is polydimethyl siloxane, which can be made more polar by increasing the percentage of phenyl groups on the polymer. • Polyethylene glycol (carbowax) is commonly used as the stationary phase for more polar analytes.

  15. Polysiloxanes

  16. Polyethylene glycol

  17. Low Bleed Phases (Arlenes)

  18. Capillary columns are a thin fused-silica capillary. Typically 10-100 m in length and 250 µm inner diameter. The stationary phase is coated on the inner surface. Varian (Agilent) Factor Four Capillary GC Columns.

  19. Open tubular columns (capillary columns) come in two basic forms. • Wall-coated open tubular (WCOT) column. WCOT columns are capillary tubes that have a thin later of the stationary phase coated along the column walls. • Support-coated open tubular (SCOT) column. In SCOT columns, the column walls are first coated with a thin layer (about 30 micrometers thick) of adsorbant solid, such as diatomaceous earth, a material which consists of single-celled, sea-plant skeletons. The adsorbant solid is then treated with the liquid stationary phase. • SCOT columns are capable of holding a greater volume of stationary phase than a WCOT column due to its greater sample capacity, • WCOT columns have greater column efficiencies.

  20. Capillary Columns

  21. Solid Support Bonded to Wall Low efficiency not used usually

  22. Liquid On Inner Wall

  23. Liquid Bonded to Support

  24. FSOT Column

  25. Capillary Columns • Capillary columns provide much higher separation efficiency than packed columns but are more easily overloaded by too much sample. • You need to dilute your sample to get good symmetrical (Gaussian) peaks • Solvent peak dominates chromatogram due to normalization of peaks.

  26. GC Column Oven

  27. MCAL GC Columns GC 3800- one columns FID detector: 1) CP-sil 8 50 m, 0.2mm, .33um 5% phenyl 95% dimethyl polysiloxane GC 3900 –one column FID detector: 1) Supelco wax-10 15m x 0.32 mm, 0.5 u Polyetyylene glycol

  28. Simple GC Computer Gas control valve Sample injector port FID, ECD, MS Detector Separation Column Temperature controlled oven Gas is He, N2 or H2

  29. Sequence of GC Injection 1. A small amount of liquid (microliters) is injectedthrough a silicon rubber septum into the heated (>200oC) GC injector that is lined with an inert glass tube. 2. The sample is immediately vaporized. 3. A pressurized, inert, carrier gas-which is continually flowing from a gas regulator through the injector and into the GC column-sweeps the gaseous sample, solvent, and analyte, onto the column. 4. Septum purge: a small ancillary flow of carrier gas bathes the underside of the injector's septum so that hot vaporized sample gases can't interact and possibly stick to the septum. This improves peak shape and reproducibility.

  30. Split / Splitless Injector Helium

  31. Split / Splitless Injector

  32. ECD detector FID Detector Injectors

  33. FID Detector • Organic compounds burning in a hydrogen- oxygen flame produce ions and electrons. These charged particles created in the combustion process create a current between the detector's electrodes. • One electrode is the metallic jet itself, the other is above the jet. The detector housing is heated so that gases produced by the combustion (mainly water) do not condense in the detector before leaving the detector chimney.

  34. FID Detector Insulator Cathode Anode Air Hydrogen Sample plus Helium from Column

  35. FID Detector

  36. The ECD or Electron Capture Detector • The ECD or electron capture detector measures electron capturing compounds (usually halogenated) by creating an electrical field in which molecules exiting a GC column can be detected by the drop in current in the field

  37. ECD Detector Electrode insulator Radioactive Beta emitter (electrons) • - - - - • - - - - - - - - - waste GC column + electrode - electrode Electron cloud insulator

  38. GC Experiment • To compare the separation of alcohols of increasing carbon number using a general purpose silicone column and a FID detector in the Varian 3800 GC. • To measured sensitivity of instrument (LOD). Why is LOD and LOQ important in analysis and research?

  39. Seperation of Alcohols C1 to C8 on carbowax column

  40. GC Exp: Fid Detector Solvent and contamination dichlorobenzene

More Related