An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. A CMOS Imager for DNA Detection Samir Parikh MASc Thesis Defense Dept. of Electrical and Computer Engineering University of Toronto 24th January, 2007

  2. Outline • Introduction • Motivation and Objectives • Design Details • Experimental Results • Conclusion • Future Work

  3. Introduction: DNA Microarrays • DNA microarrays used to detect DNA sequence concentration Chemical Processing DNA ssDNA Fragments • DNA is split into its two constituent strands • One strand is broken into fragments

  4. Introduction: Using DNA Microarrays • Within a spot multiple identical ssDNA probes are attached • Each spot is tailored to match with a particular target ssDNA sequence • target ssDNA is created from Messenger RNA extracted from a cell

  5. Introduction: DNA Detection • Solution containing target ssDNA+fluorescing dye molecule is introduced to the slide • Spots on the DNA microarray pair/unpair depending on the nucleotide sequence of the probe and target ssDNA • DNA microarray is washed to remove unpaired target ssDNA

  6. Introduction: DNA Detection • Solution containing target ssDNA+fluorescing dye molecule is introduced to the slide • Spots on the DNA microarray pair/unpair depending on the nucleotide sequence of the probe and target ssDNA • DNA microarray is washed to remove unpaired target ssDNA

  7. Introduction: Basic Microarray Scanner • Fluorescing dye molecule absorbs energy at λ1nm and emits energy at λ2nm • Light detectors are discussed in the next slide

  8. Introduction: Existing Light Detectors • Commonly used detectors in microarray scanners are: • Photomultiplier Tube (PMT) - accurate • Charge-Coupled Device (CCD) - fast

  9. Motivation and Objectives • Determine the feasibility of using standard CMOS technology for light detection and quantification • Integrated • Smaller • Cheaper • Validate the design without the use of cooling • Reduce cost related to cooling • Reduce power consumption due to cooling equip.

  10. Design Details: Microarray Scanner • Signal from entire spot captured at once

  11. Design Details: Microarray Scanner • Signal from entire spot captured at once

  12. Design Details: Microarray Scanner • Signal from entire spot captured at once

  13. Design Details: Microarray Scanner • Signal from entire spot captured at once

  14. Design Details: Microarray Scanner • Signal from entire spot captured at once

  15. Design Details: Active Pixel Sensor (APS) • 5-transistor circuit with pseudo-differential output • Pinned photodiode performs the photon-to-electron conversion • Circuits has two phases: reset and integration photons

  16. Design Details: ΔΣ Modulator • 2nd Order Discrete-Time ΔΣ • Can be combined with a decimation filter for a complete ADC • Boser-Wooley Architecture • Delaying Integrators with 1bit feedback • Folded-Cascode Op-amp used

  17. Design Details: Fabricated Chip

  18. Experimental Results: APS • Dark signal limits the integration time for the APS • Low light sensitivity sets the min # of photons detectable

  19. Experimental Results: ΔΣ Modulator • Simulation includes flicker and thermal noise • Close matching between simulation and measured

  20. Experimental Results: ΔΣ Modulator • Commercial microarray scanners have 12 to 16-bits accuracy • Sampling rate sets an upper limit on the maximum light level • Sampling rate not critical, minimum light level is more important

  21. Experimental Results: Microarray Scanner Setup

  22. Experimental Results: Microarray Scanner Setup

  23. Experimental Results: Microarray Scanner Setup

  24. Experimental Results: Microarray Scanner Setup

  25. Experimental Results: Scanner Characterization Slide • Slide contains spots with dilution series • Each spot contains fluorescing dye molecules with fixed density • Spot density (fluorophores/um2) decreases at a fixed rate Decreasing dye density

  26. Experimental Results: Microarray Scanner

  27. Experimental Results: Commercial Microarray Scanner

  28. Discussion: Microarray Scanner • Portability Potential • Microarray scanner: Smaller, integrated detector w/o cooling • Agilent scanner: PMT • Detection Limit • Microarray scanner: 4590 fluorophores/um2 • Agilent scanner: 4 fluorophores/um2 • Resolution and Scan time • Microarray scanner: Larger pixel→Entire spot imaged at once • Agilent scanner: 10μm resolution→takes longer to image a spot • Microarray scanner: Multiple pixels → short scan time • Agilent scanner: Single element → long scan time (8 min/slide)

  29. Discussion: APS • Dark signal of APS not the limiting factor • Background of the slide = 1.5 ADU/sample • Dark signal of the APS = 0.08 ADU/sample • Integration time of the APS is limited by the slide background • Improve the sensitivity of the APS beyond 2.6х10-2 lux • Increase its conversion gain • Reduce its read noise and reset noise

  30. Discussion:Optical and Mechanical • Improve optical coupling between APS ↔ fluorescing spots • Use a focusing/collimating element • Compensate for slide tilt • Reduce laser noise and drift from 3% to 0.1% • Improved power supply • Better laser control/feedback

  31. Conclusion • Standard CMOS technology shows potential to be an alternative to existing PMT/CCD detectors used in microarray scanners • The detection limit of a microarray scanner is determined by: • Mechanical and Optical Non-idealities • Detector Non-idealities

  32. Future Work • Improve the conversion gain of the APS • Reduce the read noise, and reset noise of the APS • Improve the accuracy of the ADC

  33. Thank You