1 / 13

Conditioning of MWPCs for the LHCb Muon System

Conditioning of MWPCs for the LHCb Muon System. Katharina Mair for J.-S. Graulich, H.-J. Hilke, A. Kachtchouk, K. Mair, B. Schmidt, T. Schneider LHCb Muon CERN, Switzerland. Contents: MWPCs in the LHCb Muon System MWPC Conditioning Conclusion. Muon Detectors. Calorimeters. Tracker.

tevin
Télécharger la présentation

Conditioning of MWPCs for the LHCb Muon System

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Conditioning of MWPCs for the LHCb Muon System Katharina Mair for J.-S. Graulich, H.-J. Hilke, A. Kachtchouk, K. Mair, B. Schmidt, T. Schneider LHCb Muon CERN, Switzerland • Contents: • MWPCs in the LHCb Muon System • MWPC Conditioning • Conclusion

  2. Muon Detectors Calorimeters Tracker RICH-2 Iron Filters Magnet RICH-1 Vertex Locator 5 Muon Stations Chamber Dimension MWPCs in the LHCb Muon System • Multi Wire Proportional Chambers (MWPCs): • Fast muon triggering • Muon identification • 5 Muon Stations, 4 Regions / Station • 20 different chamber sizes • 1368 chambers LHCb Muon CERN

  3. MWPC Sandwich MWPC Design 4-gap MWPC gap size: 5 mm (wire plane centered) gas mixture: Ar/CO2/CF4 (40:55:5) wire: Gold-plated Tungsten, 30 μm Ø, 250 to 310 mm wire length wire spacing: 2 mm, mechanical tension: 65 gr HV = 2.650 kV field on wires: 262 kV/cm field on cathodes 6.2 kV/cm gas gain: G ≈ 50 000 gain uniformity: ≤ 30% • panel production: • PCB coated by 35 μm copper, 5 μm nickel, 0.2 μm gold • foam injected between 2 PCBs in mould LHCb Muon CERN

  4. time resolution MWPC Requirements • fully efficient and robust level-0 high pt muon trigger: • 5-fold coincidence of at least 1 hit within 20ns in each station • high efficiency and time resolution: • high efficiency (>99%) per station requires a time resolution of ~4ns • achieved by use of 2-fold OR (double-gaps) • high rate capability: • up to 0.2 MHz/cm2for inner chambers at a luminosity of L = 5 x 1032cm-1s • ageing resistance over 10 years: • the accumulated charge < 1C/cm • spatial resolution: • we require a pt resolution < 20%, therefore we need a spatial resolution in the order of 1 cm in bending plane (6-30 mm) LHCb Muon CERN

  5. impurities on wire Chamber Conditioning • Motivation for Conditioning: • by stepwise applying positive HV on anode wire for the first time, we have difficulties to reach operating point within safe current region < 50 nA • we assume impurities on the anode wire and on the flat cathode surface responsible for self-sustaining discharge under positive HV • Conditioning Procedure: • Step 1: Inversed HV-Conditioning: negative HV up to -2300 V at anode wire • Step 2: Normal HV-Conditioning: positive HV up to +2900 V at anode wire General rule: In order not to damage any surfaces (wire and cathode) high currents must be avoided ! (nominal dark current: ~ 3 nA) • Cleaning Effects due to: • positive ion bombardment on cathode • negative ion and electron bombardment on anode • Results: • surfaces smoothened • current reduction LHCb Muon CERN

  6. time 10 min result: current reduction Step 1:Negative HV Conditioning • Procedure: • applying negative HV to wire • stepwise increasing HV up to -2300 V • Effect: • possibly electron emission from metallic tips on wire surface • positive ion bombardment on wire surface → E-field reduction due to wire surface smoothening • Advantage: • safe procedure: weak avalanche effect at this HV range ( at -2300 V: gas gain ~100) • Results: • at each HV step: fast current reduction (10 μA->100 nA) within 10 minutes • mean conditioning time at -2300 V: 45 hours • after this: positive HV = 2900 V is quickly reached (within 15 minutes) with a dark current level of 3 nA LHCb Muon CERN

