1 / 32

SC-Analyzer Enclosure Concept Exploded view

SC-Analyzer Enclosure Concept Exploded view. Integrated heating and ventilation. Insulated enclosure. Insulated door. Reagent storage with easy access. Analytical and electronic section . AMTAX sc Analytical section. 3. (1) GSE: Gas Sensitive Electrode as detector

kirby
Télécharger la présentation

SC-Analyzer Enclosure Concept Exploded view

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. SC-Analyzer Enclosure ConceptExploded view Integrated heating and ventilation Insulated enclosure Insulated door Reagent storage with easy access Analytical and electronic section

  2. AMTAX sc Analytical section 3 • (1) GSE: Gas Sensitive Electrode as detector • (2) Air pump to move liquids • (3) Dosing pump for reagent • (4) Electrolyte storage • (5) Cleaning solution • (6) 2 standard solutions • (7) Reagent • (8) Membrane storage 2 1 4 5 8 6 6 7

  3. Slot for electrode Overflow Seal Heating element Seal Motor holder (PMMA) Motor with Chamber for Inlet Inlet Inlet Fixing Magnet holder Solenoid stirring stick Sample Citric acid NaOH O ring seal Amtax sc: Measuring Chamber Rear side Heating element Temp.-observation Temp. sensor

  4. AMTAX sc Gas Sensitive Electrode (GSE) (1) Electrode (2) Removable membrane cap (3) Gas Sensitive Membrane

  5. mV  pH Electrode Electrolyte NH3 Gas selective Membrane Sample, NH4+converts to NH3 Addition of NaOH: pH  to ~12 Amtax sc: Measuring Principle NH3

  6. AMTAX Guacolda

  7. Sistema de Dosificación en el AMTAX Bomba Solución de Reactivos y limpieza Pistón Succión Estándar y muestra Válvulas fluidos Estándar y Muestra

  8. Preparación Electrodo de Medición en el AMTAX Nivel de Electrolito

  9. Bombas Dosificadoras 2 Bombas Reactivos y Limpieza Autor: Milton Martinez

  10. Bombas Dosificadoras 1.- Salida Limpieza 2 Entrada Limpieza 3.- Salida Reactivo 4.- Entrada Reactivo 1

  11. Prueba Manual de Bombas Docificadoras Test Manual Bombas: -Reactivos -Limpieza

  12. Menú de Mantención Relleno Solución -Limpieza -Reactivo -Estándar Reemplazo Electrodo/Electrolito/Membrana

  13. Datos del Electrodo

  14. Maintenance

  15. Status Indication

  16. Analizador de Amonio, Rutina de remplazo de reactivos

  17. Analizador de Amonio, Rutina de remplazo de reactivos