An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. QUALITY PATIENT CARE THROUGH PROFICIENCY TESTING Not Just Talking Quality, But Investing in QualityNaval Medical Center San DiegoLaboratory DepartmentQuality Assurance Unit

  2. What I really need to know besides my “A,B,Cs”? QA QC POCT PT

  3. OBJECTIVES • Discuss regulatory aspects and impact of proficiency testing (PT) • Approaches to effectively manage and fully utilize PT • List investigative techniques used to determine causes of poor PT performance and describe effective corrective actions • Understand how “Linearity” relate to the accreditation process • Explain techniques to perform the appropriate verifications of AMR and CRR • Common Analytical Exception codes

  4. CAP Proficiency Testing • Allows laboratories to evaluate their performance and improve the accuracy of the patient results • Results evaluated using comparable peer groups from comprehensive database of laboratories

  5. PROFICIENCY TESTING DO’s • Samples are tested on all shifts by technicians who routinely test patient samples • All surveys performed are treated as real patient samples. This includes using the same test methods for quality control, patient testing, workload accounting, verification and reporting of results. • NOTE: This only applies to analytes that are tested in-house. • Personnel who routinely analyze patient samples will integrate all PT within the routine workload

  6. Proficiency Testing DON’Ts • Test samples differently than patient specimens • Have the same people always do testing • Refer samples to another laboratory • Discuss results before reporting • Use PT sample for other purposes before submission date deadline

  7. ENROLLMENT • Required for all analytes listed in the PT Enrollment Guide - Updated annually - Available at CAP website www.cap.org -Included with order renewal forms

  8. Why perform Linearity and Calibration Verification??? • Requirements: CAP Laboratory Accreditation Program Checklist Question CHM.13400 (one of several relating to Calibration Verification) and Current CLIA Regulations Section 493.1255 for most analytes. • Linearity evaluations give information about what’s occurring “near the extremes," the high and low ends of the measurement range. • LN Surveys are particularly useful in calibration verification – identify imprecision problems before they become noticeable in other ways. • Calibration Verification provides proof that the measurement system calibration has not changed since the last calibration event.

  9. Why does AMR need to be verified? • Example: typical serum glucose method • Low calibrator = reagent blank (0 mg/dL) • High calibrator = 356 mg/dL • AMR = 10-600 mg/dL • The AMR is based on the expectation that the signal is linearly related to glucose concentration over the 10-600 mg/dL range. • Verification of the AMR expectation is required because the calibrators do not define the complete measurement range.

  10. Clinically Reportable Range • Extension of the AMR based on clinical requirements. • Established once when method is implemented • Does not need to be verified on a recurring schedule • Values outside the CRR are reported as < or > a numeric value. • Checklist question: Are dilution protocols and diluents (or concentration protocols) specified for all methods for which the CRR exceeds the AMR?

  11. AMR: how close to low and high limits • To verify acceptable clinical performance • For low limit: use clinical judgment, e.g. • creatinine, 0.2 or 0.3 mg/dL • For high limit: within 10-15% of upper limit of AMR

  12. Reviewing your LN Evaluation Report Example of LN2 Report (Calcium)

  13. Example of LN2 Report (Calcium)

  14. Example of a Non-Linear Result Loss of recovery at the high end of the AMR -Check reagents, calibrators, QC -Are the results clinically significant?

  15. PARTICIPATION • In addition to enrolling and testing samples, you must: -Ensure results are sent either electronically or via fax -Ensure they are sent by the deadline -Ensure all faxed pages are received NOTE: Non-participation (failure to receive results)= UNSATISFACTORY

  16. SATISFACTORY PT PERFORMANCE • 80% for regulated analytes and other analytes that have at least 5 challenges per PT event • Percentage varies for analytes with fewer than 5 challenges

  17. REPORTED CAUSES OF PT FAILURES

  18. CLERICAL ERRORS • Post Analytic Phase • Same importance as testing errors • Examples: • Transcription • Method/Reagent/Instrument codes • Missing Information (TNP, </>, etc)

  19. TECHNICAL ISSUES • Directly attributable to human actions: • Reconstitution • Specimen Mix-up • Improper Specimen Handling • Incorrect Instrument Handling • Failure to follow testing kit instructions • Morphologic Misinterpretations

  20. Methodologic Issues • Mechanical difficulties • Instrument Software Problems • Frequency of CALIBRATION • Inadequate Reagent Performance • Inadequate Maintenance • Other instrument malfunctions

  21. Exception Codes ANALYTICAL PROBLEM (Leave the result area blank and fill the bubble for the appropriate Exception Code). • 11 Unable to analyze (documentation be provided by laboratory). • 22 Result is outside method/ instrument reportable range. • 33 Specimen determined to be unsatisfactory after contacting the CAP. *It is the laboratory’s responsibility to document the appropriate use of these exception codes should this be requested during a laboratory inspection.

  22. Additional Tips • Keep original worksheets and instrument print-outs with the PT file • Make sure every less than acceptable result is followed up and documented • Confirm that corrective actions make sense and actually fix the problem

  23. UNSATISFACTORY PT PERFORMANCE • Laboratories that have unsatisfactory performance for 1 or more analytes on an event will receive PTES Exception Summary (PTES) report. • Investigate the problem • Determine cause • Implement Corrective Action • RESPONSE TO LAP is not required

  24. UNSUCCESSFUL PT PERFORMANCE • Is unsatisfactory performance on 2 of 3 PT events • Laboratory will receive a PT Exception Summary (PTES) report • Laboratory must suspend testing or implement plan of corrective action • LAP must approve laboratory’s action

  25. PT Monitoring by Accreditation Program

  26. CORRECTIVE ACTIONS • Design a process to verify clerical entries prior to result submission • Retrain testing personnel as to the proper procedures for sample preparation, testing and reporting • Modify/Narrow QC acceptance range to detect problems sooner • Evaluate/increase frequency of calibration • Perform Instrument Function Verification (Precision, Linearity, Correlation) • Revise Procedure to reflect corrective actions

  27. Example of Unsuccessful Proficiency Testing Performance:IgA FAILED on two out of 3 events. What will be your course of action: • Continue testing and provide CAP with credible program of corrective action • Voluntarily cease patient testing and investigate the failure • Conduct thorough investigation

  28. IgA Unity Laboratory Comparison Report <2.0 SDI/CVR

  29. IgA Levy Jennings POOR CALIBRATION

  30. Identified Cause of Failure: Poor Calibration Corrective Measures to prevent reoccurrence: • Daily review of calibration and QC • More frequent monitoring of LJ charts to catch trends and shifts that could indicate calibration curve issue • Method validation of IgA method was performed Precision Linearity Correlation Studies

  31. CONCLUSION: Proficiency Testing monitors laboratories' continuing performance in a way that: • –Improves patient care • –Protects the public’s health

  32. References: www.cap.org