An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. MICE Decay Solenoid Quench Detector Trevor Hartnett Daresbury Laboratory

  2. Outline • Existing (PSI) Quench detection system • How it works • Why we want a replacement • Replacement (DL) Quench detection system • Comparison with existing system • How existing system could meet other requirements • Progress report • When it will be ready

  3. Existing (PSI) system • System overview • 3 Crates • Quench Detector • Power Supply • Battery back-up

  4. Quench detection 7 Amplifier cards 7 Relay cards Latch and LED driver card Test points

  5. How a quench is detected

  6. Power Supply Linear power supply Switch mode ±15V and 12V power supply Circuit breaker control relays Emergency backup battery switching

  7. Battery Backup • 4 12V Lead acid batteries • Series connected to provide 48V • Capacity 7Ahrs • Charged by a 1A, 48 V battery charger • Powered by the incoming mains. • LEDs used to indicate: • Batteries charged • Mains connected

  8. Recent Problems • System is at least 27 years old • Concern that the electronic components are approaching end of life • sticking relay • capacitor failure in the magnet PSU which is of a similar age • ECS Group at DL asked to provide a Replacement QD system.

  9. Replacement (DL) System • Follow original principles of PSI design • Update components • Use a modular design • Single PCB per quench channel • At least existing system user interface

  10. Quench detection cards • Modular Design • Updated components • Isolation amplifier • Relay • Latching circuit • Isolation of voltage taps • Test points on each card

  11. Power supply • 2x 230Vac to 48Vdc PSUs • 2x 48Vdc to ±15Vdc PSUs • 2x 48Vdc to 12Vdc PSUs • Control Relays

  12. Battery Backup • The battery back-up crate will follow the existing design closely. • A 48V, 1A charger will be used to charge four 12V batteries in series. • The batteries will be 7 Ahr units.

  13. Upgrades • Variable voltage limits on quench detection • Voltage range from 10mV to 110mV • A relay off indicator LED on each card • A reset button on each card • Test points are provided on each card • Greater level of output diagnostics • Tripped channel can now be identified • Status indicators of each power supply • Increased EMC protection to prevent nuisance trips

  14. Progress • Quench Detection Cards tested • Power supply and battery backup still to be completed • Final testing of the complete system to be carried out before the end of July • Test of system with AFC in R9 at the end of July

  15. Questions?