1 / 15

2010 Tourniquet Working Group: Requirements & Test Parameters

2010 Tourniquet Working Group: Requirements & Test Parameters. CDR Tyson Brunstetter, MSC, USN Chief, Joint Medical Test & Evaluation Division Hemostatic Dressings & Tourniquets IPR 25 April 2012. Introduction. Tourniquet (TQ) Working Group Meeting 23 March 2010, Hosted by DMMPO

willis
Télécharger la présentation

2010 Tourniquet Working Group: Requirements & Test Parameters

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. 2010 Tourniquet Working Group: Requirements & Test Parameters CDR Tyson Brunstetter, MSC, USN Chief, Joint Medical Test & Evaluation Division Hemostatic Dressings & Tourniquets IPR 25 April 2012 CDR Tyson Brunstetter / 301-619-9028, DSN 343, tyson.brunstetter@amedd.army.mil Unclassified

  2. Introduction • Tourniquet (TQ) Working Group Meeting • 23 March 2010, Hosted by DMMPO • Goal: Reach consensus on TQ requirements & testing protocols • Today’s brief: To summarize recommended requirements and T&E parameters for extremity TQs CDR Tyson Brunstetter / 301-619-9028, DSN 343, tyson.brunstetter@amedd.army.mil Unclassified

  3. 2010 TQ WG: Consensus RQMTs CDR Tyson Brunstetter / 301-619-9028, DSN 343, tyson.brunstetter@amedd.army.mil Unclassified

  4. 2010 TQ WG: Consensus RQMTs -- REQUIRED ATTRIBUTES -- CDR Tyson Brunstetter / 301-619-9028, DSN 343, tyson.brunstetter@amedd.army.mil Unclassified

  5. 2010 TQ WG: Consensus RQMTs -- REQUIRED ATTRIBUTES -- CDR Tyson Brunstetter / 301-619-9028, DSN 343, tyson.brunstetter@amedd.army.mil Unclassified

  6. 2010 TQ WG: Consensus RQMTs -- DESIRED, BUT NOT REQUIRED ATTRIBUTES -- CDR Tyson Brunstetter / 301-619-9028, DSN 343, tyson.brunstetter@amedd.army.mil Unclassified

  7. 2010 TQ WG: Consensus RQMTs -- DESIRED, BUT NOT REQUIRED ATTRIBUTES -- CDR Tyson Brunstetter / 301-619-9028, DSN 343, tyson.brunstetter@amedd.army.mil Unclassified

  8. 2010 TQ WG: Consensus RQMTs -- DESIRED, BUT NOT REQUIRED ATTRIBUTES -- CDR Tyson Brunstetter / 301-619-9028, DSN 343, tyson.brunstetter@amedd.army.mil Unclassified

  9. 2010 TQ WG: Consensus Testing Efficacy-Screening Operational CDR Tyson Brunstetter / 301-619-9028, DSN 343, tyson.brunstetter@amedd.army.mil Unclassified

  10. 2010 TQ WG:Consensus Testing General Protocol – For Efficacy-Screening & OT • Collect baseline subject data (height, weight, limb circumference, baseline BP and HR), 20-min supine acclimation • Subject Clothing: Surgical scrubs and current Army ACUs • Tourniquets • Two conditions: (1) Dry and (2) soaked in blood analog solution • Tightening halted at cessation of Doppler signal; TQ pressure noted • Removed immediately if intolerable pain to the subject • Released slowly to confirm re-establishment of Doppler signal • Administrators/Subjects: Trained in TQ application prior to testing CDR Tyson Brunstetter / 301-619-9028, DSN 343, tyson.brunstetter@amedd.army.mil Unclassified

  11. 2010 TQ WG:Efficacy Testing/Screening Experiment 1 Locate & mark lower extremity popliteal artery Locate & mark location of TQ application @ mid-thigh Apply TQ, measure arterial blood flow at popliteal artery site via Doppler ultrasonography Effectiveness defined as cessation of Doppler signal Effective devices move on to Experiment 2 CDR Tyson Brunstetter / 301-619-9028, DSN 343, tyson.brunstetter@amedd.army.mil Unclassified

  12. 2010 TQ WG:Efficacy Testing/Screening Experiment 2 Locate & mark radial artery Locate & mark location of TQ application @ mid-humerus Apply TQ, measure arterial blood flow at radial artery site via Doppler ultrasonography Effectiveness defined as cessation of Doppler signal CDR Tyson Brunstetter / 301-619-9028, DSN 343, tyson.brunstetter@amedd.army.mil Unclassified

  13. 2010 TQ WG:Operational Testing Test Procedure: Upper & Lower Extremities Locate & mark extremity artery Prior to application of tourniquet, subject performs 20, eight-count body builders (jumping jack, push-up, squat jump, stand) to simulate operational duress Immediately apply TQ to own limb Measure arterial blood flow via Doppler Effectiveness defined as cessation of Doppler signal Survey assesses training, packaging, and ease of use CDR Tyson Brunstetter / 301-619-9028, DSN 343, tyson.brunstetter@amedd.army.mil Unclassified

  14. 2010 TQ WG:Operational Testing Test Procedure: Repeated w/ following variables Blood solution-soaked TQ w/dominant hand Dry TQ w/dominant hand Blood solution-soaked TQ w/non-dominant hand Dry TQ w/non-dominant hand Blood solution-soaked TQ using Night Vision Goggles Dry TQ using Night Vision Goggles CDR Tyson Brunstetter / 301-619-9028, DSN 343, tyson.brunstetter@amedd.army.mil Unclassified

  15. Questions? CDR Tyson Brunstetter Tyson.Brunstetter@amedd.army.mil (301) 619-9028 CDR Tyson Brunstetter / 301-619-9028, DSN 343, tyson.brunstetter@amedd.army.mil Unclassified