Evaluation of phakic irisfixated intraocular lenses with Zeiss Visante OCT

International Vision Correction Research Centre. E-mail:angela.ehmer@med.uni-heidelbergd.de www.lasik-uni-hd.de. Evaluation of phakic irisfixated intraocular lenses with Zeiss Visante OCT. A. Ehmer, C. Foerster, M.P. Holzer, G.U. Auffarth.

tracen
Télécharger la présentation

Evaluation of phakic irisfixated intraocular lenses with Zeiss Visante OCT

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. International Vision Correction Research Centre E-mail:angela.ehmer@med.uni-heidelbergd.de www.lasik-uni-hd.de Evaluation of phakic irisfixated intraocular lenses with Zeiss Visante OCT A. Ehmer, C. Foerster, M.P. Holzer, G.U. Auffarth The authors have no financial interests in any of the products mentioned. Dept. of Ophthalmology Ruprecht-Karls-University Heidelberg Acting-chairman: G. U. Auffarth, MD

  2. Purpose, Methods & Patients Purpose Visualisation and detection of lens position of phakic intraocular lenses in the anterior chamber of the human eye. Methods • Phakic irisfixated IOLs (Verisyse, AMO) were implanted to correct ametropia (Figure 1) • Lens position in the anterior chamber was measured with Zeiss Visante OCT (Figure 2) • ACD was measured with Zeiss IOL Master (Figure 3) and Oculus Pentacam HR (Figure 4) • Data were collected 4 to 26 months postoperatively Patients 24 eyes of 12 patients, mean age of 38.1 ± 5.9 years Figure 1 Figure 3 Figure 2 Figure 4

  3. Results I ACD preop in mm Distance cornea – pIOL in mm Figure 5 shows the pIOL in the anterior chamber. The distance from pIOL to corneal endothelium magnified in Figure 6 and pIOL to cristalline lens in Figure 7. Figure 7 Figure 6 Figure 5 A high correlation between the preoperative anterior chamber depth and the distance of the phakic IOL to corneal endothelium (r = 0.83) was found (figure 8). The deeper the anterior chamber the greater the distance between cornea and phakic IOL. Figure 8

  4. Results II Figure 10 Figure 9 • The mean distance between the phakic IOL and the endothelium of the cornea was 2.21 ± 0.31 mm (figure 9) • The mean distance between the natural crystalline lens to the phakic IOL was 0.80 ± 0.18 mm (figure 10) • High correlation was found between target refraction and refraction three months postoperatively (R = 0.98) (figure11) Figure 11

  5. Conclusion The Zeiss Visante OCT is suitable to demonstrate IOL position in the anterior chamber of the eye. The data analysed showed a confident implantation without impairing the corneal endothelium over a 26 months follow up period. • Literatur • Konstantopoulos A, Hossain P, Anderson DF.Recent advances in ophthalmic anterior segment imaging: a new era for ophthalmic diagnosis? Br J Ophthalmol. 2007 Apr;91(4):551-7. Review. • Lavanya R, Teo L, Friedman DS, Aung HT, Baskaran M, Gao H, Alfred T, Seah SK, Kashiwagi K, Foster PJ, Aung T: Comparison of anterior chamber depth measurements using the IOLMaster, Scanning Peripheral Anterior Chamber depth Analyser and Anterior Segment Optical Coherence Tomography. Br J Ophthalmol. 2007 Feb 27; • Rabsilber TM, Khoramnia R, Auffarth GU.Anterior chamber measurements using Pentacam rotating Scheimpflug camera.J Cataract Refract Surg. 2006 Mar;32(3):456-9. • Wang L, Auffarth GU. White-to-white corneal diameter measurements using the eyemetrics program of the Orbscan topography system. Dev Ophthalmol. 2002;34:141-6.