1 / 75

A-dec 300 Dual Mount on 500

A-dec 300 Dual Mount on 500. Installation. Agenda. General Information Recommended tools Installation Follow through install guide Hands-on install Adjustments Leveling. Learning Objectives. As a result of this training you will be able to:

trula
Télécharger la présentation

A-dec 300 Dual Mount on 500

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. A-dec 300 Dual Mount on 500 Installation

  2. Agenda • General Information • Recommended tools • Installation • Follow through install guide • Hands-on install • Adjustments • Leveling

  3. Learning Objectives • As a result of this training you will be able to: • Identify the different components and options of an A-dec 300 dual mount system on a 511 chair • Complete a full installation • Demonstrate how to use the installation guide • Make adjustments and level

  4. Tools • Recommended Tools: • Sizes are all SAE, no metric • Hex Key Set (inch) • Phillips & standard screw drivers • Socket wrench set (SAE) • Snap ring tool • Adjustable wrench • Needle nose pliers • Combination wrench set (SAE) • Standard pliers • Diagonal cutters • Torpedo Level

  5. 511 Chair Box 1 Move the chair from the pallet • Chair and Footswitch

  6. A-dec 511 Chair Installation • Plug in the chair and turn on • Use the three-tap feature to raise the chair • Remove the shipping pin

  7. A-dec 511 Chair Installation • Anchor the chair to the floor

  8. Anchor the Chair To Concrete To Wood Drill a 2” (50 mm) deep hole. Use a 3/8” (0.375 mm) drill. Remove any debris Place the washer on the anchor bolt and insert it into the hole. Tighten the anchor bolt until it is tight against the washer and firmly holds the baseplate to the floor. • Drill a 4” (101 mm) deep hole. Drill through the chair anchoring hole into the concrete. Use a roto-hammer drill. • Remove any debris • Drive the anchor bolt into the hole until the bolt head seats against the chair base. • Tighten the bolt until it is tight against the washer and firmly holds the baseplate to the floor.

  9. Install the Footswitch • Footswitch is optional. If the chair was ordered with a footswitch, install it next.

  10. A-dec 511 Chair Installation • Install the back support and headrest

  11. 300 Dual mount Installation

  12. 300 Dual Mount Installation • Lower the back • Install the front mount Screw the leveling bolts (from the kit) into the chair frame, leaving two threads exposed

  13. 300 Dual mount Installation • Hang the front mount on the leveling cam

  14. 300 Dual Mount Installation • Screw the nuts on the leveling bolts until they are finger tight

  15. Install the Rigid Arms Dental Light or Monitor Mount Dual Rigid Arms Delivery System

  16. Delivery System Only 2 1

  17. Dental Light or Monitor Mount Only 1

  18. Installing the Rigid Arms • Tighten the nut until the arm can move smoothly but is not loose and does not drift

  19. Installing Dual Rigid Arms Place the spacer with its solid side up and its opening toward the hydraulic tubing • Install the spacer

  20. Installing Dual Rigid Arms • Install the toeboard ramps

  21. Installing Dual Rigid Arms 2 1

  22. Level the Rigid Arms

  23. Leveling the Rigid Arms Leveling Bolts

  24. Leveling • Fully raise the chair back Place level so it is side to side on the mount

  25. Leveling

  26. Leveling Tighten the stabilizing bolts on each side of the mount until they touch the chair casting Tighten the nuts on the leveling bolts on both sides of the front mount

  27. Delivery Installation A-dec 333 Radius-Style Continental Delivery System on an A-dec 511 chair A-dec 332 Radius-Style Traditional Delivery System on an A-dec 511 chair

  28. Delivery Installation • Install the water bottle

  29. Delivery Installation • Install the flexarm and control head

  30. Delivery Installation • Control head Loosen the button head screw Rotate the control head so it is positioned over the hole for the missing leveling screw Install and tighten 1/4 – 20 x ½” leveling screw Install the 3/8” setscrew

  31. Continental Delivery • Insert whip hooks

  32. Delivery Installation • Route tubing and wiring Umbilical Wrap

  33. Delivery Installation • Route the tubing and wiring Remove the tape from the tubing

  34. Continental Tray Holder • Optional

  35. Traditional Tray Holder • Optional

  36. Dental Light Fully lower the chair base Lubricate the end of the upper rigid arm Route the light cable through the lower rigid arm Insert the upper rigid arm into the lower rigid arm until fully seated Installation is identical for the A-dec 372 or A-dec 572 Dental Light. The A-dec 372 is shown here.

  37. Dental Light Route the cable Install the umbilical wrap

  38. Monitor Mount Route the power and video cables through the mount and down through the rigid arm Install the rotation stop screw and spacer Slide the monitor mount onto the rigid arm

  39. Monitor Mount • Install protective sheath over the monitor cables and secure with cable tie

  40. Monitor Mount • Install the umbilical wrap and route the cables Route the cables Umbilical wrap

  41. Monitor Mount • Optional If the system includes a 100mm VESA standard monitor mounting pattern, use the four screws and nuts to install the adaptor plate

  42. Connections • Connect the electrical wires and data lines Secure the wiring to the bail with cable ties Plug in the red connector Connect the black wire to 0 VAC and the gray wire to 24 VAC Connect the ground wire

  43. Connections • Tubing

  44. Secure Tubing and Install Water Bottle

  45. Adjustments • Adjust delivery system spring assisted flexarm

  46. Adjustments • Install the delivery system flexarm rotation stop screw

  47. Adjustments • Adjust the control head rotation tension To increase tension, turn the key clockwise. To decrease tension, turn the key counterclockwise.

  48. Adjustments • Adjust the tray rotation tension (Traditional)

  49. Adjustments • Adjust the tray rotation tension (Continental)

  50. Adjustments • Adjust the handpieces • Remove the back cover • Turn the air coolant, water coolant, and drive air controls clockwise to turn them all the way down • Lift a handpiece from the holder • Flip the toggle to water • Step on the foot control • Adjust the water coolant flow until there is 2 drops a second • Step on the foot control • Adjust the air coolant flow until the spray is a fine mist (to increase flow turn the key counterclockwise. To decrease the flow turn the key clockwise)