The Dental X-ray Machine

The Dental X-ray Machine. Reference Reading Chapter 2: pp. 15 - 17. X-ray Machine. X-rays are produced within the dental x-ray machine The x-ray machine can be divided into 3 study areas Component Parts The x-ray tube X-ray generating apparatus. Component Parts.

mostyn
Télécharger la présentation

The Dental X-ray Machine

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. The Dental X-ray Machine Reference Reading Chapter 2: pp. 15 - 17

  2. X-ray Machine • X-rays are produced within the dental x-ray machine • The x-ray machine can be divided into 3 study areas • Component Parts • The x-ray tube • X-ray generating apparatus

  3. Component Parts • The Dental x-ray machine consists of 3 visible component parts: • The Control Panel • The Extension Arm • The Tubehead

  4. The Component Parts of the Dental X-ray Machine

  5. The Control Panel • The Control Panel contains: • On-off switch and an indicator light • The exposure button and an indicator light • Control devices to regulate the x-ray beam • Time setting • kVp setting • mA setting

  6. Control Panels Older style Newer Design

  7. The Extension Arm • The extension arm suspends the x-ray tubehead and houses the electrical wires that extend from the control panel to the tubehead • The extension arm allows for movement and positioning of the tubehead

  8. Tubehead • The tubehead is a tightly sealed, heavy metal housing that contains the x-ray tube. The x-ray tube produces the dental x-rays

  9. Tubehead • The component parts of the tubehead include: • Metal housing • Insulating oil • Tubehead seal • X-ray tube • Transformer • Aluminum disk/filter • Lead collimator • Position Indicating Device (P.I.D.)

  10. Component Parts of the Tubehead • The Metal Housing, or body, of the tubehead surrounds the x-ray tube and is filled with oil. • It protects the x-ray tube and grounds the high voltage components. Metal Housing

  11. Component Parts of the Tubehead • Insulating Oil – the oil that surrounds the x-ray tube. • It prevents overheating by absorbing the heat created when x-rays are produced

  12. Component Parts of the Tubehead • Tubehead Seal – the aluminum or leaded glass covering of the tubehead that permits the exit of x-rays from the tubehead. • It seals the oil in the tubehead

  13. Component Parts of the Tubehead • The X-ray tube – the heart of the x-ray generating system.

  14. Component Parts of the Tubehead • Aluminum Disks – sheets of 0.5 mm aluminum placed in the path of the x-ray beam. • This filters out the longer wavelength, non-penetrating x-rays. • Text – p. 15, Fig. 2-14

  15. Component Parts of the Tubehead • Lead Collimator – a lead plate with a central hole that fits directly over the opening of the metal housing where the x-rays exit the tubehead. • Restricts the size of the x-ray beam to 2.75 inches at the patient’s face.

  16. Component Parts of the Tubehead • Position Indicating Device (P.I.D.) – aims and shapes the x-ray beam as it exits the tubehead. • Sometimes called the “cone” Different tubehead lengths

  17. Cathode • Cathode is the negative electrode of the x-ray tube. • The purpose of the electrodes is to supply the electrons necessary to generate x-rays. • The cathode consists of: • Tungsten filament- generates the electrons that produce the x-rays at the tungsten target • Molybdenum cup

  18. Anode • The anode is the positive electrode within the x-ray tube • The purpose of the anode is to convert the electrons into x-ray photons at the tungsten target. • The anode consists of: • The tungsten target • The copper stem