Slim Sam on Powgen. Top: Diffraction pattern of sample in magnet. . Bottom: Diffraction pattern of sample where the background has been subtracted. . The 5T (vertical field) magnet Slim Sam was tested in powgen in February. The in plane detectors on powgen can be used for this magnet

zofia
Télécharger la présentation

Slim Sam on Powgen

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Slim Sam on Powgen Top: Diffraction pattern of sample in magnet. Bottom: Diffraction pattern of sample where the background has been subtracted. The 5T (vertical field) magnet Slim Sam was tested in powgen in February. • The in plane detectors on powgen can be used for this magnet • We observe Al diffraction from the VTI in all detectors even with the fine radial collimator. • Background subtraction can work in certain cases, but background data will need to be collected for all temperatures that sample is measured at. • Since the background diffraction peaks are all from Al, no peaks is observed beyond 2.3Å. • Users wanting to use magnet on powgen should contact Ashfia Huq to determine feasibility.