1 / 15

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY

IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology. RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. Part 15.7: Optimization of protection in radiography Practical exercise - processor. Overview / Objectives.

piper
Télécharger la présentation

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION INDIAGNOSTIC ANDINTERVENTIONAL RADIOLOGY Part 15.7: Optimization of protection in radiography Practical exercise - processor

  2. Overview / Objectives • Subject matter : quality control of the processor • Step by step procedure to be followed to implement the considered QC test: • film storage, cassette and screen, sensitometry • Interpretation of results 15.7: Optimization of protection in radiography

  3. IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology Part 15.7: Optimization of protection in radiography Topic 1: Processor evaluation (film storage, cassette and screen, sensitometry)

  4. What Minimum Equipment is Needed? • Thermometer • Sensitometer • Densitometer • Manufacturer’s cleaning solutions • Ultraviolet (“Black Light”) light 15.7: Optimization of protection in radiography

  5. Film Processor QC • Most important QC features : • proper film storage • cassette and screen care • processor chemical care • sensitometry • artifacts • processor cleanliness 15.7: Optimization of protection in radiography

  6. Film Processor QC 15.7: Optimization of protection in radiography

  7. Film Processor QC - Film Storage • Film should be stored in cool, dry conditions < 26° C, 30-60% relative humidity • Film storage area should be adequately shielded from x-ray exposure • Too low humidity allows static discharge • Storage period must not be too long • Stack film boxes vertically to avoid pressure on films 15.7: Optimization of protection in radiography

  8. Cassette and Screen Care • Clean screens periodically to avoid dust shadows • Use manufacturer’s recommended cleaning solutions • An ultraviolet (“Black Light”) light can show up dust 15.7: Optimization of protection in radiography

  9. Processor Chemical Care • Chemicals (developer and fixer) degrade with time and use • Developer in particular will oxidize (go brown) and become depleted with use resulting in poor quality (reduced contrast and film speed) films • Fixer will change pH and lose emulsion hardener 15.7: Optimization of protection in radiography

  10. Film Processor QC 15.7: Optimization of protection in radiography

  11. Processor Chemical Care • Chemicals must be replenished (continual automatic replacement) • Use manufacturer’s recommendations • Check the developer temperature daily - processing is very sensitive to temperature Note: If the number of films processed per day is low, it will be necessary to use flood replenishment. See Flood replenishment: a new method of processor control. Frank ED, Gray JE, Wilken DA. Radiol Technol. 52:271-275, 1980 15.7: Optimization of protection in radiography

  12. Sensitometry • Sensitometer and densitometer required • Essential to keep the process under control • To be performed daily • Main parameters investigated: • Base-plus-fog • mid-density • density difference (See reference to ACR Quality Control Manual, and Quality Control in Diagnostic Imaging, last slide) 15.7: Optimization of protection in radiography

  13. Film Processor QC - Artifacts • Anything on the film which is not related to the x-ray image • Examples : • dust marks, static discharge • fixer stains (poor washing) • film storage problems • processor problems (roller marks, scratches) 15.7: Optimization of protection in radiography

  14. Film Processor QC - Processor Cleanliness • All processors will eventually get dirty • Clean processor according to manufacturer’s instructions • Daily cleaning of entrance trays and cross-over racks is essential 15.7: Optimization of protection in radiography

  15. Where to Get More Information • Part XIII.8 “Optimization of protection for Mammography” Practical exercise • Flood replenishment: a new method of processor control. Frank ED, Gray JE, Wilken DA. Radiol Technol. 52:271-275, 1980 • American College of Radiology Mammography Quality Control Manual, Reston VA, 1999. • Quality Control in Diagnostic Imaging, Gray JE, Winkler NT, Stears J, Frank ED. Available at no cost. http://www.diquad.com/QC%20Book.html 15.7: Optimization of protection in radiography

More Related