  7. 8 gaps SOURCE=ON SOURCE=OFF time Step2: Positive HV Conditioning • Motivation: High Rate Capability of chambers is tested at GIF • all chambers of inner regions (highest particle flux up to 0.2 MHz/cm2) are exposed to high gamma radiation at the Gamma Irradiation Facility (GIF) at CERN for testing. • 137Cs source (photons 660 keV) similar to LHC background radiation • Outcome: • in ~ 20% of all tested gaps Malter-like emission detected • Malter effect: Thin Film Field Emission(L. Malter, Phys. Rev. 50, 1936) • thin insulator film on cathode surfaces charged up by GIF irradiation • high electric field starts e--emission (from: J. Va’vra, NIM A367, 1995) LHCb Muon CERN

  8. Time const = 5 hours • Chemistry: • cathode panel production: mould release agent ACMOIL36-4600 contains long C-chains (Isoalcine C9-C12) with silicone (5-10%) → indication for a remaining insulating film on surface • panels are cleaned by hand (Isopropyl alcohol, 4-Methyl-Pentanol, n-Hexane, demineralized water), problematic region for dirt: between cathode pads • Treatment at GIF: • positive HV applied in steps of 50-100 V for the range of 2.2 kV – 2.75 kV • high gamma ray irradiation leads to charge up of insulator spots → e--emission → positive ions • F-radicals are created due to 5% CF4 in gas mixture • remove Si by creating SiF4 molecules, that are volatile and will be removed by gas flux • Results: • exponential current reduction • conditioning time: 0.5 – 70 hours • Draw Back: • molecular bonds may recover when irradiation or HV removed → currents return, but decay is faster LHCb Muon CERN

  9. 1.000 1 0.900 I/I_initial I/I_initial 0.800 0.8 0.700 0.600 0.6 0.500 0.4 0.400 0.300 0.2 0.200 0.100 0 0.000 0 10 20 30 40 50 60 70 80 90 100 time [h] time [h] 0 20 40 60 80 current [nA] current [nA] same chamber 10 min 10 min time time • Test with 5% CF4: • gas mixture: Ar/CO2/CF4(40:55:5); HV=2.75 kV • current reduction faster • remaining current level excellent [~2 nA] • no HV trips • Test without CF4: • gas mixture: Ar/CO2(40:60); HV=2.75 kV • current reduction slow • remaining current level still too high [μA] • HV trips observed LHCb Muon CERN

  10. R Anode (Gold-plated Tungsten wires) Electrons I1(t) +HV E0 Positive ions Electron emission Thin film between cathode pads EActual=E0+EFilm Cathode pad (Gold-coated Copper foil) y= =x • Emission Physics: • first attempts to combine Malter-effect with field emission (Fowler-Nordheim) can be found in literature (A. Boyarski, NIM A535, 2004) • For our MWPCs: high currents most probably due to 2 effects: • electron emission from metallic tips on surfaces • thin film field emission → most likely from area between cathode pads • For our measurements: • measure I, V • include gas gain G(V) depending on V • approximation: ß… field enhancement factor E… electric field in [V/m] for a gold-coated wire → field factor ß*estimated from the slope LHCb Muon CERN

  11. logarithmic scale: The original I(V) relation 5000 4500 4000 current [nA] 3500 3000 2500 2000 1500 1000 500 0 2500 2550 2600 2650 2700 2750 2800 HV (V) log(I / V2) [nA / V] I / V2 M3R2#04 (Gap2) 1/V [V-1] • Measurments: • repeated measurements on the same chamber show rotation of straight lines with increasing slopes • Result: • E-field decreasing → surface cleaned • conditioning time: up to 500 hours (20 days) LHCb Muon CERN

  12. Summary • MWPC Conditioning in 2 steps: • Step1: negative HV on wire • apply negative HV up to 2300 V • check current at positive HV=2900 V: • If I < 3 nA → Step 2 • If I > 3 nA → repeat Step1 • Step 2: positive HV on wire under High Gamma Ray Irradiation • stepwise increase positive HV on wire for the range of 2.2 kV – 2.75 kV • at each step of 50-100 V switch source off to check current: • If I < 10 μA → increase HV • If I > 10 μA → repeat irradiation at same HV LHCb Muon CERN

  13. Conclusion • Achieved excellent conditioning result by applying 2 steps: • Inversed HV Conditioning • Normal HV Conditioning under High Gamma Ray Irradiation • Effects are: • successful wire cleaning • successful cathode cleaning → We can assume that the observed anomalous currents are self-suppressed during MWPC operation (high background radiation) → Therefore we are optimistic, that these currents will not be a problem for the long term operation of the LHCb Muon System LHCb Muon CERN

More Related