An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Demonstration of a Fast-neutron Detector Ray Bunker—UCSB HEP DUSEL AARM Collaboration Meeting

  2. The Neutron Detector Collaboration Dan Akerib Mike Dragowsky Chang Lee Mani Tripathi Melinda Sweany Harry Nelson Susanne Kyre& Dean White Ray Bunker Carsten Quinlan Raul Hennings-Yeomans Joel Sander Prisca Cushman Jim Beaty with support from the NSF DUSEL R&D program & thanks to the Department of Natural Resources & the staff of the Soudan Underground Laboratory!

  3. A Fast-neutron Detector—The Signal Design based on Hennings & Akerib, NIM A 574 (2007) 89-97 Light-tight Enclosure High-energy Neutron 20” Hamamatsu PMT 2” Top Lead Shield 2” Side Lead Shield ~2.2 Metric Ton Water Tank       Capture on Gadolinium 8 MeV Gamma Cascades Over 10’s of s     20 Ton Lead Target   Liberated Neutrons Hadronic Shower

  4. A Fast-neutron Detector—The Signal 100 MeV Neutron Beam Expected Number of sub-10 MeV Detectable Secondary Neutrons Detector Outline Sitting atop Pb Target FLUKA-simulated Hadronic Shower & Neutron Production by Raul Hennings-Yeomans Ray Bunker-UCSB HEP

  5. A Fast-neutron Detector—Signal Event Clustered Pulse Train Relatively Large Coincident Pulse Heights Ray Bunker-UCSB HEP

  6. A Fast-neutron Detector—Principle Background     Accidentally Coincident U/Th Gammas 2.6 MeV Endpoint 

  7. A Fast-neutron Detector—Background Event Relatively Small Coincident Pulse Heights South Tank PMT Signals Usually Spread Between Tanks Truly Random Timing North Tank PMT Signals

  8. A Fast-neutron Detector—Signal vs. Background • Primary Discriminator Based on Pulse Height • U/Th gammas < ~50 mV • Gd capture gammas > ~50 mV • Additional Discrimination Based on Pulse Timing • ~½ kHz U/Th gammas •  characteristic time ~2 ms • Gd capture time depends on • concentration •  characteristic time ~10 s • Gd captures cluster toward • beginning of event: Measured U/Th Response Gd Capture Response Calibrated with 252Cf Fission Neutrons South Tank 0.2% Gd North Tank 0.4% Gd Ray Bunker-UCSB HEP

  9. A Fast-neutron Detector—Signal vs. Background More Neutron Like More Gamma Like Background U/Th Gamma Rays 252Cf Fission Neutrons Pulse timing Likelihood Pulse Height Likelihood Ray Bunker-UCSB HEP

  10. A Fast-neutron Detector—GEANT4 Optical Properties • Muons are an excellent source of Cherenkov photons—illuminate entire detector • Use to tune MC optical properties for: • Water • Amino-g wavelength shifter • Scinteredhalon reflective panels Combination of Muon Spectral Shape & West-East Pulse Height Asymmetry Used to Break Degeneracy of Reflector’s Optical Properties Backup slides—ask me later if interested ~150 MeV Muon Peak 95% Diffuse + 5% Specular Spike for Best Agreement with Data 94% Total Reflectivity for Best Agreement with Data Event rate (arbitrary units) Pulse height (V) Stopping Muon Decay e 50 MeV Endpoint Ray Bunker-UCSB HEP

  11. A Fast-neutron Detector—Simulated Neutron Response • Estimated 252Cf Fission Neutrons: • Monoenergetic 5 MeV neutrons • Multiplicity pulled from Gaussian centered at 3.87 ( of 1.6) Single Neutron Capture Response Monte Carlo—Solid Black Data—Shaded Red 2.5 mV/photoelectron Scaling Required to Match MC to Data Implies ~½ MeV Detection Threshold Event rate (normalized) Pulse height (mV) Ray Bunker-UCSB HEP

  12. A Fast-neutron Detector—Simulated Gamma-ray Response • 1.17 & 1.33 MeV gammas from 60Co (often observe both simultaneously) • 2.5 mV/PE 252Cf scaling applied • Additional resolution required for agreement •  Gaussian smear with energy-dependent width,  ~ 0.9*sqrt(pulse height) Monte Carlo—Solid Black Data—Shaded Red Event rate (normalized) Pulse height (mV)

