Potentiometry and the Detection of Fluoride

Potentiometry and the Detection of Fluoride . pH Meter. pH meters use a ion selective glass electrode: Thin glass membrane connects sample with a reference solution Glass membrane typically made of silica Potential difference builds across membrane

rashad
Télécharger la présentation

Potentiometry and the Detection of Fluoride

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Potentiometry and the Detection of Fluoride

  2. pH Meter • pH meters use a ion selective glass electrode: • Thin glass membrane connects sample with a reference solution • Glass membrane typically made of silica • Potential difference builds across membrane • potential difference depends on ions concentrations on either side of glass membrane • pH meter selective for H+ ions

  3. Glass Membrane Electrode SiO4 Glass Membrane • Conduction is achieved by ion exchange between singly charged cations on the glass lattice with H+ from solution: H+ + Na+Glass- Na+ + H+Glass- • Then between internal solution and glass H+Glass- H+ + Glass- Cation (Na+ or H+) Oxygen Silica

  4. Ion Selective Electrodes • Highly selective for a specific ion • Selective binding to analyte by ion exchange, crystallization, or complexation • Types of Ion Selective Electrodes • Crstyalline Membrane Electrodes • Ex: LaF3 for F- • Non crystalline Membrane Electrodes • Ex: Silicate glass for Na+ and H+

  5. Lanthanum Fluoride Electrode • Crystalline membrane electrode of LaF3, a natural conductor • Doping with EuF2 which has one less F- anion • Vacancies in crystal structure allow for ion “hopping” and thus a conduction of charge • Interference by: • hydroxide ion at pH > 8 Vacancy LaF3 LaF2+ + F- Ion Hopping in a Doped Lanthium Fluoride Electrode F- anions move through vacancies. Eu2+ F- La3+

  6. Selectivity • Electrodes can also be sensitive to ions of the same charge • can cause interference • need to evaluate sensitivity of your technique to other interferents • Activity, not concentration, is directly measured • Total ionic strength of buffer kept constant to minimize ionic effects

  7. Potentiometry • Measure potential difference develops from fluoride ions present • Use potentiometer: high impedance voltmeter that draw virtually no electrical current • Ion concentration can be determined from the measured potential using the Nernst Equation: • Can be used to determine fluoride concentration in drinking water E = E0-(RT/nF)lnQ E is the potential difference E0 is the standard cell potential Q = [ion outside]/[ion inside] R, F are constants T is temperature n is # of e-

  8. Fluoridation • Since 1945, communities across the United States have been adding fluoride (F-) to public water supplies. • Fluoride fights tooth decay • Optimal fluoride level recommended for the prevention of tooth decay: 0.7 - 1.2 parts per million

  9. Fluoridation One of three compounds typically added: Na+-F • Sodium Fluoride: crystalline form easy to handle, but expensive; reference standard • Fluorosilicic acid: liquid by product of phosphate fertilizer manufacture; expensive distribution (shipping) • Sodium Fluorosilicate: powder, fine crystal; easy to ship Fluoridation does not change taste, smell, or appearance of drinking water.

  10. Experimental Results • Calibration curve constructed using public water samples. • Higher concentrations of fluoride are associated with more negative potential • Linear until very low concentrations Fluoride Ion (ppm) Millivolts

  11. Experimental Results • INSERT HUMAN DATA