1 / 17

Geant4 in Brachytherapy

Geant4 in Brachytherapy. Brachytherapy: Brief Overview Clinical applications Basic research Ultrafast & biology applications Issues for the work group. Brachytherapy: Overview. Brachy : Greek word for “short” → Fast (targeted) radiation treatment → Catheter-driven insertion system

rodd
Télécharger la présentation

Geant4 in Brachytherapy

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Geant4 in Brachytherapy Brachytherapy: Brief Overview Clinical applications Basic research Ultrafast & biology applications Issues for the work group G4NAMU @ AAPM, Seattle, WA July 24-28, 2005

  2. Brachytherapy: Overview • Brachy: Greek word for “short” → Fast (targeted) radiation treatment → Catheter-driven insertion system → breast, head, neck, colon, prostate, esophagus, eyes etc. • Typical sources • Gamma & beta emitters • Small: few mm3 • Geometry: cylindrical (Iridium-192, Palladium-103) G4NAMU @ AAPM, Seattle, WA July 24-28, 2005

  3. 3.5 mm 1.1 mm 0.6 mm 5 mm 2 m Modeling of encapsulated 192Ir source(Courtesy of McGill University) MicroSelectron Classic HDR source (part no. 080950) G4NAMU @ AAPM, Seattle, WA July 24-28, 2005

  4. Radial dose and anisotropy functions of encapsulated source(Courtesy of McGill University) Anisotropy function GEANT4: ±1% agreement with Williamson and Li Radial dose function GEANT4: ~0.4% higher than Williamson and Li (Med. Phys. 22, 809-819, 1995) G4NAMU @ AAPM, Seattle, WA July 24-28, 2005

  5. Modeling of rectal applicator(Courtesy of McGill University) • made of silicone rubber • allows for insertion of shielding materials We collect phase space files of photons reaching the applicator surface and use the data for dose calculations in patient geometry. G4NAMU @ AAPM, Seattle, WA July 24-28, 2005

  6. Isodose distributions(Courtesy of McGill University) Presence of lead shielding reduces the dose distribution in a lucite phantom by up to 20%, and it is confirmed by ion chamber measurements. G4NAMU @ AAPM, Seattle, WA July 24-28, 2005

  7. Octree geometry compression – DICOM images Voxel-based geometry • Objectives: • Reduce the number of voxels • Keep the critical information • Indexing of the density distributions using a density gradient threshold: • High resolution kept only in heterogeneous area • Can easily reduce the number of voxels by 75% • Does not affect dose distribution V. Hubert-Tremblay, MSc project G4NAMU @ AAPM, Seattle, WA July 24-28, 2005

  8. Octree geometry compression – DICOM images Examples: CT phantom + thorax patient Voxel number v/s DGT G4NAMU @ AAPM, Seattle, WA July 24-28, 2005

  9. Validation of the model: radial dose fonction Prostate brachytherapy - LDR LDR 125I source, Amersham 6711 J.-F. Carrier, Ph.D. project G4NAMU @ AAPM, Seattle, WA July 24-28, 2005

  10. Prostate brachytherapy - LDR MC simulations on clinical treatment plans RED PLAN • 77 seeds • 0.6 mCi activity • 1.7% of the D90 value is lost due to interseed attenuation This interseed attenuation is not considered in TG43 (clinical) dose calculations 1) Interseed attenuation 1.7% 3.3% BLUE PLAN • 137 seeds • 0.3 mCi activity • 3.3% of the D90 value is lost due to interseed attenuation G4NAMU @ AAPM, Seattle, WA July 24-28, 2005

  11. Prostate brachytherapy - LDR MC simulations on clinical treatment plans 2) Medium composition: CT-based MC G4 DICOM reader[1] Organ contours Seed positions CT image G4 geometry D90: 5-7% differences between water MC and CT-based MC [1] http://www.physmed.phy.ulaval.ca/phys_med/DICOM G4NAMU @ AAPM, Seattle, WA July 24-28, 2005

  12. 32 z 1 z y Hampton University(Collaboration: NIST, EVMS) Scintillating Fiber based Beta Detector → Absolute calibration of sources Lawrence Tynes, Ph.D.: Detector Nnenna Onumah, Ph.D.: Geant4 G4NAMU @ AAPM, Seattle, WA July 24-28, 2005

  13. Hampton University(Collaboration: Proxima Therapeutics, dePaul) Active Mammosite → Absolute real-time position measurement (within ±1 mm) → Absolute real-time dose measurement → Modelization of the dose distribution using CT scan data & 4D phantom Jacquelyn Winston, M.Sc.: Detector Rachel Black, Ph.D.: Detector, Geant4 & VTK G4NAMU @ AAPM, Seattle, WA July 24-28, 2005

  14. Hampton University(Collaboration: Jefferson Lab, EVMS, ODU) Breast cancer cells Collimator and Sr-90 source Dipole Monoenergetic Brachytherapy Sources → Energy dependence of irradiated breast cancer cells Ariano Munden, M.Sc. G4NAMU @ AAPM, Seattle, WA July 24-28, 2005

  15. ALLS FacilityUltrafast Laser & Biomedical Research 10 Hz, 300 mJ, 25 fs Advanced Laser Light Source • Varennes, Canada • Dynamics of biological processes 100 Hz, 100 mJ, 25 fs G4NAMU @ AAPM, Seattle, WA July 24-28, 2005

  16. Issues for the Working Group • Physics • Electromagnetic shower • Multiple scattering • Experimental cross sections (JLab data, others) • Theory (?) • Treatment verification • Bench marks • Examples • Micro-dosimmetry • Simulation at molecular level • Mono-energetic dose distribution • Other topics • Group organization • Therapy Physics (I. Chetty, MC in Treatment Planning) • Where to go from here? • To be discussed!!! G4NAMU @ AAPM, Seattle, WA July 24-28, 2005

  17. Prostate brachytherapy - HDR HDR 192Ir source, MicroSelectron Preliminary results Ejection spectrum keV A. Allard, Summer Project G4NAMU @ AAPM, Seattle, WA July 24-28, 2005