1 / 10

Status of the S5 calibration

Status of the S5 calibration. Michael Landry For the Calibration Committee LSC September 2008. Outline. h(f) Status V3 summary. V4 status - improvements. Expectations for S6 calibration. Next talk: Xavier Siemens: h(t) report V3 h(t) - validation, systematics, hardware injections. S6.

lam
Télécharger la présentation

Status of the S5 calibration

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Status of the S5 calibration Michael Landry For the Calibration Committee LSC September 2008 G080469-00-Z

  2. Outline • h(f) Status • V3 summary. • V4 status - improvements. • Expectations for S6 calibration. • Next talk: Xavier Siemens: h(t) report • V3 h(t) - validation, systematics, hardware injections. • S6. G080469-00-Z

  3. V3 h(f) Calibration • Valid throughout S5, from 40-3000 Hz. • Employed in several publications (e.g. Crab, GRB070201, SGR). • DC values from V4 employed in final V3 calibration. G080469-00-Z

  4. Reminder of DC issues • Significant effort has gone into understanding the DC calibrations. • at the outset of S5, DC methods (3) disagreed at the 20% level. • Now, the multiple DC methods agree within the one-sigma error bars of the official method G080469-00-Z

  5. Resultant V3 strains • DC calibration shifts • H1: 7.4%  • H2: 6.2%  • L1: 4%  • High-water BNS marks for S5 shift accordingly to: • H1 15.1 Mpc  • H2 7.3 Mpc  • L1: 15.7 Mpc  G080469-00-Z

  6. V3, V4 OLG Comparisons L1 V3 V4 G080469-00-Z

  7. Epochs Changes in Im() occur due to changes in the detector configuration or time-delay shifts. We see time-delay shifts after front-end reboots. (<30 s) No new epochs for V4 G080469-00-Z

  8. Status of V4 • Many improvements in OLG model, measurement agreement. • Time delay validation by direct measurements • Code mods and bug fixes (Pade approximation, sign error in complex conjugation) • Residuals aid high-frequency agreement • Request from burst group to go to 6.5kHz (as high as we can go, given the butterfly resonance). • L1 models complete, factors being generated. • H1, H2 models nearly done. G080469-00-Z

  9. What’s left? • Finalize H1, H2 models and generate files. • Produce error budget. • Release models for h(t) generation. • Release S5 V4 calibration documents. • Get reviewed. • Publish calibration papers. G080469-00-Z

  10. Expectations for S6 • New sensing scheme will modify the calibration some for S6. Otherwise, propose doing nothing new for this science run. Leverage what was learned in S5. • Have to have photon calibrators running in S6. • Our early calibrations for S5 were accurate at the level of 10% (plus systematics). Expect the same for S6, but with systematics correctly accounted for. • Need access to the machines prior to run in order to obtain h(f) and h(t) for online analyses. • Clearly we need to move to a quicker release of finalized h(t) for searches. • Recruitment of volunteers for h(t) was successful (see following talk), we’re recruiting for h(f). • DAC summit: we’ll ask the search groups, what calibration can they live with. G080469-00-Z