Eichrom Resin Columns: A New breakthrough
180 likes | 861 Vues
Eichrom Resin Columns: A New breakthrough. Anil Thakkar Eichrom Users’ Group Workshop May 3, 2005 Oak Ridge, TN. Background details. Stripping issues of Th on TEVA prepacked columns reported by numerous customers
Eichrom Resin Columns: A New breakthrough
E N D
Presentation Transcript
Eichrom Resin Columns: A New breakthrough Anil Thakkar Eichrom Users’ Group Workshop May 3, 2005 Oak Ridge, TN
Background details • Stripping issues of Th on TEVA prepacked columns reported by numerous customers • Investigated all the components of the prepacaked column to trouble shoot the problem: resin, column body, resin bead size, frits….
Previously discussed investigations…. • Column frits identified as the main cause of the problem • Investigations on frits led to the understanding of different frit types: • Frits with ‘ionic surfactant’ • Impact Th elution from TEVA columns • Impact low level U elution on UTEVA columns • Frits with ‘no surfactant’ • Measured slow flow rates at gravity • Mechanical pressure required to start the flow • Measured inconsistent flow rates
Additional issues reported • Poor elution of low level U on UTEVA columns reported • An inconsistent elution of Th on UTEVA columns has been observed
Additional research continues… A ‘non-ionic surfactant’ frit material found and put to test
TEVA Column: Performance Test • Columns packed with non-ionic frit material • Measured performance of columns over a period oftime
TEVA Columns: % Pu Yields with 0.5M HCl* *INEEL procedure uses 0.5M HCl to elute Pu
UTEVA, TRU, Sr and Ln columns tested after 2 weeks of packing
UTEVA Columns: % U & Th Yields *Th eluted with 5M HCl/0.05M oxalic acid
Improvements & Summary • Non-ionic surfactant frits improve performance • Th elution on TEVA • Th elution on UTEVA • Pu elution on TEVA • No impact observed on the performance of the following products with non-ionic frits • Am & Pu on TRU resin • Tc-99, Am on TEVA resin • Sr elution on Sr resin • Ac elution on Ln resin
Future resin column inventory • Preliminary evaluation of prepacked column products with the new frits (non-ionic surfactant) has been extremely satisfactory • New frit inventory for future production has been ordered • Current inventory has old frits (ionic surfactant) in the prepacked column products • Any users of TEVA resin and UTEVA resin prepacked columns may consider cartridges in the interim • Contact Eichrom for further assistance