An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. FDA Workshop-External Defibrillators Quality Systems Practices and Adverse Reporting John Collins AHA/ASHE

  2. Nine Hospital Clinical Engineering Program Started in 1997

  3. Risk Assessment Medical Devices • Function • Physical risk • Maintenance

  4. Advent of the Internet FDA database introduced on web site

  5. Revisit Risk Assessment Physical Risk using FDA data

  6. MAUDE RESULTS 1997-2004 DEATH INJURY Anesthesia Unit 1 Blood Cell Processor 1 Dialysis Unit 3 Hypohyperthermia Unit 1 Laser, Excimer 1 Linear accelerator 1 Monitor, Physiological 1 Pacemaker, External 1 Phaecoemulsifier 5 Phototherapy Unit 1 Pump, Athrombic 1 Pump, IABP 3 Treadmill 2 Defibrillator 10 Heart Lung Unit 1 LVAD 1 O2 Regulator 1 Patient Lift 1 Pump, Infusion 2 Telemetry 2 Ventilator 4

  7. Maude Physical Risk Findings • Incidents concern specific models • Incidents for a model occur a single time (not recurrent) • Very few devices repeat with different mfg/model

  8. Comparison of Physical Risk Scores

  9. Maintenance Risks • Originally was manufacturer’s PM schedule • Revised to guess work on device repairs • Five years accumulated work requests for 9 hospitals/>50,000 devices • Revise risk based on data

  10. Comparison of Maintenance Risk Scores

  11. What are the causes of medical device repairs?

  12. Repair History -9 Hospitals-5years>50,000 Devices

  13. Defibrillator Maude Reports

  14. Defibrillator Recalls • 2000 0 • 2001 0 • 2002 0 • 2003 0 • 2004 0 • 2005 5 • 2006 2 • 2007 3 • 2008 1 • 2009 3 • 2010 3

  15. Incident Report FindingsDefibrillators 2008-2009 • malfunction 865 41 • under investigation 500 92 • failure 457 373 • replace 427 674 • shut down 126 15 • fault 103 255 • defective 55 205 • board 37 42

  16. Suggestion • Develop a design so that the device has a constant monitoring process at critical points on the circuit boards with the ability to send messages concerning imminent failure