Dialysis Water Treatment Systems

Dialysis Water Treatment Systems. Rational, Function, Water Testing and Documentation. Introduction.

Télécharger la présentation

Dialysis Water Treatment Systems

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Dialysis Water Treatment Systems Rational, Function, Water Testing and Documentation

  2. Introduction • AAMI (Association for the Advancement of Medical Instrumentation) since 1982 has set the standards for all Hemodialysis Equipment, Water Treatment, Hemodialysis Concentrates and Hemodialyzer Reprocessing • AAMI states that the medical director has ultimate responsibility for the water treatment system and all testing and performance of that system • Water Treatment Systems installed after May 30,1997 must are subject to the FDA 510(k) approval process • It is imperative that a dialogue be maintained with the feedwater suppliers

  3. Rational Hemodialysis Water Treatment Standards are an evolutionary process dictated by data and information gathered on the effects of water on patient mortality and morbidity. Sadly many patients have died over the years as a result of ignorance, carelessness , bad luck and the failure to follow the standards then in effect. Education of all members of the dialysis treatment team will allow us to be more cognizant of the dangers involved in the treatment process, and encourage us to do our best for our patients

  4. Temperature Blending Valve • Mixes hot and cold water to maintain a constant temperature of 77 F or 25 C • This prevents harm to patients and damage to the RO (reverse osmosis ) membranes • Product water output will decrease as the temperature decreases • A temperature gauge is placed downstream and is monitored

  5. Blending Valve and Booster Pump

  6. Booster Pump • Maintains a constant flow and pressure to the Pretreatment components and the RO • Pressure gauges are placed before and after the pump and are monitored

  7. Sediment Filters • Filters larger particles from the feedwater • This protects the downstream pretreatment equipment and the RO • This system is routinely backwashed at night by the use of an automatic timer and is connected to the RO with lockout protection • Pre and post pressure gauges are used and monitored

  8. Sediment Filter

  9. Water Softener • Removes Calcium and Magnesium from the feedwater to prevent scaling on the Carbon filters and the RO • An ion exchange resin is used that attracts Calcium and Magnesium and releases a Sodium ion • The softener is regenerated at night by an automatic timer and is connected to the RO with lockout protection • Pre and post pressure gauges are used and monitored • Water hardness testing is done and documented at the beginning and end of the treatment day

  10. Water Softener

  11. Carbon Filters • Removes chlorine, chloramines and organic compounds by the use of GAC (granular activated charcoal) • Two tanks in series are used • Feedwater must contact the carbon for a EBTC (empty bed contact time) of at least 10 minutes • The carbon cannot be regenerated, but a flushing is done at night with an automatic timer • Pre and post pressure gauges are used and monitored • Chlorine and chloramine testing must done and documented before each patient treatment

  12. Carbon Filters

  13. Reverse Osmosis System • Purifies large volumes of relatively pure water at low cost • Usually has a 5 micron filter that removes carbon fines and resin beads • Water pump forces pretreated water at high pressure through semi-permeable membranes • Efficiently removes > 95% of ions • Removes bacteria but not endotoxins • Has alarm system to protect patient

  14. RO Data Sheet

  15. RO Control Box

  16. RO Display Screen

  17. DI • Deionization is generally used as a backup in case of RO failure • Most systems use a dual bed (cation and anion) system of ion exchange resins • Produces water purer than a RO but is very expensive • Does not remove bacteria or endotoxins • Must be regenerated offsite from vendors that deal with medical or food grade systems • Resistivity must be continuously monitored to 1 megohm with an audible and visual alarm in patient area and a product stop or divert

  18. DI Control Box and Resistivity Meter

  19. Ultrafilters • Used to remove bacteria and endotoxins from product water • Required downstream from a DI system

  20. Ultrafilters

  21. Ultraviolet Irradiators • Used to control bacteria in a water storage and distribution system • The intensity of the UV light must be monitored and alarmed (visual) to insure sufficient kill intensity • Ultrafilters are recommended downstream from the UV light

  22. RO Water Storage Tank • Water storage tanks are used when a direct feed system isn’t feasible or recommended • Must have a conical bottom, tight fitting lid and a hydrophobic 0.2 micron air filter • The recirculating pumps run 24/7 to keep bacteria from forming on the walls of the tank and in the • The water flow velocity must be at least 3 feet per second (1.5 feet per second for direct feed)

  23. RO Water Storage

  24. Water Testing AAMI Chemical Analysis • A test of 21 chemical contaminants performed at least once a year ,and when there are feed water changes. See AAMI standards for limits Bacteria Colony counts • Performed monthly per AAMI standards • Action Level > 50 CFU/mL • Maximum Level 200 CFU/mL LAL Endotoxin • Performed monthly per AAMI standards • Action Level > 1.0 EU/mL • Maximum Level 2.0 EU/mL An Action process must de developed to deal with the situation when limits for any water tests are exceeded

  25. Documentation • Chemical Contaminants, Bacteria Colony Counts and LAL Endotoxin should be consolidated to allow for trend analysis and government audits • A Water Treatment System Log should include but not limited to data from all pressure and temperature gauges, RO data screen, post softener hardness results, brine tank inspection and action, chloramine test results and any action, filter change, cultures performed, cleaning and/or disinfection completed on the water treatment system (see handouts)