Pulse Oximetry Screening: UK Survey and BWH Experience

Pulse Oximetry Screening: UK Survey and BWH Experience. Quad Network Study Day Anju Singh, SV Rasiah, Andy Ewer 29/11/2012. Neonatal pulse oximetry screening: a national survey. Kang et al. Arch Dis Child Fetal Neonatal Ed  2011;96:F312.

Télécharger la présentation

Pulse Oximetry Screening: UK Survey and BWH Experience

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Pulse Oximetry Screening: UK Survey and BWH Experience Quad Network Study Day Anju Singh, SV Rasiah, Andy Ewer 29/11/2012

  2. Neonatal pulse oximetry screening: a national survey. Kang et al. Arch Dis Child Fetal Neonatal Ed 2011;96:F312 7% routine use of pulse oximetry to supplement postnatal examination

  3. Pulse oximetry screening for critical congenital heart defectsin asymptomatic newborn babies: a systematic review andmeta-analysis.Shakila Thangaratinam, Kiritrea Brown, Javier Zamora, Khalid S Khan, Andrew K Ewer Lancet. 2012 Jun 30;379(9835):2459-64. doi: 10.1016/S0140-6736(12)60107-X. Epub 2012 May 2 • 13 eligible studies • Sensitivity: 76.5% (95% CI 67·7–83·5) • Specificity: 99·9% (99·7–99·9) • False-positive rate of 0·14% (0·06–0·33) • False-positive rate for critical CHD • Before 24 h 0·50 [0·29–0·86] • After 24 h 0·05% [0·02–0·12] p=0·0017 Pulse oximetry is highly specifi c for detection of critical congenital heart defects with moderate sensitivity, that meets criteria for universal screening.

  4. Pulse Oximetry Screening for critical congenital heart defects: A UK national surveySingh A , Ewer A • Who does the Routine Screening? • Who intends to do it? • Who doesn’t? • Anticipated Barriers?

  5. Pulse Oximetry Screening for critical congenital heart defects: A UK national surveySingh A , Ewer A • 204/ 204 (100%) Units responded • Routine screening: 36 (18%) units • In process of introducing screening: 8 units • Considering routine screening: 111 (70%) units

  6. Pulse Oximetry Screening for critical congenital heart defects: A UK national surveySingh A , Ewer A • Commonest Concerns • Resource issues: • Cost : 63% • Staff Time: 28% • Availability of Echocardiography: 25% • Staff Training: 24% • Lack of national and local guidelines: 36% • Excess False Positives: 10% • Delayed discharge: 5% • Cross infection: 3%

  7. Pulse Oximetry Screening for critical congenital heart defects: A UK national surveySingh A , Ewer A • Reasons for units not considering screening (49 units) • Staffing: 57% • False Positives: 55% • Availability of echocardiography: 33% • Cost: 31% • Unconvinced by evidence: 22% • Adequate current screening methods: 18%

  8. Pulse Oximetry Screening for critical congenital heart defects: A UK national surveySingh A , Ewer A • Threshold Saturation for positive test: 90-97% • 20/36: 95% • Postductal Saturations Only: 18 • Screening Time: • Before discharge: 55% • Within 48Hours: 4% • Before 24 hours: 13/14 • After 24 hours: 1

  9. Pulse Oximetry Screening for critical congenital heart defects: A UK national surveySingh A , Ewer A Conclusion Shift of opinion of among UK Neonatologists regarding pulse oximetry screening with a significant majority now in favour, albeit with some reservations

  10. The impact of pre- discharge pulse oximetry screening in a Regional Neonatal unit Singh A, Rasiah SV, Ewer A • To evaluate the impact of routine pulse oximetry screening on the rate of unexpected admissions and need for echocardiography. • To review the outcomes of babies admitted as a result of positive pulse oximetry screening.

  11. Routine pre-discharge pulse oximetry screening at BWH • Pre and postductal Sats • Abnormal Test: <95% or a difference of >2% • Expedited Clinical Examination • Repeat Pulse Ox in 1-2 hrs • Test Positive • 2 abnormal pulse ox readings • 1 abnormal pulse ox reading + abnormal Clinical Exam

  12. Methods • Retrospective review of all unexpected admissions to the unit • April’10 –March’12 • Review of • Indication for admission • Clinical diagnosis • Management • Outcome

  13. CHD Classification Results • Total admissions: 2137 • Unexpected admissions: 1021 • Test positive pulse oximetry: 123 (12%) • Congenital heart lesions: • Critical CHD: 4 • Serious CHD: 1 • Significant CHD: 3 • Critical - HLHS, PA/IVS, TGA or IAA or dying and/or intervention in 1st month with CoA, AS, PS, ToF, PA/VSD or TAPVD • Serious – Requiring intervention in 1st year • Significant - Requiring FU > 6 mths or drug Rx

  14. Other significant diagnosis • Congenital Pneumonia- ↑inflam markers ± +ve culture, X-Ray changes, O2 requirement, abs ≥ 5 days. • Sepsis - ↑ inflammatory markers ± culture +ve, abs ≥ 5 days • MAS – h/o meconium , respiratory distress, O2 requirement, X-Ray changes • TTN requiring oxygen: Tachypnoea with X-Ray changes of fluid retention, oxygen requirement, no rise in inflam markers or +ve culture • Congenital pneumonia: 33 • Sepsis: 17 • PPHN: 8 • MAS: 3 • TTN requiring oxygen: 21 • Hyperinsulinaemia: 1 • Pneumothorax: 1 • Depressed skull fracture: 1 • Early onset jaundice: 1

  15. Results • Transitional circulation: 29 (23%) • No collapse in the postnatal wards during study period • Echocardiograms performed for Test Postive pulse Ox: 39/123 (32%) • Abnormal ECHO’s: 16/39 (41%)

  16. Conclusions • Test positive pulse oximetry resulted in approx one admission per week • It leads to a modest increase in the number of echocardiograms performed. • Routine use of Pulse oximetry identifies babies with illnesses, which if not identified early could potentially lead to postnatal collapse

  17. References • Ewer AK, et al. Pulse oximetry as a screening test for congenital heart defects in newborn infants: the PulseOx test accuracy study. The Lancet 2011 Aug 27;378(9793):785-94. • Kang et al. Neonatal pulse oximetry screening: a national survey. Arch Dis Child Fetal Neonatal Ed 2011;96:F312.