Doppler Broadening Spectroscopy:

Doppler Broadening Spectroscopy:. Analysis of Rubber Composites. Chun Yang Mentor: Dr. C.A. Quarles. Initial Goal. To confirm non-uniformity found in rubber samples by V.O. Jobando and John Rhoads from previous Lifetime Analysis work.

ouida
Télécharger la présentation

Doppler Broadening Spectroscopy:

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Doppler BroadeningSpectroscopy: Analysis of Rubber Composites Chun Yang Mentor: Dr. C.A. Quarles

  2. Initial Goal • To confirm non-uniformity found in rubber samples by V.O. Jobando and John Rhoads from previous Lifetime Analysis work. • Composite rubber used in industry are possibly not homogeneous. • Positron Annihilation Spectroscopy could be used to find non-uniformities in mixed rubber. • Doppler Broadening was chosen in particular for better analysis of composition differences.

  3. Some Background… • Doppler Broadening PAS involves the momentum distribution of electrons annihilated with positrons. • Greater momentums of annihilation increase Doppler shift in the signal of the photon energy released. • Para-positronium formed have a very low momentum of annihilation. • Increasing carbon will decrease positronium formation. • Fewer low-momentum distributions. • DB signals directly related to uniformity.

  4. Unexpected Challenges…

  5. …Overcome! • S-Parameter fluctuation appeared to be related to dead time.

  6. Rate Dependency • Counting rate, which is dependent on the amplifier settings, is directly affected by the distance of the sample from the detector. • Dead Time is affected by both the number of channels counted and counting rate. • The number of channels counted is dependent on the MCA card settting.

  7. Once Again… • Reexamined SnSSBR rubber with a reasonably stable counting rate.

  8. Uniformity Data • Rubber with differing phr (parts per hundred rubber) of N774 carbon black analyzed.

  9. Concentration Data • Analysis conducted on N774 and N115 samples. • Concentration of Carbon Black affects S-parameter.

  10. Technique Refinement Metal Covering Consider Temperature Look for non-uniformity based on mixing time. Possible application for faster material analysis due to rate dependence. Looking Ahead…

  11. Acknowledgements • I would like to thank: • The NSF for providing the funding for this program (and paying me). • The TCU Department of Physics and Astronomy for their support. • Dr. C.A. Quarles for guiding me on this project.