1 / 4

Calibration Report Guidelines for Acceptance Criteria

Learn about setting acceptance criteria in calibration reports, referencing standards like ISO 8655-3 to evaluate systematic and random errors.

may-mason
Télécharger la présentation

Calibration Report Guidelines for Acceptance Criteria

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Check the calibration report (Volume)

  2. กำหนดเกณฑ์การยอมรับ • กำหนดจาก • Tolerance บริษัทผู้ผลิตหากค่า Max. Systematic error ไม่เกินค่ายอมรับถือว่าผ่าน

  3. กำหนดเกณฑ์การยอมรับ • 2. อ้างอิงจากมาตรฐาน ต่างๆ เช่น ISO 8655-3 โดยพิจารณาจากค่า error กับเกณฑ์ยอมรับได้ของปริมาตรบรรจุสูงสุดโดยดูเพราะ Max. Systematic error และ Max. Random error ต้องไม่เกินค่ายอมรับได้ • 3. ผู้ใช้งานกำหนดขึ้นเอง

  4. Error of Calibration Error = Systematic Error Standard Deviation = Random Error

More Related