UK Radiotherapy Trials QA (RTTAQ) Group

UK Radiotherapy Trials QA (RTTAQ) Group. SBRT trial QA programmes. GHG meeting - ESTRO May 2017. QA Programme: NHSE CtE (England only) (NHS England Commissioning through Evaluation). UK Consensus on Normal Tissue Dose Constraints for Stereotactic Radiotherapy.

montym
Télécharger la présentation

UK Radiotherapy Trials QA (RTTAQ) Group

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. UK Radiotherapy Trials QA (RTTAQ) Group SBRT trial QA programmes GHG meeting - ESTRO May 2017

  2. QA Programme: NHSE CtE (England only)(NHS England Commissioning through Evaluation)

  3. UK Consensus on Normal Tissue Dose Constraints for Stereotactic Radiotherapy Collaborative work: Trialists, RTTQA and UK SABR consortium Adopted by the NHSE CtE and in clinical use for current and future SABR trials Submitted for publication: UK Consensus on Normal Tissue Dose Constraints for Stereotactic Radiotherapy Hanna GG, Patel R, Aitken A, et al UK stereotactic Ablative Radiotherapy trials normal tissue dose constraints tolerance consensus. 2016; Radiother. Oncol. 119:1; S191

  4. Current and future trials

  5. Trial QA activities

  6. Lung audit • 37centres audited • As required for trial /commissioning programme • Single machine, single energy • Motion management not assessed • Postal audit using CIRS Model 002LFC IMRT thorax phantom • EBT3 GafChromic film - planar dosimetry • Alanine dosimeters – dose at target volume centre • CT dataset of phantom with pre-defined volumes (ITV) sent to centres for planning • Alanine results showed planned and delivered doses were within ±3.0% for 34/37 SABR plans

  7. Spinal audit • SABR Spine dosimetry audit • CIRS E2E SBRT phantom • Dosimetry • Axial GafChromic Film (yellow) • Alanine and MicroDiamond detectors (orange) • Measurement in bone CTV (red) and water equivalent cord

  8. Outlining benchmark • Same cases for all centres • Reviewed against a standard by study CI +/- TMG members • Drop down menus: • No variation (per-protocol) • Acceptable (variation) • Unacceptable (variation)

  9. Planning benchmark • Same cases for all centres • Reviewed against protocol prescription and mandatory and optimal dose constraints by RTTQA

  10. Individual case review • Prospective: • Turnaround time 48-72hrs • Outlines submitted for review prior to plan • Amendments to outlines and plan implemented and re-reviewed before treatment • Pressure to avoid delays to planned treatment start date • Report format same as benchmark case review • Retrospective: • Timely retrospective - within first week of treatment • Standard retrospective - during trial • No direct feedback to centres

  11. What has worked …and what hasn’t! • Audit logistics • Borrowing phantoms • Coordination of audit aspects from multiple centres • Audit limitation • Non standard local plan, harder to provide feedback • No movement • Workload • Many treatment sites • Underestimated number of clinicians • Huge workload for individual centres and RTTQA • Planning aspects more challenging with higher failure rate compared to outlining • Lung case relatively easy as well established. • Liver case hardest to set up – motion management issues. Liver contouring high failure rate. • Audit cost and familiarity (borrowing local phantoms) • Consistency of treatment across all centres • Streamlining - faster trial QA approval • Positive feedback from majority of centres • Enabled inexperienced centres to clinically implement SABR with support • Learning experience for centres with similar equipment • Large data set • An overall view of what is achievable with specific equipment combinations • Facilitate future development of plan metrics

  12. Collaborative work

  13. CORE - International QA