  13. A Fast-neutron Detector—Simulated U/Th Background Response • Throw Ruddick spectrum from cavern walls • Apply scaling and energy-dependent smearing indicated by 252Cf and 60Co • Ruddick spectrum is softer than observed data • Enhancing 2.6 MeV endpoint resolves discrepancy •  Implies that cavern/materials near detector have 40% more Thorium in U/Th ratio Event rate (normalized) Keith Ruddick 1996-NuMI-L-210 Monte Carlo—Solid Black Data—Shaded Red Pulse height (mV) Gammas/second/sq.m GEANT Event rate (normalized) Monte Carlo—Black Data—Red Pulse height (mV) Gamma energy (MeV) Ray Bunker-UCSB HEP

  14. A Fast-neutron Detector—Concluding Remarks • Constructed a water Cherenkov, Gd-loaded high-energy neutron detector • Response to U/Th & 60Co gammas, muons, and 252Cf fission neutrons understood via GEANT4 • Demonstrated ability to separate signal from background • Have operated in Soudan Mine since November 2009... calibration + neutron-search data • Rough analysis of search data shows a clear excess of high-multiplicity events! • Goals: • Absolute flux measurement & Monte Carlo Benchmarking: MCNP, FLUKA, GEANT4, … • Unfold energy spectrum from multiplicity distribution Background Signal Ray Bunker-UCSB HEP

  15. A Fast-neutron Detector—Multiplicity = Energy? FLUKA Demonstration of Secondary Neutron Multiplicity Dependence on Energy of Primary Raul Hennings-Yeomans Underground Neutron Flux Mei & Hime Phys. Rev. D73 (2006) ? ?

  16. A Fast-neutron Detector—Multiplicity 27 Candidate Event # 2314 from 2nd Fast-neutron Run: South Tank PMT Traces — Channel 1—South East PMT Pulse height (volts) — Channel 2—South West PMT Triggering Pulses Ray Bunker-UCSB HEP

  17. A Fast-neutron Detector—Installation Electronics Rack Source Tubes Ray Bunker-UCSB HEP

  18. A Fast-neutron Detector—Installation Cheap Labor Water Tanks Ray Bunker-UCSB HEP

  19. A Fast-neutron Detector—Installation 20” KamLAND Phototubes Ray Bunker-UCSB HEP

  20. A Fast-neutron Detector—Muon Response • Large dE/dx events (>80% of all recorded events) • Large initial pulse with prominent after pulsing • Large individual channel multiplicities, but few coincidences Ray Bunker-UCSB HEP

  21. A Fast-neutron Detector—GEANT4 Optical Properties of Water • Water absorption and refractive index taken from LUXSim package: • Refraction  The equation for the refractive index is evaluated by D. T. Huibers, 'Models for the wavelength dependence of the index • of refraction of water', Applied Optics 36 (1997) p.3785. The original equation comes from X. Qua and E. S. Fry, 'Empirical • equation for the index of refraction of seawater", Applied Optics 34 (1995) p.3477. • Absorption: • 200-320 nm: T.I. Quickenden& J.A. Irvin, 'The ultraviolet absorption spectrum of liquid water', J. Chem. Phys. 72(8) (1980) p4416. • 330 nm: A rough average between 320 and 340 nm. Very subjective. • 340-370 nm: F.M. Sogandares and E.S. Fry, 'Absorption spectrum (340-640 nm) of pure water. Photothermal measurements', • Applied Optics 36 (1997) p.8699. • 380-720 nm: R.M. Pope and E.S. Fry, 'Absorption spectrum (380-700 nm) of pure water. II. Integrating cavity measurements', • Applied Optics 36 (1997) p.8710.

  22. A Fast-neutron Detector—GEANT4 Optical Properties 20” KamLANDPhototubes (~17” photocathode) ~20% Peak Quantum Efficiency Amino-g Wavelength Shifter Absorbs UV, Emits Blue (most Cherenkov photons are UV) >2 Increase in Light Yield Ray Bunker-UCSB HEP