Eichrom Resin Columns: A New breakthrough

Eichrom Resin Columns: A New breakthrough. Anil Thakkar Eichrom Users’ Group Workshop May 3, 2005 Oak Ridge, TN. Background details. Stripping issues of Th on TEVA prepacked columns reported by numerous customers

Pat_Xavi
Télécharger la présentation

Eichrom Resin Columns: A New breakthrough

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Eichrom Resin Columns: A New breakthrough Anil Thakkar Eichrom Users’ Group Workshop May 3, 2005 Oak Ridge, TN

  2. Background details • Stripping issues of Th on TEVA prepacked columns reported by numerous customers • Investigated all the components of the prepacaked column to trouble shoot the problem: resin, column body, resin bead size, frits….

  3. Previously discussed investigations…. • Column frits identified as the main cause of the problem • Investigations on frits led to the understanding of different frit types: • Frits with ‘ionic surfactant’ • Impact Th elution from TEVA columns • Impact low level U elution on UTEVA columns • Frits with ‘no surfactant’ • Measured slow flow rates at gravity • Mechanical pressure required to start the flow • Measured inconsistent flow rates

  4. Additional issues reported • Poor elution of low level U on UTEVA columns reported • An inconsistent elution of Th on UTEVA columns has been observed

  5. Additional research continues… A ‘non-ionic surfactant’ frit material found and put to test

  6. TEVA Column: Performance Test • Columns packed with non-ionic frit material • Measured performance of columns over a period oftime

  7. TEVA Columns: % Th Yields

  8. TEVA Columns: % Pu Yields with 0.5M HCl* *INEEL procedure uses 0.5M HCl to elute Pu

  9. TEVA Columns: % Am Yields(Am/ Lanthanide Separation)

  10. TEVA Columns: % Tc-99 Yields

  11. UTEVA, TRU, Sr and Ln columns tested after 2 weeks of packing

  12. UTEVA Columns: % U & Th Yields *Th eluted with 5M HCl/0.05M oxalic acid

  13. TRU Columns: % Am & Pu Yields

  14. Sr Columns: % Sr Yields & Recoveries

  15. Ln Columns: % Ac-228 Yields

  16. Improvements & Summary • Non-ionic surfactant frits improve performance • Th elution on TEVA • Th elution on UTEVA • Pu elution on TEVA • No impact observed on the performance of the following products with non-ionic frits • Am & Pu on TRU resin • Tc-99, Am on TEVA resin • Sr elution on Sr resin • Ac elution on Ln resin

  17. Future resin column inventory • Preliminary evaluation of prepacked column products with the new frits (non-ionic surfactant) has been extremely satisfactory • New frit inventory for future production has been ordered • Current inventory has old frits (ionic surfactant) in the prepacked column products • Any users of TEVA resin and UTEVA resin prepacked columns may consider cartridges in the interim • Contact Eichrom for further assistance