Tower Driver CERN scope measurement March ‘05 E.Ladygin
Points of measurement HEC quadrant structure E.Ladygin
Signal waveforms : towers – A6,A7,A8,A9 E.Ladygin
Signal saturation E.Ladygin
TDB output noise waveform E.Ladygin
Noise autocorrelation function E.Ladygin
Summed signal E.Ladygin
Conclusion 1 • Calibration signal waveforms should be computed for the time constants and amplification and compared with expected signals E.Ladygin
Conclusion 2 • Calibration delay time distribution has been measured • Maximal difference between delays in a module is about 17ns • Maximal difference between delays in a longitudinal segment is about 9ns (~1.8m) E.Ladygin
Conclusion 3 • Saturated signal has been measured • No surprise E.Ladygin
Conclusion 4 • Noise has been measured – here is seen the big digital pickup noise, mainly 40MHz • Where is from? FEB? Trigger system by the cable to the oscilloscope? E.Ladygin
Conclusion 5 • Has been found that the numbering of mixer channels is not the same on both side of FEB, there is mirror sequence • J.Ban promised that every mixer can be programmed separately • We need a cable check program for trigger chain E.Ladygin