1 / 8

Measurements and Comparisons of IPMs for Ecloud Studies in the LHC: 2007 Results

This report details the measurements conducted using the IPM (Interferometric Profile Monitor) systems in 2007, focusing on both the IPMH and IPMV versions for controlled transverse blow-up studies. Specific cases include one bunch on a long injection flat bottom, four bunches accelerated to 450 GeV/c, and three bunches up to 270 GeV/c. The results reveal significant variances in measured values compared to BWS, particularly under varying energy conditions. Conclusions indicate the need for a detailed comparison and further analysis in future studies, solidifying the IPM’s role in real-time observation of emittance changes.

yuma
Télécharger la présentation

Measurements and Comparisons of IPMs for Ecloud Studies in the LHC: 2007 Results

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. SPS MEASUREMENTS WITH IPMs IN 2007 J. Koopman and E. Métral • SFTPRO1 • LHC 25 ns (without and with controlled transverse blow-up for ecloud studies) • IONS • 1 bunch on the long injection flat bottom • 4 bunches up to 450 GeV/c • 3 bunches up to 270 GeV/c • Conclusions and outlook

  2. SFTPRO1 (07/11/07) Gaussian of constant normalized emittance in blue x = 3.56 mm BIPMH51634 x = 87.9 m x = 0.81 mm

  3. LHC 25 ns (08/11/07) WITHOUT controlled BU WITH controlled BU (ecloud studies) Gaussian of constant normalized emittance in blue x = 2.68 mm x = 1.51 mm x = 2.10 mm

  4. IONS: 1 bunch on the long injection flat bottom (30/10/07) BIPMV51734 y = 91.2 m y  2.7 mm Comparison with BWS meas.: ~ 0.55 m  Good agreement!

  5. IONS: 4 bunches up to 450 GeV/c (30/10/07) y = 3.2 mm y = 1.3 mm Comparison with BWS meas.: After 1st inj.  0.55 m 12800 ms  0.83 m  ~ 4 times bigger value at 450 GeV/c with IPMV!

  6. IONS: 3 bunches up to 270 GeV/c (24/10/07) After 2nd inj After 1st inj y = 2.99 mm y = 3.00 mm After 3rd inj y = 3.25 mm y = 1.28 mm  ~ 2.5 times bigger value at 270 GeV/c with IPMV!

  7. CONCLUSIONS AND OUTLOOK (1/2) • Measurements have been performed in 2007 with both IPMH and IPMV (not working at the same time!) • Good agreement found with BWS at low energy • However, (with the ions) at top energy bigger values were measured with the IPMs wrt the BWS • At 270 GeV/c  A factor of ~ 2.5 • At 450 GeV/c  A factor of ~ 4 • Another issue which has to be understood is the value of the signal after multiple injections

  8. CONCLUSIONS AND OUTLOOK (2/2) • Next year a detailed comparison between the IPMs and the BWS should be made  The IPM is really a nice tool to observe “online” any blow-up of the transverse emittance (or modification of the profile) at any time along the cycle WITH controlled BU (ecloud studies) Gaussian of constant normalized emittance in blue

More Related